Radiation dose assessement on dual and 16 slices MDCT using Monte Carlo simulation

被引:0
|
作者
Mkimel, M. [1 ,2 ]
Mesradi, M. R. [1 ]
El Baydaoui, R. [1 ]
Toufique, Y. [1 ]
Halimi, A. [1 ]
Hilal, M. [1 ]
Saad, E. [1 ]
Hilali, A. [1 ]
机构
[1] Hassan 1st Univ, High Inst Hlth Sci, Lab Technol & Med Sci, Settat, Morocco
[2] Publ Hosp, Dept Radiol, Settat, Morocco
关键词
CTDI; GATE; MDCT; PMMA; effective dose; CONVERSION FACTORS; GATE;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Computed Tomography Dose Index (CTDI) is an important factor to estimate effective dose using the conversion factors. In this work, we estimated CTDI with Monte Carlo simulation using the (GATE/GEANT4) code. Two PMMA phantom diameters, 16 cm and 32 cm, have been used to simulate, respectively, the patient's head and body. The simulation was made at various kVp and mAs levels for dual and 16 slices Multi Detector Computed Tomography MDCT. Experimental CTDI measurements were performed with a calibrated pencil ion chamber and the two PMMAs. the comparison results between the simulated, measured and displayed values show slight differences. Indeed, a difference of less than 4% is noticed for the 2 slices and less than 2% for the 16 slices. The effectiveness of our method shows that GATE can be used to optimize the effective dose to reduce radiological risks in CT clinical applications.
引用
收藏
页码:885 / 889
页数:5
相关论文
共 50 条
  • [1] Calculation of radiation dose to the lens of the eye using Monte Carlo simulation
    Meriç, N
    APPLIED RADIATION AND ISOTOPES, 2001, 55 (04) : 557 - 560
  • [2] Variability of surface and center position radiation dose in MDCT: Monte Carlo simulations using CTDI and anthropomorphic phantoms
    Zhang, Di
    Savandi, Ali S.
    Demarco, John J.
    Cagnon, Chris H.
    Angel, Erin
    Turner, Adam C.
    Cody, Dianna D.
    Stevens, Donna M.
    Primak, Andrew N.
    McCollough, Cynthia H.
    McNitt-Gray, Michael F.
    MEDICAL PHYSICS, 2009, 36 (03) : 1025 - 1038
  • [3] Monte Carlo simulation to assess fetal dose from MDCT imaging using patient based voxelized models
    Angel, E.
    Wellnitz, C.
    Goodsitt, M.
    DeMarco, J.
    Cagnon, C.
    Cody, D.
    Stevens, D.
    McCollough, C.
    Primak, A.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2006, 33 (06) : 2211 - 2211
  • [4] Scatter radiation dose profile evaluation in computed tomography using Monte Carlo simulation
    Mohammadi, Gh R. Fallah
    Hesamnezhad, L.
    Mahdavi, M.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (04): : 813 - 818
  • [5] Radiation dose from MDCT using Monte Carlo simulations: Estimating fetal dose due to pulmonary embolism scans accounting for overscan
    Angel, E.
    Wellnitz, C.
    Goodsitt, M.
    DeMarco, J.
    Cagnon, C.
    Ghatali, M.
    Cody, D.
    Stevens, D.
    McCollough, C.
    Primak, A.
    McNitt-Gray, M.
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [6] SBRT Dose Verification Using Monte Carlo Simulation
    Luo, W.
    Xie, X.
    McGarry, R.
    Molloy, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3796 - 3796
  • [7] THE EFFECT OF GOLD NANOPARTICLES ON RADIATION DOSE DISTRIBUTION IN BREAST CANCER USING MONTE CARLO SIMULATION
    Bush, H. S.
    Eisa, M. H.
    Ramizy, A.
    Ashari, M.
    Haroun, Kh M.
    Aldaghri, O.
    Abd-Alla, M. D.
    JOURNAL OF OPTOELECTRONIC AND BIOMEDICAL MATERIALS, 2020, 12 (04): : 101 - 108
  • [8] Radiation dose distribution around a mobile linear accelerator Mobetron using Monte Carlo simulation
    Garcia-Cases, F.
    Patino, G.
    Perez-Calatayud, J.
    Ballester, F.
    Granero, D.
    Vijande, J.
    2016 GLOBAL MEDICAL ENGINEERING PHYSICS EXCHANGES/PAN AMERICAN HEALTH CARE EXCHANGES (GMEPE/PAHCE), 2016,
  • [9] Monte Carlo simulation of breast imaging using synchrotron radiation
    Fitousi, N. T.
    Delis, H.
    Panayiotakis, G.
    MEDICAL PHYSICS, 2012, 39 (04) : 2069 - 2077
  • [10] Monte Carlo Simulation of Dose Enhancement in Gold Nanoparticle Mediated Radiation Therapy
    Bashir, S.
    Koreshi, Z.
    MEDICAL PHYSICS, 2017, 44 (06) : 2982 - 2982