Flow Test to Predict Early Hypotony and Hypertensive Phase After Ahmed Glaucoma Valve (AGV) Surgical Implantation
被引:10
|作者:
Cheng, Jason
[1
]
Beltran-Agullo, Laura
[2
]
Buys, Yvonne M.
[3
]
Moss, Edward B.
[3
]
Gonzalez, Johanna
[3
]
Trope, Graham E.
[3
]
机构:
[1] Khoo Tech Puat Hosp, Singapore, Singapore
[2] Inst Catala Retina, Barcelona, Spain
[3] Univ Toronto, Dept Ophthalmol & Vis Sci, Toronto Western Hosp, Toronto, ON, Canada
关键词:
Ahmed glaucoma valve;
glaucoma;
surgery;
hypotony;
hypertensive phase;
CLINICAL-EXPERIENCE;
DRAINAGE DEVICES;
CLOSING PRESSURE;
MOLTENO IMPLANT;
SINGLE-PLATE;
IN-VITRO;
BAERVELDT;
POLYPROPYLENE;
OUTCOMES;
D O I:
10.1097/IJG.0000000000000318
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
Purpose: To assess the validity of a preimplantation flow test to predict early hypotony [intraocular pressure (IOP) <= 5mm Hg on 2 consecutive visits and hypertensive phase (HP) (IOP > 21 mm Hg) after Ahmed Glaucoma Valve (AGV) implantation. Patients and Methods: Prospective interventional study on patients receiving an AGV. A preimplantation flow test using a gravity-driven reservoir and an open manometer was performed on all AGVs. Opening pressure (OP) and closing pressure (CP) were defined as the pressure at which fluid was seen to flow or stop flowing through the AGV, respectively. OP and CP were measured twice per AGV. Patients were followed for 12 weeks. Results: In total, 20 eyes from 19 patients were enrolled. At 12 weeks the mean IOP decreased from 29.2 +/- 9.1 to 16.8 +/- 5.2 mm Hg (P < 0.01). The mean AGV OP was 17.5 +/- 5.4 mm Hg and the mean CP was 6.7 +/- 2.3 mm Hg. Early (within 2 wk postoperative) HP occurred in 37% and hypotony in 16% of cases. An 18 mm Hg cutoff for the OP gave a sensitivity of 0.71, specificity of 0.83, positive predictive value of 0.71, and negative predictive value of 0.83 for predicting an early HP. A 7 mm Hg cutoff for the CP yielded a sensitivity of 1.0, specificity of 0.38, positive predictive value of 0.23, and negative predictive value of 1.0 for predicting hypotony. Conclusions: Preoperative OP and CP may predict early hypotony or HP and may be used as a guide as to which AGV valves to discard before implantation surgery.
引用
收藏
页码:493 / 496
页数:4
相关论文