Rapid determination of C4-acylcarnitine and C5-acylcarnitine isomers in plasma and dried blood spots by UPLC-MS/MS as a second tier test following flow-injection MS/MS acylcarnitine profile analysis

被引:33
作者
Forni, Sabrina [1 ]
Fu, Xiaowei [1 ]
Palmer, Susan E. [2 ]
Sweetman, Lawrence [1 ]
机构
[1] Baylor Res Inst, Inst Metab Dis, Dallas, TX 75226 USA
[2] Univ Oklahoma, HSC, Dept Pediat, Genet Sect, Oklahoma City, OK 73104 USA
关键词
Short-chain acylcarnitines; Short-chain acyl-CoA dehydrogenase deficiency; Isovaleric acidemia; Isobutyryl-CoA dehydrogenase deficiency; 2-Methylbutyryl-CoA dehydrogenase deficiency; Pivaloylcarnitine; Ultra performance liquid chromatography mass spectrometry; TANDEM MASS-SPECTROMETRY; DEHYDROGENASE-DEFICIENCY; ISOVALERIC ACIDEMIA; DIAGNOSIS; POPULATION; METABOLISM;
D O I
10.1016/j.ymgme.2010.05.012
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Flow-injection MS/MS methods for elevated acylcarnitines are routinely performed in most newborn screening and biochemical genetics laboratories; however this technique cannot distinguish between isobaric compounds; therefore, chromatographic separation is required to quantitate isomers for differential diagnosis of some inborn errors of metabolism. Methods: A UPLC-MS/MS method has been developed for the simultaneous quantitation of isobutyrylcarnitine and butyrylcarnitine, and a second UPLC-MS/MS method for the quantitation of isovalerylcarnitine, (S) and (R) 2-methylbutyrylcarnitine, pivaloylcarnitine and valerylcarnitine. Plasma and dried blood spots samples are extracted with methanol and derivatized with butanolic HCI. Deuterium labeled internal standards are used for quantitation. Separation is obtained using a methanol/water gradient with a C18 BEH, 1 x 100 mm, 1.7 mu m UPLC column, at 60 degrees C; run time is less than 10 min. The isomers are detected with a Quattro Premier triple quadrupole, with electrospray ionization in positive ion mode. Results: Intra-day precision in plasma and dried blood spots ranged from 1.4% to 14% and accuracy from 88% to 114% respectively for butyrylcarnitine and isobutyrylcarnitine. Precision for the isomers of C5-acylcarnitine ranged from 1.3% to 15% and accuracy 87% to 119%, respectively in plasma or dried blood spots. Inter-day precision was within 20% at each concentration of isobutyrylcarnitine and butyrylcarnitine. Precision for 2-methylbutyrylcarnitine and isovalerylcarnitine at concentrations above the normal range was within 24%. Conclusions: Two diagnostic tests based on the separation of C4-acylcarnitine and C5-acylcarnitine isomers by UPLC-MS/MS provide fast differential diagnosis of SCAD deficiency versus IBCD deficiency and IVA versus 2-MBCD deficiency. The separation of C5-acylcarnitines can reveal false elevation due to pivalic acid-containing antibiotics. Abnormal newborn screen results due to pivalate-generating prodrug antibiotics of maternal origin were confirmed. This separation of isomers can resolve multiple diagnostic challenges in both newborn screening and in cases with ambiguous metabolic test results. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:25 / 32
页数:8
相关论文
共 16 条
[1]  
Abendur JE, 1998, J INHERIT METAB DIS, V21, P624
[2]   Pivalate-generating Prodrugs and carnitine homeostasis in man [J].
Brass, EP .
PHARMACOLOGICAL REVIEWS, 2002, 54 (04) :589-598
[3]   The application of tandem mass spectrometry to neonatal screening for inherited disorders of intermediary metabolism [J].
Chace, DH ;
Kalas, TA ;
Naylor, EW .
ANNUAL REVIEW OF GENOMICS AND HUMAN GENETICS, 2002, 3 :17-45
[4]  
Fernandes J., 2006, INBORN METABOLIC DIS, VFourth
[5]   Separation and identification of plasma short-chain acylcarnitine isomers by HPLC/MS/MS for the differential diagnosis of fatty acid oxidation defects and organic acidemias [J].
Ferrer, Isaac ;
Ruiz-Sala, Pedro ;
Vicente, Yolanda ;
Merinero, Begona ;
Perez-Cerda, Celia ;
Ugarte, Magdalena .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 860 (01) :121-126
[6]   Simultaneous quantification of acylcarnitine isomers containing dicarboxylic acylcarnitines in human serum and urine by high-performance liquid chromatography/electrospray ionization tandem mass spectrometry [J].
Maeda, Yasuhiro ;
Ito, Tetsuya ;
Suzuki, Atsuko ;
Kurono, Yukihisa ;
Ueta, Akihito ;
Yokoi, Kyoko ;
Sumi, Satoshi ;
Togari, Hajime ;
Sugiyama, Naruji .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (05) :799-806
[7]   Falsely elevated C4-carnitine as expression of glutamate formiminotransferase deficiency in tandem mass spectrometry newborn screening [J].
Malvagia, S ;
La Marca, G ;
Casetta, B ;
Gasperini, S ;
Pasquini, E ;
Donati, MA ;
Zammarchi, E .
JOURNAL OF MASS SPECTROMETRY, 2006, 41 (02) :263-265
[8]   Prospective diagnosis of 2-methylbutyryl-CoA dehydrogenase deficiency in the Hmong population by newborn screening using tandem mass spectrometry [J].
Matern, D ;
He, M ;
Berry, SA ;
Rinaldo, P ;
Whitley, CB ;
Madsen, PP ;
van Calcar, SC ;
Lussky, RC ;
Andresen, BS ;
Wolff, JA ;
Vockley, J .
PEDIATRICS, 2003, 112 (01) :74-78
[9]   The frequency of short-chain acyl-CoA dehydrogenase gene variants in the US population and correlation with the C4-acylcarnitine concentration in newborn blood spots [J].
Nagan, N ;
Kruckeberg, KE ;
Tauscher, AL ;
Bailey, KS ;
Rinaldo, P ;
Matern, D .
MOLECULAR GENETICS AND METABOLISM, 2003, 78 (04) :239-246
[10]   Development of a newborn screening follow-up algorithm for the diagnosis of isobutyryl-CoA dehydrogenase deficiency [J].
Oglesbee, Devin ;
He, Miao ;
Majumder, Nilanjana ;
Vockley, Jerry ;
Ahmad, Ayesha ;
Angle, Brad ;
Burton, Barbara ;
Charrow, Joel ;
Ensenauer, Regina ;
Ficicioglu, Can H. ;
Keppen, Laura Davis ;
Marsden, Deborah ;
Tortorelli, Silvia ;
Hahn, Si Houn ;
Matern, Dietrich .
GENETICS IN MEDICINE, 2007, 9 (02) :108-116