Comparison of the Determination of a Low-Concentration Active Ingredient in Pharmaceutical Tablets by Backscatter and Transmission Raman Spectrometry

被引:19
作者
Townshend, Nichola [1 ,2 ]
Nordon, Alison [1 ,2 ]
Littlejohn, David [1 ,2 ]
Myrick, Michael [3 ]
Andrews, John [4 ]
Dallin, Paul [4 ]
机构
[1] Univ Strathclyde, Dept Pure & Appl Chem, WestCHEM, Glasgow G1 1XL, Lanark, Scotland
[2] Univ Strathclyde, CPACT, Glasgow G1 1XL, Lanark, Scotland
[3] Univ S Carolina, Dept Chem & Biochem, Columbia, SC 29208 USA
[4] Clairet Sci Ltd, Northampton NN3 6AP, England
基金
英国工程与自然科学研究理事会;
关键词
QUANTITATIVE-DETERMINATION; SPECTROSCOPY; STATE; RESOLUTION; SUBSTANCE; DEPTH; FORMS;
D O I
10.1021/ac203447k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A total of 383 tablets of a pharmaceutical product were analyzed by backscatter and transmission Raman spectrometry to determine the concentration of an active pharmaceutical ingredient (API), chlorpheniramine maleate, at the 2% m/m (4 mg) level. As the exact composition of the tablets was unknown, external calibration samples were prepared from chlorpheniramine maleate and microcrystalline cellulose (Avicel) of different particle size. The API peak at 1594 cm(-1) in the second derivative Raman spectra was used to generate linear calibration models. The API concentration predicted using backscatter Raman measurements was relatively insensitive to the particle size of Avicel. With transmission, however, particle size effects were greater and accurate prediction of the API content was only possible when the photon propagation properties of the calibration and sample tablets were matched. Good agreement was obtained with HPLC analysis when matched calibration tablets were used for both modes. When the calibration and sample tablets are not chemically matched, spectral normalization based on calculation of relative intensities cannot be used to reduce the effects of differences in physical properties. The main conclusion is that although better for whole tablet analysis, transmission Raman is more sensitive to differences in the photon propagation properties of the calibration and sample tablets.
引用
收藏
页码:4671 / 4676
页数:6
相关论文
共 44 条
[11]   Picosecond time-resolved Raman spectroscopy of solids: Capabilities and limitations for fluorescence rejection and the influence of diffuse reflectance [J].
Everall, N ;
Hahn, T ;
Matousek, P ;
Parker, AW ;
Towrie, M .
APPLIED SPECTROSCOPY, 2001, 55 (12) :1701-1708
[12]   Measurement of Spatial Resolution and Sensitivity in Transmission and Backscattering Raman Spectroscopy of Opaque Samples: Impact on Pharmaceutical Quality Control and Raman Tomography [J].
Everall, Neil ;
Priestnall, Ian ;
Dallin, Paul ;
Andrews, John ;
Lewis, Ian ;
Davis, Kevin ;
Owen, Harry ;
George, Michael W. .
APPLIED SPECTROSCOPY, 2010, 64 (05) :476-484
[13]   Comparison of multivariate methods for quantitative determination with transmission Raman spectroscopy in pharmaceutical formulations [J].
Fransson, Magnus ;
Johansson, Jonas ;
Sparen, Anders ;
Svensson, Olof .
JOURNAL OF CHEMOMETRICS, 2010, 24 (11-12) :674-680
[14]   Characterisation of transmission Raman spectroscopy for rapid quantitative analysis of intact multi-component pharmaceutical capsules [J].
Hargreaves, Michael D. ;
Macleod, Neil A. ;
Smith, Mark R. ;
Andrews, Darren ;
Hammond, Stephen V. ;
Matousek, Pavel .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2011, 54 (03) :463-468
[15]   Application of on-line Raman spectroscopy for characterizing relationships between drug hydration state and tablet physical stability [J].
Hausman, DS ;
Cambron, RT ;
Sakr, A .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2005, 299 (1-2) :19-33
[16]   Analysis of the effect of particle size on polymorphic quantitation by Raman spectroscopy [J].
Hu, Yuerong ;
Wikstrom, Hakan ;
Byrn, Stephen R. ;
Taylor, Lynne S. .
APPLIED SPECTROSCOPY, 2006, 60 (09) :977-984
[17]   Characterization of different laser irradiation methods for quantitative Raman tablet assessment [J].
Johansson, J ;
Pettersson, S ;
Folestad, S .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2005, 39 (3-4) :510-516
[18]   Quantitative transmission Raman spectroscopy of pharmaceutical tablets and capsules [J].
Johansson, Jonas ;
Sparen, Anders ;
Svensson, Olof ;
Folestad, Staffan ;
Claybourn, Mike .
APPLIED SPECTROSCOPY, 2007, 61 (11) :1211-1218
[19]   Direct, non-destructive quantitative measurement of an active pharmaceutical ingredient in an intact capsule formulation using Raman spectroscopy [J].
Kim, Jaejin ;
Noh, Jaegeun ;
Chung, Hoeil ;
Woo, Young-Ah ;
Kemper, Mark S. ;
Lee, Youngil .
ANALYTICA CHIMICA ACTA, 2007, 598 (02) :280-285
[20]   New reliable Raman collection system using the wide area illumination (WAI) scheme combined with the synchronous intensity correction standard for the analysis of pharmaceutical tablets [J].
Kim, Minjung ;
Chung, Hoeil ;
Woo, Youngah ;
Kemper, Mark .
ANALYTICA CHIMICA ACTA, 2006, 579 (02) :209-216