Pulmonary nodules at chest CT: Effect of computer-aided diagnosis on radiologists' detection performance

被引:218
作者
Awai, K
Murao, K
Ozawa, A
Komi, M
Hayakawa, H
Hori, S
Nishimura, Y
机构
[1] Kinki Univ, Sch Med, Dept Radiol, Osaka 5898511, Japan
[2] Computat Sci & Engn Ctr, Fujitsu, Chiba, Japan
[3] Rinku Gen Med Ctr, Dept Radiol, Osaka, Japan
[4] Image Guided Therapy Clin, Osaka, Japan
关键词
computers; diagnostic aid; diagnostic radiology; observer performance; lung; CT; nodule;
D O I
10.1148/radiol.2302030049
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the effect of computer-aided diagnosis (CAD) on radiologists' detection of pulmonary nodules. MATERIALS AND METHODS: Fifty chest computed tomographic (CT) examination cases were used. The mean nodule size was 0.81 cm +/- 0.60 (SD) (range, 0.3-2.9 cm). Alternative free-response receiver operating characteristic (ROC) analysis with a continuous rating scale was used to compare the observers' performance in detecting nodules with and without use of CAD. Five board-certified radiologists and five radiology residents participated in an observer performance study. First they were asked to rate the probability of nodule presence without using CAD; then they were asked to rate the probability of nodule presence by using CAD. RESULTS: For all radiologists, the mean areas under the best-fit alternative free-response ROC curves (A(z)) without and with CAD were 0.64 +/- 0.08 and 0.67 +/- 0.09, respectively, indicating a significant difference (P < .01). For the five board-certified radiologists the mean A(z) values without and with CAD were 0.63 +/- 0.08 and 0.66 +/- 0.09, respectively, indicating a significant difference (P < .01). For the five resident radiologists, the mean A(z) values without and with CAD were 0.66 +/- 0.04 and 0.68 +/- 0.04, respectively, indicating a significant difference (P = .02). At observer performance analyses, there were no significant differences in A(z) values obtained either without (P = .61) or with (P = .88) CAD between the board-certified radiologists and the residents. For all radiologists, in the detection of pulmonary nodules 1.0 cm in diameter or smaller, the mean A(z) values without and with CAD were 0.60 +/- 0.11 and 0.64 +/- 0.11, respectively, indicating a significant difference (P < .01). CONCLUSION: Use of the CAD system improved the board-certified radiologists' and residents detection of pulmonary nodules at chest CT. ((C))RSNA, 2004.
引用
收藏
页码:347 / 352
页数:6
相关论文
共 27 条
[1]  
AKIYAMA M, 1998, HDB IMAGE ANAL, P578
[2]   Computerized detection of pulmonary nodules on CT scans [J].
Armato, SG ;
Giger, ML ;
Moran, CJ ;
Blackburn, JT ;
Doi, K ;
MacMahon, H .
RADIOGRAPHICS, 1999, 19 (05) :1303-1311
[3]  
ASTOLA J, 1994, DIGITAL IMAGE PROCES, P1
[4]   FREE-RESPONSE METHODOLOGY - ALTERNATE ANALYSIS AND A NEW OBSERVER-PERFORMANCE EXPERIMENT [J].
CHAKRABORTY, DP ;
WINTER, LHL .
RADIOLOGY, 1990, 174 (03) :873-881
[5]   DIGITAL AND CONVENTIONAL CHEST IMAGING - A MODIFIED ROC STUDY OF OBSERVER PERFORMANCE USING SIMULATED NODULES [J].
CHAKRABORTY, DP ;
BREATNACH, ES ;
YESTER, MV ;
SOTO, B ;
BARNES, GT ;
FRASER, RG .
RADIOLOGY, 1986, 158 (01) :35-39
[6]   PULMONARY NODULES - IMPROVED DETECTION WITS VASCULAR SEGMENTATION AND EXTRACTION WITH SPIRAL CT [J].
CROISILLE, P ;
SOUTO, M ;
COVA, M ;
WOOD, S ;
AFEWORK, Y ;
KUHLMAN, JE ;
ZERHOUNI, EA .
RADIOLOGY, 1995, 197 (02) :397-401
[7]   Screening for early lung cancer with low-dose spiral CT: Prevalence in 817 asymptomatic smokers [J].
Diederich, S ;
Wormanns, D ;
Semik, M ;
Thomas, M ;
Lenzen, H ;
Roos, N ;
Heindel, W .
RADIOLOGY, 2002, 222 (03) :773-781
[8]   COMPUTERIZED DETECTION OF PULMONARY NODULES IN COMPUTED-TOMOGRAPHY IMAGES [J].
GIGER, ML ;
BAE, KT ;
MACMAHON, H .
INVESTIGATIVE RADIOLOGY, 1994, 29 (04) :459-465
[9]   Early Lung Cancer Action Project: overall design and findings from baseline screening [J].
Henschke, CI ;
McCauley, DI ;
Yankelevitz, DF ;
Naidich, DP ;
McGuinness, G ;
Miettinen, OS ;
Libby, DM ;
Pasmantier, MW ;
Koizumi, J ;
Altorki, NK ;
Smith, JP .
LANCET, 1999, 354 (9173) :99-105
[10]   Computer-aided diagnosis for pulmonary nodules based on helical CT images [J].
Kanazawa, K ;
Kawata, Y ;
Niki, N ;
Satoh, H ;
Ohmatsu, H ;
Kakinuma, R ;
Kaneko, M ;
Moriyama, N ;
Eguchi, K .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1998, 22 (02) :157-167