Cochlear Implant Electrode Tip Fold-Over: Our Experience With Long and Flexible Electrode

被引:11
作者
Hoegerle, Catalina [1 ]
Englhard, Anna [1 ]
Simon, Florian [1 ]
Grueninger, Ivo [1 ]
Mlynski, Robert [2 ]
Hempel, John-Martin [1 ]
Mueller, Joachim [1 ]
机构
[1] Ludwig Maximilians Univ Munchen, Dept Otorhinolaryngol Head & Neck Surg, Marchioninistr 15, D-81377 Munich, Germany
[2] Rostock Univ, Dept Otorhinolaryngol Head & Neck Surg Otto Korne, Med Ctr, Rostock, Germany
关键词
Cochlear implant; Electrode tip fold-over; Lateral wall electrode; Modiolar-hugging electrode; EARLY OUTCOMES; X-RAY; SLIM;
D O I
10.1097/MAO.0000000000003362
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: The aim of this study was to retrospectively investigate if there is any incidence of electrode tip fold-over with 31.5 nun long and flexible lateral wall electrodes implanted in two high-volume Cochlear Implant (CI) centers in Germany. In addition, a detailed literature review was performed to capture all the peer-reviewed publications reporting on tip fold-over with CI electrodes from different CI brands for comparison. Methods: Post-operative X-ray images of FLEX SOFT electrode from MED-EL in Stenver's view were retrospectively investigated for the presence of electrode tip fold-over from 378 consecutive cases in two high-volume CI centers in Germany. All patients were implanted between 2010 and 2018 by three individual experienced CI surgeons using round window and extended round window approach for CI electrode insertion. A literature review was performed following a thorough PubMed (https://www.ncbi.nlm.nih.gov/pubmedo search using the keywords "cochlear implant electrode tip fold-over" or "cochlear implant electrode tip roll-over" to capture articles that were published until December 2020 in English language only. Articles selection was based on electrode-related issues investigated only in-patient cases applying imaging modality. Those studies investigated tip fold-over in cadaveric temporal bones and cases with inner-ear malformation excluded. Results: No single case of tip fold-over was clinically detected from the retrospective investigation of post-operative X-ray images from 378 consecutive cases. The electrode angular insertion depth as measured applying the cochlear coordinate system, varied from a minimum of 560 degrees to a maximum of 720 degrees. The literature review on the tip fold-over issue resulted in 24 peer-reviewed published articles in total. Tip fold-over with pre-curved modiolar-hugging electrodes was reported in 85 cases out of 1,606 implantations making an incidence rate of 5.3%. With the straight lateral wall electrodes, the tip fold-over was reported in four cases out of 398 implantations making an incidence rate of 1%, not including the number of implantations reported in the current study. Otherwise it would be 0.5%. Conclusion: Electrode tip fold-over with 31.5 mm long flexible lateral wall electrodes is highly exceptional and this can be generalized to any of the straight lateral wall electrodes from any CI brand. The literature review on tip fold-over revealed an incidence rate of 5.3% with pre-curved or modiolar-hugging electrodes and 1% with straight lateral wall electrodes from CI brands. Including this series of 0% tip fold-over, the incidence rate of electrode tip fold-over with LW electrode type would be 0.5%.
引用
收藏
页码:64 / 71
页数:8
相关论文
共 32 条
[11]   Surgical Experience and Early Outcomes With a Slim Perimodiolar Electrode [J].
Friedmann, David R. ;
Kamen, Emily ;
Choudhury, Baishakhi ;
Roland, J. Thomas, Jr. .
OTOLOGY & NEUROTOLOGY, 2019, 40 (03) :E304-E310
[12]   Incidence for Tip Foldover During Cochlear Implantation [J].
Gabrielpillai, Jennis ;
Burck, Iris ;
Baumann, Uwe ;
Stoever, Timo ;
Helbig, Silke .
OTOLOGY & NEUROTOLOGY, 2018, 39 (09) :1115-1121
[13]  
Garaycochea O., 2017, BRAZ J OTORHINOLAR, V86, pS38
[14]   Comparative study of two different perimodiolar and a straight cochlear implant electrode array: surgical and audiological outcomes [J].
Garaycochea, Octavio ;
Manrique-Huarte, Raquel ;
Lazaro, Carlos ;
Huarte, Alicia ;
Prieto, Carlos ;
Alvarez de Linera-Alperi, Marta ;
Manrique, Manuel .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2020, 277 (01) :69-76
[15]   A COCHLEAR FREQUENCY-POSITION FUNCTION FOR SEVERAL SPECIES - 29 YEARS LATER [J].
GREENWOOD, DD .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 87 (06) :2592-2605
[16]   Spread of Excitation Measurements for the Detection of Electrode Array Foldovers: A Prospective Study Comparing 3-Dimensional Rotational X-ray and Intraoperative Spread of Excitation Measurements [J].
Grolman, Wilko ;
Maat, Albert ;
Verdam, Froukje ;
Simis, Yvonne ;
Carelsen, Bart ;
Freling, Nicole ;
Tange, Rinze A. .
OTOLOGY & NEUROTOLOGY, 2009, 30 (01) :27-33
[17]   The evaluation of a slim perimodiolar electrode: surgical technique in relation to intracochlear position and cochlear implant outcomes [J].
Heutink, Floris ;
Verbist, Berit M. ;
Mens, Lucas H. M. ;
Huinck, Wendy J. ;
Mylanus, Emmanuel A. M. .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2020, 277 (02) :343-350
[18]   Preservation of residual hearing after cochlear implant surgery with slim modiolar electrode [J].
Iso-Mustajarvi, Matti ;
Sipari, Sini ;
Lopponen, Heikki ;
Dietz, Aarno .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2020, 277 (02) :367-375
[19]   Intraoperative Conebeam CT for Assessment of Intracochlear Positioning of Electrode Arrays in Adult Recipients of Cochlear Implants [J].
Jia, H. ;
Torres, R. ;
Nguyen, Y. ;
De Seta, D. ;
Ferrary, E. ;
Wu, H. ;
Sterkers, O. ;
Bernardeschi, D. ;
Mosnier, I. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2018, 39 (04) :768-774
[20]   Use of intraoperative CT scanning for quality control assessment of cochlear implant electrode array placement [J].
Labadie, Robert F. ;
Schefano, Antonio D. ;
Holder, Jourdan T. ;
Dwyer, Robert T. ;
Rivas, Alejandro ;
O'Malley, Matthew R. ;
Noble, Jack H. ;
Dawant, Benoit M. .
ACTA OTO-LARYNGOLOGICA, 2020, 140 (03) :206-211