Validation of Uniaxial and Triaxial Accelerometers for the Assessment of Physical Activity in Preschool Children

被引:89
作者
Adolph, Anne L. [1 ]
Puyau, Maurice R. [1 ]
Vohra, Firoz A. [1 ]
Nicklas, Theresa A. [1 ]
Zakeri, Issa F. [2 ]
Butte, Nancy F. [1 ]
机构
[1] Baylor Coll Med, Dept Pediat, USDA ARS Childrens Nutr Res Ctr, Houston, TX 77030 USA
[2] Drexel Univ, Dept Epidemiol & Biostat, Philadelphia, PA 19104 USA
关键词
accelerometry; calorimetry; sedentary behavior; light physical activity; moderate physical activity; vigorous physical activity; ACTIVITY RATING-SCALE; ACTIVITY MONITOR; CALIBRATION; CARS;
D O I
10.1123/jpah.9.7.944
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Purpose: Given the unique physical activity (PA) patterns of preschoolers, wearable electronic devices for quantitative assessment of physical activity require validation in this population. Study objective was to validate uniaxial and triaxial accelerometers in preschoolers. Methods: Room calorimetry was performed over 3 hours in 64 preschoolers, wearing Actical, Actiheart, and RT3 accelerometers during play, slow, moderate, and fast translocation. Based on activity energy expenditure (AEE) and accelerometer counts, optimal thresholds for PA levels were determined by piecewise linear regression and discrimination boundary analysis. Results: Established HR cutoffs in preschoolers for sedentary/light, light/moderate and moderate/vigorous levels were used to define AEE (0.015, 0.054, 0.076 kcal.kg(-1).min(-1)) and PA ratio (PAR; 1.6, 2.9, 3.6) thresholds, and accelerometer thresholds. True positive predictive rates were 77%, 75%, and 76% for sedentary; 63%, 61%, and 65% for light; 34%, 52%, and 49% for moderate; 46%, 46%, and 49% for vigorous levels. Due to low positive predictive rates, we combined moderate and vigorous PA. Classification accuracy was improved overall and for the combined moderate-to-vigorous PA level (69%, 82%, 79%) for Actical, Actiheart, and RT3, respectively. Conclusion: Uniaxial and triaxial accelerometers are acceptable devices with similar classification accuracy for sedentary, light, and moderate-to-vigorous levels of PA in preschoolers.
引用
收藏
页码:944 / 953
页数:10
相关论文
共 27 条
[11]   Calibration of accelerometer output for adults [J].
Matthews, CE .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2005, 37 (11) :S512-S522
[12]  
MOON JK, 1995, J NUTR, V125, P220
[13]   Physical activity in preschoolers - Understanding prevalence and measurement issues [J].
Oliver, Melody ;
Schofield, Grant M. ;
Kolt, Gregory S. .
SPORTS MEDICINE, 2007, 37 (12) :1045-1070
[14]   Validation and calibration of an accelerometer in preschool children [J].
Pate, Russell R. ;
Almeida, Maria J. ;
McIver, Kerry L. ;
Pfeiffer, Karin A. ;
Dowda, Marsha .
OBESITY, 2006, 14 (11) :2000-2006
[15]   Validation and calibration of the actical accelerometer in preschool children [J].
Pfeiffer, KA ;
McIver, KL ;
Dowda, M ;
Almeida, MJCA ;
Pate, RR .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2006, 38 (01) :152-157
[16]   Physical activity assessment with accelerometers: An evaluation against doubly labeled water [J].
Plasqui, Guy ;
Westerterp, Klaas R. .
OBESITY, 2007, 15 (10) :2371-2379
[17]   CHILDRENS ACTIVITY RATING-SCALE (CARS) - DESCRIPTION AND CALIBRATION [J].
PUHL, J ;
GREAVES, K ;
HOYT, M ;
BARANOWSKI, T .
RESEARCH QUARTERLY FOR EXERCISE AND SPORT, 1990, 61 (01) :26-36
[18]  
Puyau MR, 2004, MED SCI SPORT EXER, V36, P1625, DOI 10.1249/01.MSS.0000139898.30804.60
[19]   An objective method for measurement of sedentary behavior in 3-to 4-year olds [J].
Reilly, JJ ;
Coyle, J ;
Kelly, L ;
Burke, G ;
Grant, S ;
Paton, JY .
OBESITY RESEARCH, 2003, 11 (10) :1155-1158
[20]  
Schofield W N, 1985, Hum Nutr Clin Nutr, V39 Suppl 1, P5