Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales

被引:409
作者
Smith, Alison J. [1 ]
Dieppe, Paul [2 ,3 ]
Vernon, Kelly [1 ]
Porter, Martyn
Blom, Ashley W. [1 ]
机构
[1] Univ Bristol, Sch Clin Sci, Musculoskeletal Res Unit, Bristol BS10 5NB, Avon, England
[2] Univ Plymouth, Peninsula Med Sch, Plymouth PL4 8AA, Devon, England
[3] Univ Exeter, Peninsula Med Sch, Exeter, Devon, England
关键词
DIFFERENT BEARING SURFACES; ARTHROPLASTY; OSTEOARTHRITIS; DIAMETER; IMPLANTS; SURVIVAL; THA;
D O I
10.1016/S0140-6736(12)60353-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Total hip replacement (THR) is extremely common. Some prostheses fail, particularly in younger patients, and need to be revised, most commonly for loosening secondary to wear or dislocation. Surgeons have tried to address these problems by implanting large diameter metal-on-metal bearing surfaces. Our aim was to assess if metal-on-metal bearing surfaces lead to increased implant survival compared with other bearing surfaces in stemmed THR and, additionally, if larger head sizes result in improved implant survival. Methods We analysed the National Joint Registry of England and Wales for primary hip replacements (402 051, of which 31 171 were stemmed metal-on-metal) undertaken between 2003 and 2011. Our analysis was with a multivariable flexible parametric survival model to estimate the covariate-adjusted cumulative incidence of revision adjusting for the competing risk of death. Findings Metal-on-metal THR failed at high rates. Failure was related to head size, with larger heads failing earlier (3.2% cumulative incidence of revision [95% CI 2.5-4.1] for 28 mm and 5.1% [4.2-6.2] for 52 mm head at 5 years in men aged 60 years). 5 year revision rates in younger women were 6.1% (5.2-7.2) for 46 mm metal-on-metal compared with 1.6% (1.3-2.1) for 28 mm metal-on-polyethylene. By contrast, for ceramic-on-ceramic articulations larger head sizes were associated with improved survival (5 year revision rate of 3.3% [2.6-4.1] with 28 mm and 2.0% [1.5-2.7] with 40 mm for men aged 60 years). Interpretation Metal-on-metal stemmed articulations give poor implant survival compared with other options and should not be implanted. All patients with these bearings should be carefully monitored, particularly young women implanted with large diameter heads. Since large diameter ceramic-on-ceramic bearings seem to do well we support their continued use.
引用
收藏
页码:1199 / 1204
页数:6
相关论文
共 29 条
[21]  
National Joint Registry of England and Wales, 8 ANN REP 2011
[22]   Survivorship of second-generation metal-on-metal primary total hip replacement [J].
Neuerburg, C. ;
Impellizzeri, F. ;
Goldhahn, J. ;
Frey, Ph. ;
Naal, F. D. ;
von Knoch, M. ;
Leunig, M. ;
von Knoch, F. .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2012, 132 (04) :527-533
[23]   Thresholds for indirect DNA damage across cellular barriers for orthopaedic biomaterials [J].
Parry, Michael C. ;
Bhabra, Gevdeep ;
Sood, Aman ;
Machado, Filipa ;
Cartwright, Laura ;
Saunders, Margaret ;
Ingham, E. ;
Newson, R. ;
Blom, Ashley W. ;
Case, Charles P. .
BIOMATERIALS, 2010, 31 (16) :4477-4483
[24]   Flexible parametric proportional-hazards and proportional-odds models for censored survival data, with application to prognostic modelling and estimation of treatment effects [J].
Royston, P ;
Parmar, MKB .
STATISTICS IN MEDICINE, 2002, 21 (15) :2175-2197
[25]   Comparative assessment of implantable hip devices with different bearing surfaces: systematic appraisal of evidence [J].
Sedrakyan, Art ;
Normand, Sharon-Lise T. ;
Dabic, Stefan ;
Jacobs, Samantha ;
Graves, Stephen ;
Marinac-Dabic, Danica .
BMJ-BRITISH MEDICAL JOURNAL, 2011, 343
[26]   Optimal bearing surfaces for total hip replacement in the young patient: a meta-analysis [J].
Shetty, Vijay ;
Shitole, Bhushan ;
Shetty, Gautam ;
Thakur, Harshad ;
Bhandari, Mohit .
INTERNATIONAL ORTHOPAEDICS, 2011, 35 (09) :1281-1287
[27]   The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements [J].
Smith, SL ;
Dowson, D ;
Goldsmith, AAJ .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H2) :161-170
[28]   The failure of survivorship [J].
Wylde, V. ;
Blom, A. W. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :569-570
[29]   Survival of Hard-on-Hard Bearings in Total Hip Arthroplasty: A Systematic Review [J].
Zywiel, Michael G. ;
Sayeed, Siraj A. ;
Johnson, Aaron J. ;
Schmalzried, Thomas P. ;
Mont, Michael A. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (06) :1536-1546