Electrical impedance spectroscopy for detection of bacterial cells in suspensions using interdigitated microelectrodes

被引:129
|
作者
Yang, Liju [1 ,2 ]
机构
[1] N Carolina Cent Univ, Biomfg Res Inst & Technol Enterprise, Durham, NC 27707 USA
[2] N Carolina Cent Univ, Dept Pharmaceut Sci, Durham, NC 27707 USA
关键词
impedance spectroscopy; bacterial detection; bacterial suspension; Salmonella; ESCHERICHIA-COLI O157-H7; IMMUNOMAGNETIC SEPARATION; SALMONELLA-TYPHIMURIUM; RAPID DETECTION; BIOSENSOR; SURFACE; MICROFLUIDICS; MICROBIOLOGY; CONDUCTIVITY; ELECTRODES;
D O I
10.1016/j.talanta.2007.10.018
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study, we present a new, simple and rapid impedance method to detect bacterial cells by making use of the impedance properties of bacterial cell suspensions using interdigitated microelectrodes. It was found that bacterial cell suspensions in deionized (DI) water with different cell concentrations could generate different electrical impedance spectral responses, whereas cell suspensions in phosphate buffered saline (PBS) solution could not produce any significant differences in impedance spectra in response to different cell concentrations. In DI water suspensions, impedance at I kHz decreased with the increasing cell concentrations in the suspensions. The impedance of cell suspensions in DI water was discussed and found that it was resulted from the cell wall charges and the release of ions or other osmolytes from the cells. A linear relationship between the impedance and the logarithmic value of the cell concentration was found in the cell concentration range from 10(6) to 10(10) cfu/ml, which can be expressed by a regression equation of Z (k Omega) = -2.06 log C (cells/20 mu l) + 5.23 with R-2 = 0.98. The detection limit was calculated to be 3.45 x 10(6) cfu/ml, which is comparable with many label-free immunosensors for detection of pathogenic bacteria reported in the literature. To achieve the selectivity of this method, we also demonstrated the feasibility of integrating magnetic separation to this impedance method. This study has demonstrated that bacterial cell concentration can be inferred by measuring the impedance of cell suspensions in DI water. This new detection mechanism could be an alternative to current impedance methods that have been reported for the detection of bacterial cells, e.g. impedance microbiology and electrical/electrochemical impedance biosensors. Published by Elsevier B.V.
引用
收藏
页码:1621 / 1629
页数:9
相关论文
共 50 条
  • [1] Interdigitated array microelectrodes based impedance biosensors for detection of bacterial cells
    Varshney, Madhukar
    Li, Yanbin
    BIOSENSORS & BIOELECTRONICS, 2009, 24 (10): : 2951 - 2960
  • [2] Amyloid Beta Detection by Faradaic Electrochemical Impedance Spectroscopy Using Interdigitated Microelectrodes
    Park, Jin Soo
    Kim, Hye Jin
    Lee, Ji-Hoon
    Park, Jung Ho
    Kim, Jinsik
    Hwang, Kyo Seon
    Lee, Byung Chul
    SENSORS, 2018, 18 (02):
  • [3] Rapid diagnosis of mycobacterium tuberculosis with electrical impedance spectroscopy in suspensions using interdigitated microelectrode
    Pourmir, A. R.
    Bahrmand, A. R.
    Far, S. H. Ettefagh
    Tasbiti, A. R. Hadizadeh
    Yari, Sh.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2016, 71 (07) : 676 - 684
  • [4] Rapid diagnosis of mycobacterium tuberculosis with electrical impedance spectroscopy in suspensions using interdigitated microelectrode
    A. R. Pourmir
    A. R. Bahrmand
    S. H. Ettefagh Far
    A. R. Hadizadeh Tasbiti
    Sh. Yari
    Journal of Analytical Chemistry, 2016, 71 : 676 - 684
  • [5] Electrical Impedance Spectroscopy for Detection of Cells in Suspensions Using Microfluidic Device with Integrated Microneedles
    Mansor, Muhammad Asraf
    Takeuchi, Masaru
    Nakajima, Masahiro
    Hasegawa, Yasuhisa
    Ahmad, Mohd Ridzuan
    APPLIED SCIENCES-BASEL, 2017, 7 (02):
  • [6] Rapid bacterial detection with an interdigitated array electrode by electrochemical impedance spectroscopy
    Kim, Seonghwan
    Yu, Guiduk
    Kim, Taeyoung
    Shin, Kyusoon
    Yoon, Jeyong
    ELECTROCHIMICA ACTA, 2012, 82 : 126 - 131
  • [7] Electrical characterization of phytoplankton suspensions using impedance spectroscopy
    Margaret R. Jett
    Mohamed Z. Rashed
    Susan P. Hendricks
    Stuart J. Williams
    Journal of Applied Phycology, 2021, 33 : 1643 - 1650
  • [8] Electrical characterization of phytoplankton suspensions using impedance spectroscopy
    Jett, Margaret R.
    Rashed, Mohamed Z.
    Hendricks, Susan P.
    Williams, Stuart J.
    JOURNAL OF APPLIED PHYCOLOGY, 2021, 33 (03) : 1643 - 1650
  • [9] The Effect of 3D Interdigitated Microelectrodes on Electrochemical Impedance Spectroscopy
    Chang, Fu-Yu
    Chen, Ming-Kun
    Jang, Ling-Sheng
    Wang, Min-Haw
    2014 INTERNATIONAL CONFERENCE ON INFORMATION SCIENCE, ELECTRONICS AND ELECTRICAL ENGINEERING (ISEEE), VOLS 1-3, 2014, : 676 - +
  • [10] Highly sensitive three-dimensional interdigitated microelectrode for microparticle detection using electrical impedance spectroscopy
    Chang, Fu-Yu
    Chen, Ming-Kun
    Wang, Min-Haw
    Jang, Ling-Sheng
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2016, 49 (07)