Usefulness of biodegradable stents constructed of poly-/-lactic acid monofilaments in patients with benign esophageal stenosis

被引:98
作者
Saito, Yasuharu [1 ]
Tanaka, Toyohiko
Andoh, Akira
Minematsu, Hideki
Hata, Kazunori
Tsujikawa, Tomoyuki
Nitta, Norihisa
Murata, Kiyoshi
Fujiyama, Yoshihide
机构
[1] Shiga Univ Med Sci, Div Endoscop Med, Otsu, Shiga 5202192, Japan
[2] Shiga Univ Med Sci, Dept Radiol, Otsu, Shiga 5202192, Japan
[3] Shiga Univ Med Sci, Dept Internal Med, Otsu, Shiga 5202192, Japan
关键词
stents; esophageal cancer; endoscopic; submucosal dissection;
D O I
10.3748/wjg.v13.i29.3977
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
AIM: To report 13 patients with benign esophageal stenosis treated with the biodegradable stent. METHODS: We developed a Ultraflex-type stent by knitting poly-/-lactic acid monofilaments. RESULTS: Two cases were esophageal stenosis caused by drinking of caustic liquid, 4 cases were due to surgical resection of esophageal cancers, and 7 cases were patients with esophageal cancer who received the preventive placement of biodegradable stents for post-endoscopic mucosal dissection (ESD) stenosis. The preventive placement was performed within 2 to 3 d after ESD. In 10 of the 13 cases, spontaneous migration of the stents occurred between 10 to 21 d after placement. In these cases, the migrated stents were excreted with the feces, and no obstructive complications were experienced. In 3 cases, the stents remained at the proper location on d 21 after placement. No symptoms of re-stenosis were observed within the follow-up period of 7 mo to 2 years. Further treatment with balloon dilatation or replacement of the biodegradable stent was not required. CONCLUSION: Biodegradable stents were useful for the treatment of benign esophageal stenosis, particularly for the prevention of post-ESD stenosis. (c) 2007 WJG. All rights reserved.
引用
收藏
页码:3977 / 3980
页数:4
相关论文
共 21 条
[1]   Expandable metallic stents should not be used in the treatment of benign esophageal strictures [J].
Ackroyd, R ;
Watson, DI ;
Devitt, PG ;
Jamieson, GG .
JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2001, 16 (04) :484-487
[2]   ABSORBABLE IMPLANTS FOR THE FIXATION OF FRACTURES [J].
BOSTMAN, OM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (01) :148-153
[3]   Self-expanding plastic stents for benign esophageal lesions [J].
Evrard, S ;
Le Moine, O ;
Lazaraki, G ;
Dormann, A ;
El Nakadi, I ;
Devière, J .
GASTROINTESTINAL ENDOSCOPY, 2004, 60 (06) :894-900
[4]   Palliation of patients with malignant gastroduodenal obstruction with self-expanding metallic stents: the treatment of choice? [J].
Holt, AP ;
Patel, M ;
Ahmed, MM .
GASTROINTESTINAL ENDOSCOPY, 2004, 60 (06) :1010-1017
[5]   The use of biodegradable PGA stents to judge the risk of post-TURP incontinence in patients with combined bladder outlet obstruction and overactive bladder [J].
Knutson, T ;
Pettersson, S ;
Dahlstrand, C .
EUROPEAN UROLOGY, 2002, 42 (03) :262-267
[6]  
KULKARNI RK, 1966, ARCH SURG-CHICAGO, V93, P839
[7]   PRELIMINARY EXPERIENCE WITH ENDOSCOPIC INTRALESIONAL STEROID INJECTION THERAPY FOR REFRACTORY UPPER GASTROINTESTINAL STRICTURES [J].
LEE, M ;
KUBIK, CM ;
POLHAMUS, CD ;
BRADY, CE ;
KADAKIA, SC .
GASTROINTESTINAL ENDOSCOPY, 1995, 41 (06) :598-601
[8]  
Oyama Tsuneo, 2005, Clin Gastroenterol Hepatol, V3, pS67, DOI 10.1016/S1542-3565(05)00291-0
[9]   Biodegradable external tracheal stents and their use in a rabbit tracheal reconstruction model [J].
Robey, TC ;
Eiselt, PM ;
Murphy, HS ;
Mooney, DJ ;
Weatherly, RA .
LARYNGOSCOPE, 2000, 110 (11) :1936-1942
[10]   New tubular bioabsorbable knitted airway stent: Feasibility assessment for delivery and deployment in a dog model [J].
Saito, Y ;
Minami, KI ;
Kaneda, H ;
Okada, T ;
Maniwa, T ;
Araki, Y ;
Imamura, H ;
Yamada, H ;
Igaki, K ;
Tamai, H .
ANNALS OF THORACIC SURGERY, 2004, 78 (04) :1438-1440