Reproducibility and reliability of hypoglycaemic episodes recorded with Continuous Glucose Monitoring System (CGMS) in daily life

被引:32
作者
Hoi-Hansen, T [1 ]
Pedersen-Bjergaard, U [1 ]
Thorsteinsson, B [1 ]
机构
[1] Hillerod Hosp, Dept Internal Med F, DK-3400 Hillerod, Denmark
关键词
continuous glucose monitoring system; hypoglycaemia; self-monitored blood glucose; Type; 1; diabetes;
D O I
10.1111/j.1464-5491.2005.01552.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim Continuous glucose monitoring may reveal episodes of unrecognized hypoglycaemia. We evaluated reproducibility and reliability of hypoglycaemic episodes recorded in daily life by the Medtronic MiniMed Continuous Glucose Monitoring System (CGMS). Methods Twenty-nine adult patients with Type 1 diabetes underwent 6 days of continuous subcutaneous glucose monitoring, applying one CGMS on each side of the abdomen. Blood glucose was measured by HemoCue B-Glucose Analyzers six times daily and two different 4-point calibration sets were generated (set A and B). Using these calibration sets, CGMS raw data were recalibrated generating four different CGMS data sets [left-A (left side of abdomen, calibration set A), left-B, right-A and right-B]. Agreement between CGMS data sets was evaluated during hypoglycaemic events, comparing CGMS readings = 2.2 mmol/l with nadir values from corresponding CGMS data sets. CGMS readings were also compared with independent self-monitored blood glucose (SMBG) values. Results With hypoglycaemia (CGMS readings = 2.2 mmol/l) in calibration set left-A, values below 3.5 mmol/l were present in 99% (95% CI: 95-100%) of samples in left-B, 91% (95% CI: 84-96%) of samples in right-A, and 90% (95% CI: 83-95%) of samples in right B. In 84% of these episodes (95% CI: 59-96%) independent SMBG values were below 3.5 mmol/l. Difference in duration was observed with a median difference of 20 min; (left-A vs. right-B). Conclusion Hypoglycaemic episodes recorded by CGMS are reproducible and agreement with independent SMBG values is acceptable for retrospective recording of hypoglycaemic events with CGMS.
引用
收藏
页码:858 / 862
页数:5
相关论文
共 23 条
  • [1] Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor
    Boyne, MS
    Silver, DM
    Kaplan, J
    Saudek, CD
    [J]. DIABETES, 2003, 52 (11) : 2790 - 2794
  • [2] Subcutaneous glucose sensor values closely parallel blood glucose during insulin-induced hypoglycaemia
    Caplin, NJ
    O'Leary, P
    Bulsara, M
    Davis, EA
    Jones, TW
    [J]. DIABETIC MEDICINE, 2003, 20 (03) : 238 - 241
  • [3] Cheyne E H, 2002, Diabetes Technol Ther, V4, P607, DOI 10.1089/152091502320798222
  • [4] Hypoglycaemia: The limiting factor in the glycaemic management of Type I and Type II Diabetes
    Cryer, PE
    [J]. DIABETOLOGIA, 2002, 45 (07) : 937 - 948
  • [5] Accuracy of the continuous glucose monitoring system in inpatient and outpatient conditions
    Djakouré-Platonoff, C
    Radermercker, R
    Reach, G
    Slama, G
    Selam, JL
    [J]. DIABETES & METABOLISM, 2003, 29 (02) : 159 - 162
  • [6] Delay in onset of awareness of acute hypoglycemia and of restoration of cognitive performance during recovery
    Evans, ML
    Pernet, A
    Lomas, J
    Jones, J
    Amiel, SA
    [J]. DIABETES CARE, 2000, 23 (07) : 893 - 897
  • [7] FREQUENCY OF SEVERE HYPOGLYCEMIA IN PATIENTS WITH TYPE-1 DIABETES WITH IMPAIRED AWARENESS OF HYPOGLYCEMIA
    GOLD, AE
    MACLEOD, KM
    FRIER, BM
    [J]. DIABETES CARE, 1994, 17 (07) : 697 - 703
  • [8] Gross T, 2003, DIABETES, V52, pA95
  • [9] Gross TM, 2000, DIABETOLOGIA, V43, pA5
  • [10] Clinical performance of CGMS in type 1 diabetic patients treated by continuous subcutaneous insulin infusion using insulin analogs
    Guerci, B
    Floriot, M
    Böhme, P
    Durain, D
    Benichou, M
    Jellimann, S
    Drouin, P
    [J]. DIABETES CARE, 2003, 26 (03) : 582 - 589