A monoclonal immunoglobulin M not visible on the profile of serum capillary electrophoresis

被引:0
作者
Szymanowicz, A. [1 ]
Doche, C. [2 ]
Schemitick, I. [3 ]
机构
[1] Ctr Hosp Roanne, Biochim Lab, Roanne, France
[2] Ctr Hosp Chambery, Biochim Lab, Chambery, France
[3] Lab Savoy Schemitick, Albertville, MN USA
关键词
capillary electrophoresis; monoclonal IgM; gammopathy; anomaly of migration; recommendation;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
We report a case of an invisible monoclonal IgM on the profile of the serum electrophoresis carried Out by the Paragon CZE 2000 (R) system. This case is about a 84 years old man followed for a Waldenstrom disease. The level of the monoclonal IgM given at the time of the preceding visit one year before was 10 g/L. The electrophoresis is reproduced in two other laboratories, using respectively the Hydrasys (R) and Capillarys (R) techniques which have both shown the presence of the monoclonal peak. At this occasion the frequency of these anomalies of migration and the recommendations to avoid the erroneous interpretation of the electrophoresis are discussed. The vigilance and the competence of the biologist are essential to avoid this rare pitfall, induced by the new principle of the capillary electrophoresis.
引用
收藏
页码:465 / 469
页数:5
相关论文
共 12 条
[1]  
Blade Joan, 1998, British Journal of Haematology, V102, P1115, DOI 10.1046/j.1365-2141.1998.00930.x
[2]  
Blessum C, 1999, ANN BIOL CLIN-PARIS, V57, P643
[3]  
Bossuyt X, 2001, CLIN CHEM, V47, P1477
[4]   Detection of monoclonal proteins by capillary zone electrophoresis:: Comparison of 2 multichannel automated systems [J].
Bossuyt, Xavier ;
Marien, Godelieve .
CLINICAL CHEMISTRY, 2007, 53 (01) :152-153
[5]  
BRADWELL AR, 2006, SERUM FREE LIGNT CHA
[6]  
Henskens Y, 1998, CLIN CHEM, V44, P1184
[7]  
Jenkins MA, 1996, CLIN CHEM, V42, P1886
[8]  
Katzmann JA, 1998, AM J CLIN PATHOL, V110, P503
[9]  
Le Carrer D, 2005, SPECTRA BIOL, V146, P47
[10]   Clinical capillary zone electrophoresis of serum proteins:: Balancing high sensitivity and high specificity [J].
Mariën, G ;
Vranken, G ;
Demuylder, M ;
Blanckaert, N ;
Bossuyt, X .
CLINICAL CHEMISTRY, 2003, 49 (08) :1419-1420