Magnetic resonance imaging and gynecological devices

被引:11
作者
Correia, Lucia [1 ]
Ramos, Ana Beatriz [1 ]
Machado, Ana Isabel [1 ]
Rosa, Duarte [1 ]
Marques, Carlos [1 ]
机构
[1] Hosp Geral de Santo Antonio, Ctr Hosp Porto, Oporto, Portugal
关键词
Magnetic resonance imaging; Intrauterine devices; Implanon (R); Essure (R); INTRAUTERINE CONTRACEPTIVE DEVICES; METALLIC IMPLANTS; SAFETY; IUD; MRI;
D O I
10.1016/j.contraception.2011.10.011
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Background: Performing magnetic resonance imaging (MRI) on women with gynecological devices is a completely accepted practice. The goal of our review is to assess how safe it is to perform MRI on women using contraceptive implants or devices. Study Design: Literature review, searching in PubMed-Medline/Ovid for the following keywords: magnetic resonance imaging, intrauterine devices, Implanon (R) and Essure (R). Results: Though plastic devices do not represent a contraindication to the use of the technique, those including metallic components have been submitted to several tests, after which they were classified as MR Conditional (devices presenting no risks in MR-specific environments) by the Food and Drug Administration. Thus, the use of MRI can be safely advised to women with this type of device as long as the magnetic resonance equipment is <= 3.0 T. Conclusions: Presently, there is no scientific evidence that contraindicates performing MRI on women with any kind of gynecological device. Therefore, this procedure is safe as long as it is performed under previously tested conditions. (C) 2012 Elsevier Inc. All rights reserved.
引用
收藏
页码:538 / 543
页数:6
相关论文
共 18 条
[1]   Safety of intrauterine contraceptive devices during MR imaging [J].
Hess, T ;
Stepanow, B ;
Knopp, MV .
EUROPEAN RADIOLOGY, 1996, 6 (01) :66-68
[2]   MR IMAGING IN PATIENTS WITH METALLIC IMPLANTS [J].
LAAKMAN, RW ;
KAUFMAN, B ;
HAN, JS ;
NELSON, AD ;
CLAMPITT, M ;
OBLOCK, AM ;
HAAGA, JR ;
ALFIDI, RJ .
RADIOLOGY, 1985, 157 (03) :711-714
[3]  
Lee JKT, 2006, COMPUTED BODY TOMOGR, V20, P1341
[4]   INTRAUTERINE CONTRACEPTIVE DEVICES - MR IMAGING [J].
MARK, AS ;
HRICAK, H .
RADIOLOGY, 1987, 162 (02) :311-314
[5]  
Nunes MCF, 2009, CIENCIA RURAL, V39, P1287
[6]  
Palácio Glaucia Andrade e Silva, 2002, Radiol Bras, V35, P371, DOI 10.1590/S0100-39842002000600011
[7]   Lack of interaction between magnetic resonance imaging and the Copper-T380A IUD [J].
Pasquale, SA ;
Russer, TJ ;
Foldesy, R ;
Mezrich, RS .
CONTRACEPTION, 1997, 55 (03) :169-173
[8]   AAPM/RSNA physics tutorial for residents - Fundamental physics of MR imaging [J].
Pooley, RA .
RADIOGRAPHICS, 2005, 25 (04) :1087-1099
[9]   New metallic implant used for permanent contraception in women: Evaluation of MR safety [J].
Shellock, FG .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 178 (06) :1513-1516
[10]   Biomedical implants and devices: Assessment of magnetic field interactions with a 3.0-Tesla MR system [J].
Shellock, FG .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 16 (06) :721-732