BIOCOMPATIBILITY AND MECHANICAL PROPERTY OF LARS ARTIFICIAL LIGAMENT WITH TISSUE INGROWTH

被引:10
作者
Wang, Cho-Lin [1 ]
Hsiao, Chih-Kun [3 ]
Hsu, Ar-Tyan [2 ]
Dung, Ching-Ze [1 ]
Chang, Chih-Han [1 ]
机构
[1] Natl Cheng Kung Univ, Inst Biomed Engn, Tainan 701, Taiwan
[2] Natl Cheng Kung Univ, Dept Phys Therapy, Tainan 701, Taiwan
[3] E Da Hosp, Dept Med Res, Kauhsiung 824, Taiwan
关键词
LARS artificial ligament; anterior cruciate ligament; tissue ingrowth; mechanical property; ANTERIOR CRUCIATE LIGAMENT; NEW-GENERATION; RECONSTRUCTION; MATRICES; KNEE;
D O I
10.1142/S0219519411004514
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Tissue ingrowth into the implanted artificial ligaments is to be expected after reconstructive procedures and several groups have reported soft tissue ingrowth of the implanted LARS (Ligament Augmentation and Reconstruction System, Dijon, France) artificial ligament. Up to now the influence of the tissue ingrowth on the mechanical properties of the artificial ligament is uncertain. The purpose of this research was to study the mechanical property after tissue ingrowth on LARS artificial ligament. Five LARS ligaments were implanted subcutaneously in the abdomen of five pigs. After six months of implantation, four successful implants were explanted and tested. The results showed that fibroblasts and collagen fibers had grown into the unknotted middle part of the LARS. However, 12.2 +/- 2.2% of fibres were surrounded by foreign body giant cells. Tensile tests showed that the explanted LARS possessed similar elastic characteristics as native ligaments. The tensile strength of the explanted LARS decreased by 23.53 +/- 18.04% (p < 0.001) and the elongation of the middle part of the explanted LARS increased by 69.84 +/- 38.38% (p < 0.002) in comparison with unimplanted control ligaments. Since there were not apparent surface cracking on the LARS fibers, it was inferred that activities of foreign-body giant cells might have weakened the strength and adversely affected the mechanical properties of the fibers of the LARS.
引用
收藏
页数:13
相关论文
共 21 条
[1]   LIGAMENTOUS RESTRAINTS TO ANTERIOR-POSTERIOR DRAWER IN THE HUMAN KNEE - BIOMECHANICAL STUDY [J].
BUTLER, DL ;
NOYES, FR ;
GROOD, ES .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (02) :259-270
[2]   BIOLOGICAL RESPONSE TO LIGAMENT WEAR PARTICLES [J].
CLAES, LE ;
LUDWIG, J ;
MARGEVICIUS, KJ ;
DURSELEN, L .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :35-41
[3]  
Freeman Joseph W., 2008, Recent Patents on Biomedical Engineering, V1, P18, DOI 10.2174/1874764710801010018
[4]   Analysis of retrieved polymer fiber based replacements for the ACL [J].
Guidoin, MF ;
Marois, Y ;
Bejui, J ;
Poddevin, N ;
King, MW ;
Guidoin, R .
BIOMATERIALS, 2000, 21 (23) :2461-2474
[5]  
Hsu HC, 2007, MID TAIWAN J MED, V12, P191
[6]   BIOCOMPATIBILITY AND INGROWTH OF TREVIRA PROSTHESES FOLLOWING REPLACEMENT OF THE CRUCIATE LIGAMENTS [J].
KOCK, HJ ;
STURMER, KM .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1992, 30 (04) :CE13-CE16
[7]   Patient satisfaction needs as related to knee stability and objective findings after ACL reconstruction using the LARS artificial ligament [J].
Lavoie, P ;
Fletcher, J ;
Duval, N .
KNEE, 2000, 7 (03) :157-163
[8]  
Liu ZT, 2009, INT ORTHOP
[9]   A prospective randomized study of anterior cruciate ligament reconstruction - A comparison of patellar tendon and quadruple-strand semitendinosus/gracilis tendons fixed with bioabsorbable interference screws [J].
Maletis, Gregory B. ;
Cameron, Sheri L. ;
Tengan, Joann J. ;
Burchette, Raoul J. .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2007, 35 (03) :384-394
[10]   INVIVO BIOCOMPATIBILITY STUDIES .5. INVIVO LEUKOCYTE INTERACTIONS WITH BIOMER [J].
MARCHANT, RE ;
MILLER, KM ;
ANDERSON, JM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (09) :1169-1190