Effect of tissue density on PET spatial resolution in magnetic environment

被引:1
|
作者
Li, Chong [1 ,2 ]
Cao, Xingzhong [1 ,2 ]
Liu, Fuyan [1 ,2 ]
Zhang, Peng [1 ]
Zhang, Zhiming [1 ,2 ]
Wei, Long [1 ,2 ]
Wang, Baoyi [1 ,2 ]
机构
[1] Chinese Acad Sci, Inst High Energy Phys, Beijing 100049, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100039, Peoples R China
来源
EUROPEAN PHYSICAL JOURNAL PLUS | 2020年 / 135卷 / 01期
关键词
MONTE-CARLO-SIMULATION; WHOLE-BODY PET/MR; POSITRONS;
D O I
10.1140/epjp/s13360-020-00198-x
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Simultaneous positron emission tomography (PET) and magnetic resonance imaging (MRI) can provide a strong magnetic field to improve the PET spatial resolution by restricting the positron range. In this study, an experimental system and a Geant4 model were built to investigate the difference in improved spatial resolution by using a magnetic field for PET in different human tissues. Experiment results indicated that the magnetic field enabled a greater compressive effect on the positron distribution in a low-density medium. From the simulation results, the magnetic compression was only effective in the direction perpendicular to the magnetic field. Additionally, the saturated relative change rates of the mean positron range were reached for all tissues with increasing magnetic field, and the saturated values exhibited an overall negative correlation with the tissue density. However, despite the considerable difference in the absolute range in different tissues, the saturated values were only in a very narrow range of approximately 45% to 50%, compared to the final relative change rates. It was revealed that the magnetic field can be a more effective method for improving the PET imaging quality for low-density tissues, and there should be an upper threshold for improving the PET spatial resolution using a magnetic field. A range-corrected algorithm associated with tissue density for PET-MRI systems can probably be developed based on the results.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] The effect of photon incident angle on spatial resolution with thick crystal hybrid PET.
    Glick, SJ
    Stodilka, RZ
    JOURNAL OF NUCLEAR MEDICINE, 2000, 41 (05) : 56P - 56P
  • [22] Measuring the effect of count rate on PET spatial resolution using a uniform cylinder phantom
    Lodge, Martin
    Rahmim, Arman
    Wong, Dean
    Wahl, R.
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [23] Effect of smart mA CT attenuation correction on PET/CT scanner spatial resolution
    Belakhlef, Sam
    Church, Clifford
    Fortenbery, Edwin
    Lakhanpal, Suresh
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [24] Effect of electrode density and measurement noise on the spatial resolution of cortical potential distribution
    Ryynänen, ORM
    Hyttinen, JAK
    Laarne, PH
    Malmivuo, JA
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (09) : 1547 - 1554
  • [25] Effect of spatial, temporal, and velocity resolution on Doppler tissue image derived indices
    Rojo-Alvarez, JL
    Bermejo, J
    Rodrígez-González, AB
    Yotti, R
    Martínez-Fernández, MA
    García-Fernández, MA
    Antoranz, JC
    Computers in Cardiology 2004, Vol 31, 2004, 31 : 317 - 320
  • [26] Effect of magnetic susceptibility variations on the spatial resolution of tumor mapping by proton magnetic resonance spectroscopy
    Rajagopalan, B
    MEDICAL PHYSICS, 2003, 30 (06) : 1505 - 1506
  • [27] Effect of magnetic nanoparticle internalization on cell density in skeletal muscle tissue
    Kirihara, Yuji
    Miyata, Hiromu
    Peng, Zugui
    Shimba, Kenta
    Miyamoto, Yoshitaka
    Shimizu, Kazunori
    Honda, Hiroyuki
    Yagi, Tohru
    IEEJ Transactions on Electronics, Information and Systems, 2021, 141 (07): : 795 - 801
  • [28] An RPC-PET prototype with high spatial resolution
    Blanco, A
    Carolino, N
    Correia, CMBA
    Marques, RF
    Fonte, P
    González-Díaz, D
    Lindote, A
    Lopes, MI
    Macedo, MP
    Policarpo, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 533 (1-2): : 139 - 143
  • [29] Spatial resolution of human RPC-PET system
    Couceiro, M.
    Crespo, P.
    Mendes, L.
    Ferreira, N.
    Ferreira Marques, R.
    Fonte, P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 661 : S156 - S158
  • [30] A Modified Spatial Resolution Formula for DOI-PET
    Xia, Yan
    Ma, Tianyu
    Liu, Yaqiang
    Wang, Shi
    Shao, Yiping
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2632 - 2635