Performance evaluation of a high-sensitivity large-aperture small-animal PET scanner: ClairvivoPET

被引:38
作者
Mizuta, Tetsuro [1 ]
Kitamura, Keishi [1 ,2 ]
Iwata, Hiroshi [2 ]
Yamagishi, Yoshiyuki [1 ]
Ohtani, Atsushi [1 ]
Tanaka, Kazumi [1 ]
Inoue, Yoshihiro [1 ]
机构
[1] Shimadzu Co Ltd, Med Syst Div, Dept Res & Dev, Nakagyo Ku, Kyoto 6048511, Japan
[2] Shimadzu Co Ltd, Corp Strategy Planning Dept, Kyoto 6048511, Japan
关键词
small-animal PET; performance evaluation; time activity curves; molecular imaging; kinetic measurement;
D O I
10.1007/s12149-008-0127-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective In this study, we evaluated the performance of a newly commercialized small-animal positron emission tomography (PET) scanner, ClairvivoPET, which provides significant advantages in spatial resolution, sensitivity, and quantitative accuracy. Methods This scanner consists of depth of interaction detector modules with a large axial extent of 151 mm and an external Cs-137 source for attenuation correction. Physical performances, resolution, sensitivity, scatter fraction (SF), counting rate including noise equivalent count (NEC) rate, quantitative accuracy versus activity strength, and transmission accuracy, were measured and evaluated. Animal studies were also performed. Results Transaxial spatial resolution, measured with a capillary tube, was 1.54 mm at the center and 2.93 mm at a radial offset of 40 mm. The absolute sensitivity was 8.2% at the center, and SFs for mouse-and rat-sized phantoms were 10.7% and 24.2%, respectively. Peak NEC rates for mouse-and rat-sized uniform cylindrical phantoms were 328 kcps at 173 kBq/ml and 119 kcps at 49 kBq/ml, respectively. The quantitative stability of emission counts against activity strength was within 2% over 5 half-lives, ranging from 0.6 MBq to 30 MBq. Transmission measurement based on segmented attenuation correction allowed 6-min and 10-min scans for mouse-and rat-sized cylindrical phantoms, respectively. Rat imaging injected with F-18-NaF resulted in visibility of fine bone structures, and mouse imaging injected with F-18-D-fluoromethyl tyrosine demonstrated the feasibility of using this system to obtain simultaneous time activity curves from separate regions, such as for the heart and tumors. Conclusions ClairvivoPET is well suited to quantitative imaging even with short scan times, and will provide a number of advantages in new drug development and for kinetic measurement in molecular imaging.
引用
收藏
页码:447 / 455
页数:9
相关论文
共 50 条
[21]   PET performance evaluation of MADPET4: a small animal PET insert for a 7 T MRI scanner [J].
Omidvari, Negar ;
Cabello, Jorge ;
Topping, Geoffrey ;
Schneider, Florian R. ;
Paul, Stephan ;
Schwaiger, Markus ;
Ziegler, Sibylle I. .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (22) :8671-8692
[22]   Performance evaluation of the fully engineered YAP-(S)PET scanner for small animal imaging [J].
Del Guerra, A ;
Bartoli, A ;
Belcari, N ;
Herbert, D ;
Motta, A ;
Vaiano, A ;
Di Domenico, G ;
Sabba, N ;
Moretti, E ;
Zavattini, G ;
Lazzarotti, M ;
Sensi, L ;
Larobina, M ;
Uccelli, L .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) :1078-1083
[23]   Micro Insert: A Prototype Full-Ring PET Device for Improving the Image Resolution of a Small-Animal PET Scanner [J].
Wu, Heyu ;
Pal, Debashish ;
Song, Tae Yong ;
O'Sullivan, Joseph A. ;
Tai, Yuan-Chuan .
JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (10) :1668-1676
[24]   PET performance evaluation of the small-animal Hyperion IID PET/MRI insert based on the NEMA NU-4 standard [J].
Hallen, Patrick ;
Schug, David ;
Weissler, Bjoern ;
Gebhardt, Pierre ;
Salomon, Andre ;
Kiessling, Fabian ;
Schulz, Volkmar .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (06)
[25]   Evaluation of Scatter Fraction and Count Rate Performance of Two Small-animal PET scanners using dedicated phantoms [J].
Prasad, Rameshwar ;
Zaidi, Habib .
2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, :3639-3645
[26]   Performance Evaluation of microPET: A High-Resolution Lutetium Oxyorthosilicate PET Scanner for Animal Imaging [J].
Chatziioannou, Anion F. ;
Cherry, Simon R. ;
Shao, Yiping ;
Silverman, Robert W. ;
Meadors, Ken ;
Farquhar, Thomas H. ;
Pedarsani, Marjan ;
Phelps, Michael E. .
JOURNAL OF NUCLEAR MEDICINE, 2020, 61 :155S-165S
[27]   Performance evaluation of microPET: A high-resolution lutetium oxyorthosilicate PET scanner for animal imaging [J].
Chatziioannou, AF ;
Cherry, SR ;
Shao, YP ;
Silverman, RW ;
Meadors, K ;
Farquhar, TH ;
Pedarsani, M ;
Phelps, ME .
JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (07) :1164-1175
[28]   Radiosynthesis of [124I]Iodometomidate and Biological Evaluation Using Small-Animal PET [J].
Herbert Kvaternik ;
Thomas Wanek ;
Friedrich Hammerschmidt ;
Ilse Zolle ;
Reingard Aigner ;
Claudia Kuntner .
Molecular Imaging and Biology, 2014, 16 :317-321
[29]   Radiosynthesis of [124I]Iodometomidate and Biological Evaluation Using Small-Animal PET [J].
Kvaternik, Herbert ;
Wanek, Thomas ;
Hammerschmidt, Friedrich ;
Zolle, Ilse ;
Aigner, Reingard ;
Kuntner, Claudia .
MOLECULAR IMAGING AND BIOLOGY, 2014, 16 (03) :317-321
[30]   Performance analysis of a low-cost small animal PET/SPECT scanner [J].
Guerra, Pedro ;
Rubio, Jose L. ;
Ortuno, Juan E. ;
Kontaxakis, Georgios ;
Ledesma, Maria J. ;
Santos, Andres .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) :98-101