Optimization of image quality and accuracy of low iodine concentration quantification as function of kVp pairs for abdominal imaging using dual-source CT: A phantom study

被引:12
作者
Dabli, Djamel [1 ]
Frandon, Julien [1 ]
Hamard, Aymeric [1 ]
Belaouni, Asmaa [1 ]
Addala, Takieddine [1 ]
Beregi, Jean-Paul [1 ]
Greffier, Joel [1 ]
机构
[1] Univ Montpellier, Dept Med Imaging, CHU Nimes, Med Imaging Grp Nimes,EA 2992, Nimes, France
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2021年 / 88卷
关键词
Dual-Source CT scan; Iodine quantification; Virtual monochromatic images; Contrast-to-noise ratio; ENERGY CT; LIVER-LESIONS; DIAGNOSIS; CELL;
D O I
10.1016/j.ejmp.2021.07.008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To determine the suitable kVp pair for optimal image quality of the virtual monochromatic images (VMIs) and iodine quantification accuracy at low concentration, using a third generation dual-source CT (DSCT). Materials and methods: Multi-energy CT phantoms with and without body rings were scanned with a DSCT using four kVp pairs (tube "A"/"B" voltage): 100/Sn150, 90/Sn150, 80/Sn150 and 70/Sn150 kVp. The reference mAs was adjusted to obtain a CTDIvol close to 11 mGy. HU values accuracy (RMSDHU), noise (SD) and contrast-tonoise ratio (CNR) of iodine inserts of 0.5, 1, 2 and 5 mg/mL concentrations were assessed on VMIs at 40/50/ 60/70 keV. Iodine quantification accuracy was assessed using the RMSDiodine and iodine bias (IBiodine). Results: The RMSDHU decreased when the tube "A" voltage increased. The mean noise value increased significantly with tube "A" voltage (p < 0.001) but decreased between 80/Sn150 and 90/Sn150 kVp for the small phantom (1.1 +/- 0.1%; p = 0.047). The CNR significantly decreased with tube "A" voltage (p < 0.001), except between 80/Sn150 and 90/Sn150 kVp for all inserts and between 90/Sn150 kVp and 100/Sn150 kVp for the 1.0 and 0.5 mg/mL inserts in the large phantom. In the small phantom, no significant difference was found between 80/Sn150 kVp and 90/Sn150 kVp for all inserts and between 80/Sn150, 90/Sn150 and 100/Sn150 kVp for the 1 and 0.5 mg/mL inserts. The RMSDiodine and IBiodine decreased as the tube "A" voltage of the kVp pair increased. Conclusion: The kVp pair of 70/Sn150 led to better image quality in VMIs and sufficient iodine accuracy.
引用
收藏
页码:285 / 292
页数:8
相关论文
共 21 条
[1]   Review of Clinical Applications for Virtual Monoenergetic Dual-Energy CT [J].
Albrecht, Moritz H. ;
Vogl, Thomas J. ;
Martin, Simon S. ;
Nance, John W. ;
Duguay, Taylor M. ;
Wichmann, Julian L. ;
De Cecco, Carlo N. ;
Varga-Szemes, Akos ;
van Assen, Marly ;
Tesche, Christian ;
Schoepf, U. Joseph .
RADIOLOGY, 2019, 293 (02) :260-271
[2]   Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data [J].
Greffier, J. ;
Si-Mohamed, S. ;
Dabli, D. ;
de Forges, H. ;
Hamard, A. ;
Douek, P. ;
Beregi, J. P. ;
Frandon, J. .
EUROPEAN RADIOLOGY, 2021, 31 (07) :5324-5334
[3]   Low kV versus dual-energy virtual monoenergetic CT imaging for proven liver lesions: what are the advantages and trade-offs in conspicuity and image quality? A pilot study [J].
Hanson, G. Jay ;
Michalak, Gregory J. ;
Childs, Robert ;
McCollough, Brian ;
Kurup, Anil N. ;
Hough, David M. ;
Frye, Judson M. ;
Fidler, Jeff L. ;
Venkatesh, Sudhakar K. ;
Leng, Shuai ;
Yu, Lifeng ;
Halaweish, Ahmed F. ;
Harmsen, W. Scott ;
McCollough, Cynthia H. ;
Fletcher, J. G. .
ABDOMINAL RADIOLOGY, 2018, 43 (06) :1404-1412
[4]   Dual-Energy CT Applications in the Abdomen [J].
Heye, Tobias ;
Nelson, Rendon C. ;
Ho, Lisa M. ;
Marin, Daniele ;
Boll, Daniel T. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (05) :S64-S70
[5]   Intermanufacturer Comparison of Dual-Energy CT Iodine Quantification and Monochromatic Attenuation: A Phantom Study [J].
Jacobsen, Megan C. ;
Schellingerhout, Dawid ;
Wood, Cayla A. ;
Tamm, Eric P. ;
Godoy, Myrna C. ;
Sun, Jia ;
Cody, Dianna D. .
RADIOLOGY, 2018, 287 (01) :224-234
[6]   Clinical significance of dual-energy CT-derived iodine quantification in the diagnosis of metastatic LN in colorectal cancer [J].
Kato, T. ;
Uehara, K. ;
Ishigaki, S. ;
Nihashi, T. ;
Arimoto, A. ;
Nakamura, H. ;
Kamiya, T. ;
Oshiro, T. ;
Ebata, T. ;
Nagino, M. .
EJSO, 2015, 41 (11) :1464-1470
[7]  
Krauss B, 2015, INVEST RADIOL, V50, P114, DOI 10.1097/RLI.0000000000000109
[8]   Applications of dual energy computed tomography in abdominal imaging [J].
Lestra, T. ;
Mule, S. ;
Millet, I. ;
Carsin-Vu, A. ;
Taourel, P. ;
Hoeffel, C. .
DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2016, 97 (06) :593-603
[9]   Dual-Energy CT Iodine Mapping and 40-keV Monoenergetic Applications in the Diagnosis of Acute Bowel Ischemia [J].
Lourenco, Pedro D. M. ;
Rawski, Ryan ;
Mohammed, Mohammed F. ;
Khosa, Faisal ;
Nicolaou, Savvas ;
McLaughlin, Patrick .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2018, 211 (03) :564-570
[10]   State of the Art: Dual-Energy CT of the Abdomen [J].
Marin, Daniele ;
Boll, Daniel T. ;
Mileto, Achille ;
Nelson, Rendon C. .
RADIOLOGY, 2014, 271 (02) :327-342