Tensile and Shear Loading Stability of All-Inside Meniscal Repairs An In Vitro Biomechanical Evaluation

被引:15
作者
Brucker, Peter U. [1 ]
Favre, Philippe [1 ]
Puskas, Gabor J. [1 ]
von Campe, Arndt [1 ]
Meyer, Dominik C. [1 ]
Koch, Peter P. [1 ]
机构
[1] Univ Zurich, Dept Orthopaed Balgrist, CH-8008 Zurich, Switzerland
关键词
meniscus; meniscal repair; all-inside meniscal repair device; tensile load; shear load; biomechanics; SUTURE ANCHORS; FAILURE STRENGTHS; PULL-OUT; MICROVASCULATURE; DISTRACTION; REFIXATION; SYSTEMS; DEVICES; LESIONS; WINNER;
D O I
10.1177/0363546510368131
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Most biomechanical studies for evaluation of the structural properties of meniscal repairs have been performed in tensile loading scenarios perpendicular to the circumferential meniscal fibers. However, meniscal repair constructs are also exposed to shear forces parallel to the circumferential meniscal fibers during healing, particularly in the midportion of the meniscus. Hypothesis: Material properties of meniscal repair devices cannot be extrapolated from tensile to shear load scenarios. Study Design: Controlled laboratory study. Methods: In 84 harvested and isolated bovine lateral menisci following removal of adjacent soft tissue, a standardized vertical lesion was set followed by repair using all-inside flexible (FasT-Fix, FasT-Fix AB, RapidLoc) and rigid (Meniscus Screw, Meniscus Arrow) meniscal repair devices. Vertical and horizontal 2.0 Ethibond sutures were used as controls. The repaired meniscal construct was tested in a tensile (parallel to the axis of the tested repair device) and shear load scenario (perpendicular to the axis of the tested repair device) at 5 mm/min and 37 degrees C environmental temperature. Maximum load to failure, stiffness, and failure mode were recorded. Results: The absolute load to failure values of each repair device in the shear scenario were only marginally different from the tensile load scenario. However, the stiffness of several tested devices was markedly reduced in the shear scenario. In both scenarios, large differences of the load to failure and the stiffness between the implant types up to 5-fold were found (P<.05). The failure mode of several all-inside flexible repair devices was different in the shear versus tensile load scenario, while the failure mode of the rigid systems was similar in both scenarios. Conclusion: All-inside meniscal repair devices exposed to shear load scenarios have comparable maximum loads to failures as tensile load scenarios. However, the stiffness of the majority of the flexible meniscal repair implants in a shear load scenario is markedly reduced. The applied scenario also affects the failure mode in several flexible meniscal repair devices. Clinical Relevance: Meniscal repair devices with sufficient stiffness and stability against shear loads may be favored for meniscal repair, especially within the midportion of the meniscus where shear loads occur during healing.
引用
收藏
页码:1838 / 1844
页数:7
相关论文
共 20 条
[1]   Failure strength of a new meniscus arrow repair technique: Biomechanical comparison with horizontal suture [J].
AlbrechtOlsen, P ;
Lind, T ;
Kristensen, G ;
Falkenberg, B .
ARTHROSCOPY, 1997, 13 (02) :183-187
[2]   MICROVASCULATURE OF THE HUMAN MENISCUS [J].
ARNOCZKY, SP ;
WARREN, RF .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1982, 10 (02) :90-95
[3]   Tensile fixation strengths of absorbable meniscal repair devices as a function of hydrolysis time - An in vitro experimental study [J].
Arnoczky, SP ;
Lavagnino, M .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2001, 29 (02) :118-123
[4]   WINNER OF THE 1982 ODONOGHUE AWARD - THE MICROVASCULATURE OF THE MENISCUS AND ITS RESPONSE TO INJURY - AN EXPERIMENTAL-STUDY IN THE DOG [J].
ARNOCZKY, SP ;
WARREN, RF .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1983, 11 (03) :131-141
[5]   Meniscal repair devices [J].
Barber, FA ;
Herbert, MA .
ARTHROSCOPY, 2000, 16 (06) :613-618
[6]   Biomechanical investigations of different meniscal repair implants in comparison with horizontal sutures on human meniscus [J].
Becker, R ;
Schröder, M ;
Stärke, C ;
Urbach, D ;
Nebelung, W .
ARTHROSCOPY, 2001, 17 (05) :439-444
[7]   Distraction forces on repaired bucket-handle lesions in the medial meniscus [J].
Becker, Roland ;
Brettschneider, Olaf ;
Groebel, Karl-Heinz ;
von Versen, Ruediger ;
Staerke, Christian .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2006, 34 (12) :1941-1947
[8]   Pull-out strength and stiffness of meniscal repair using absorbable arrows or Ti-Cron vertical and horizontal loop sutures [J].
Boenisch, UW ;
Faber, KJ ;
Ciarelli, M ;
Steadman, JR ;
Arnoczky, SP .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1999, 27 (05) :626-631
[9]   Biomechanical comparison of the FasT-Fix meniscal repair suture system with vertical mattress sutures and meniscus arrows [J].
Borden, P ;
Nyland, J ;
Caborn, DNM ;
Pienkowski, D .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2003, 31 (03) :374-378
[10]   Pull-out and shear failure strengths of arthroscopic meniscal repair systems [J].
Fisher, SR ;
Markel, DC ;
Koman, JD ;
Atkinson, TS .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2002, 10 (05) :294-299