The practical application of scintillation dosimetry in small-field photon-beam radiotherapy

被引:23
作者
Burke, Elisa [1 ]
Poppinga, Daniela [1 ]
Schoenfeld, Andreas A. [1 ]
Harder, Dietrich [2 ]
Poppe, Bjoern [1 ]
Looe, Hui Khee [1 ]
机构
[1] Carl von Ossietzky Univ Oldenburg, Pius Hosp, Univ Clin Med Radiat Phys, Med Campus, Oldenburg, Germany
[2] Georg August Univ, Med Phys & Biophys, Gottingen, Germany
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2017年 / 27卷 / 04期
关键词
Plastic scintillators; Cerenkov light correction; in-phantom field output factors; dose profile scans in water phantoms; EXRADIN W1 SCINTILLATOR; PLASTIC SCINTILLATOR; RELATIVE DOSIMETRY; FIBER DOSIMETER; DETECTORS; RADIATION; RESOLUTION; MV;
D O I
10.1016/j.zemedi.2016.11.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Plastic scintillation detectors are a new instrument of stereotactic photon-beam dosimetry. The clinical application of the plastic scintillation detector Exradin W1 at the Siemens Artiste and Elekta Synergy accelerators is a matter of current interest. In order to reduce the measurement uncertainty, precautions have to be taken with regard to the geometrical arrangement of the scintillator, the light guide fiber and the photodiode in the radiation field. To determine the "Cerenkov light ratio" CLR with a type A uncertainty below 1%, the Cerenkov calibration procedure for small-field measurements based on the two-channel spectral method was used. Output factors were correctly measured with the W1 for field sizes down to 0.5 x 0.5 cm(2) with a type A uncertainty of 1.8%. Measurements of small field dose profiles and percentage depth dose curves were carried out with the W1 using automated water phantom profile scans, and a type A uncertainty for dose maxima of 1.4% was achieved. The agreement with a synthetic diamond detector (microDiamond, PTW Freiburg) and a plane parallel ionization chamber (Roos chamber, PTW Freiburg) in relative dose measurements was excellent. In oversight of all results, the suitability of the plastic scintillation detector Exradin W1 for clinical dosimetry under stereotactic conditions, in particular the tried and tested procedures for CLR determination, output factor measurement and automated dose profile scans in water phantoms, have been confirmed.
引用
收藏
页码:324 / 333
页数:10
相关论文
共 38 条
[1]   A new formalism for reference dosimetry of small and nonstandard fields [J].
Alfonso, R. ;
Andreo, P. ;
Capote, R. ;
Huq, M. Saiful ;
Kilby, W. ;
Kjaell, P. ;
Mackie, T. R. ;
Palmans, H. ;
Rosser, K. ;
Seuntjens, J. ;
Ullrich, W. ;
Vatnitsky, S. .
MEDICAL PHYSICS, 2008, 35 (11) :5179-5186
[2]  
[Anonymous], 2016, 68098 DIN
[3]  
[Anonymous], 2012, 8064300 STAND IM EXR
[4]  
[Anonymous], 2008, 68002 DIN
[5]   Measurement accuracy and Cerenkov removal for high performance, high spatial resolution scintillation dosimetry [J].
Archambault, L ;
Beddar, AS ;
Gingras, L ;
Roy, R ;
Beaulieu, L .
MEDICAL PHYSICS, 2006, 33 (01) :128-135
[6]   Detector to detector corrections: A comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams [J].
Azangwe, Godfrey ;
Grochowska, Paulina ;
Georg, Dietmar ;
Izewska, Joanna ;
Hopfgartner, Johannes ;
Lechner, Wolfgang ;
Andersen, Claus E. ;
Beierholm, Anders R. ;
Helt-Hansen, Jakob ;
Mizuno, Hideyuki ;
Fukumura, Akifumi ;
Yajima, Kaori ;
Gouldstone, Clare ;
Sharpe, Peter ;
Meghzifene, Ahmed ;
Palmans, Hugo .
MEDICAL PHYSICS, 2014, 41 (07)
[7]   A high-precision, high-resolution and fast dosimetry system for beta sources applied in cardiovascular brachytherapy [J].
Bambynek, M ;
Flühs, D ;
Quast, U ;
Wegener, D ;
Soares, CG .
MEDICAL PHYSICS, 2000, 27 (04) :662-667
[8]   Current status of scintillation dosimetry for megavoltage beams [J].
Beaulieu, L. ;
Goulet, M. ;
Archambault, L. ;
Beddar, S. .
7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
[9]  
Beaulieu L, 2013, IEEE NSS MIC RTSD, P83
[10]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .2. PROPERTIES AND MEASUREMENTS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1901-1913