Characterization of a Three Wavelength Photoacoustic Soot Spectrometer (PASS-3) and a Photoacoustic Extinctiometer (PAX)

被引:73
作者
Nakayama, Tomoki [1 ,2 ]
Suzuki, Hiroyuki [1 ,2 ]
Kagamitani, Satomi [1 ,2 ]
Ikeda, Yuka [1 ,2 ]
Uchiyama, Akihiro [3 ]
Matsumi, Yutaka [1 ,2 ]
机构
[1] Nagoya Univ, Solar Terr Environm Lab, Nagoya, Aichi 4648601, Japan
[2] Nagoya Univ, Grad Sch Sci, Nagoya, Aichi 4648601, Japan
[3] Japan Meteorol Agcy, Meteorol Res Inst, Tsukuba, Ibaraki, Japan
关键词
photoacoustic spectrometer (PAS); reciprocal nephelometer; absorption coefficient; scattering coefficient; aerosol optical property; calibration; AEROSOL OPTICAL-PROPERTIES; COMPLEX REFRACTIVE-INDEX; BLACK CARBON AEROSOL; PULSED CAVITY RING; LIGHT-ABSORPTION; BROWN CARBON; INTEGRATING NEPHELOMETER; CALIBRATION; SPECTROSCOPY; 3-WAVELENGTH;
D O I
10.2151/jmsj.2015-016
中图分类号
P4 [大气科学(气象学)];
学科分类号
0706 ; 070601 ;
摘要
Laboratory experiments were conducted for characterizing the performance of two commercially available instruments employed for the measurement of light absorption and scattering coefficients of aerosols at lambda = 405, 532, and 781 nm (using three-wavelength photoacoustic soot spectrometer; PASS-3) as well as at 375 nm (using photoacoustic extinctiometer; PAX) based on photoacoustic spectroscopy and reciprocal nephelometry, respectively. The calibration factors (conversion factors from the readout to real values) associated with scattering measurements, estimated using gaseous molecules, mono-disperse polystyrene latex and ammonium sulfate particles, and/or polydisperse ammonium sulfate particles, are in good agreement with one another, typically within 5 %, 5 %, and 10 % at 375, 405, and 781 nm, respectively. In contrast, a significant particle size dependency was observed for the calibration factors at 532 nm, which is possibly because of a combination of differences in the polarization states of the lasers relative to the scattering planes and large truncation angle. Considering the estimated effective truncation angle, the typical uncertainties in calibration factors for scattering when measuring non-or weakly light-absorbing particles, with volume-based geometrical diameters of less than 700 nm, were estimated to be 12 %, 7 %, 34 %, and 17 %, at 375, 405, 532, and 781 nm, respectively. The typical uncertainties in the calibration factors for absorption measurements, which were determined using poly-disperse propane soot particles, were estimated to be 6 %, 4 %, 8 %, and 11 %, at 375, 405, 532, and 781 nm, respectively. The calibration factors for absorption determined by the poly-disperse soot particles at 375 and 405 nm were 48 % and 36 % smaller than those by light absorption of NO2 molecules possibly because of NO2 photolysis, although good agreement was observed at 532 nm. These results suggest that the photolysis effect should be taken into account when light absorption by NO2 is used for calibration at 375 and 405 nm.
引用
收藏
页码:285 / 308
页数:24
相关论文
共 61 条
[1]   Integrating nephelometer with a low truncation angle and an extended calibration scheme [J].
Abu-Rahmah, A. ;
Arnott, W. P. ;
Moosmuller, H. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2006, 17 (07) :1723-1732
[2]  
Air Resource Specialists Inc., 2005, REEV HFC 134A SUVA S
[3]   A novel multi-wavelength photoacoustic spectrometer for the measurement of the UV-vis-NIR spectral absorption coefficient of atmospheric aerosols [J].
Ajtai, Tibor ;
Filep, Agnes ;
Schnaiter, Martin ;
Linke, Claudia ;
Vragel, Marlen ;
Bozoki, Zoltan ;
Szabo, Gabor ;
Leisner, Thomas .
JOURNAL OF AEROSOL SCIENCE, 2010, 41 (11) :1020-1029
[4]  
Anderson TL, 1996, J ATMOS OCEAN TECH, V13, P967, DOI 10.1175/1520-0426(1996)013<0967:PCOAHS>2.0.CO
[5]  
2
[6]   Determining aerosol radiative properties using the TSI 3563 integrating nephelometer [J].
Anderson, TL ;
Ogren, JA .
AEROSOL SCIENCE AND TECHNOLOGY, 1998, 29 (01) :57-69
[7]   Black carbon or brown carbon?: The nature of light-absorbing carbonaceous aerosols [J].
Andreae, M. O. ;
Gelencser, A. .
ATMOSPHERIC CHEMISTRY AND PHYSICS, 2006, 6 :3131-3148
[8]  
[Anonymous], 2011, JPL PUBLICATION
[9]   Nitrogen dioxide and kerosene-flame soot calibration of photoacoustic instruments for measurement of light absorption by aerosols [J].
Arnott, WP ;
Moosmüller, H ;
Walker, JW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (12) :4545-4552
[10]  
Arnott WP, 1999, ATMOS ENVIRON, V33, P2845