Pulse Wave Imaging Coupled With Vector Flow Mapping: A Phantom, Simulation, and In Vivo Study

被引:25
作者
Karageorgos, Grigorios Marios [1 ]
Apostolakis, Iason-Zacharias [1 ]
Nauleau, Pierre [1 ]
Gatti, Vittorio [1 ]
Weber, Rachel [1 ]
Kemper, Paul [1 ]
Konofagou, Elisa E. [1 ,2 ]
机构
[1] Columbia Univ, Biomed Engn Dept, New York, NY 10027 USA
[2] Columbia Univ, Dept Radiol, New York, NY 10027 USA
基金
美国国家卫生研究院;
关键词
Arterial wall stiffness; flow mapping in vivo; fluid structure interaction (FSI) simulations; pulse wave imaging (PWI); speckle tracking; vector Doppler; vector flow field; vector flow imaging; WALL SHEAR-STRESS; NONINVASIVE VASCULAR ELASTOGRAPHY; BLOOD VELOCITY ESTIMATION; COMMON CAROTID ARTERIES; TRANSVERSE OSCILLATION; MECHANICAL-PROPERTIES; HIGH-FREQUENCY; ULTRASOUND; DOPPLER; ATHEROSCLEROSIS;
D O I
10.1109/TUFFC.2021.3074113
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Pulse wave imaging (PWI) is an ultrasound imaging modality that estimates the wall stiffness of an imaged arterial segment by tracking the pulse wave propagation. The aim of the present study is to integrate PWI with vector flow imaging, enabling simultaneous and co-localized mapping of vessel wall mechanical properties and 2-D flow patterns. Two vector flow imaging techniques were implemented using the PWI acquisition sequence: 1) multiangle vector Doppler and 2) a cross-correlation-based vector flow imaging (CC VFI) method. The two vector flow imaging techniques were evaluated in vitro using a vessel phantom with an embedded plaque, along with spatially registered fluid structure interaction (FSI) simulations with the same geometry and inlet flow as the phantom setup. The flow magnitude and vector direction obtained through simulations and phantom experiments were compared in a prestenotic and stenotic segment of the phantom and at five different time frames. In most comparisons, CC VFI provided significantly lower bias or precision than the vector Doppler method (p < 0.05) indicating better performance. In addition, the proposed technique was applied to the carotid arteries of nonatherosclerotic subjects of different ages to investigate the relationship between PWI-derived compliance of the arterial wall and flow velocity in vivo. Spearman's rank-order test revealed positive correlation between compliance and peak flow velocity magnitude (r(s) = 0.90 and p < 0.001), while significantly lower compliance (p < 0.01) and lower peak flow velocity magnitude (p < 0.001) were determined in older (54-73 y.o.) compared with young (24-32 y.o.) subjects. Finally, initial feasibility was shown in an atherosclerotic common carotid artery in vivo. The proposed imaging modality successfully provided information on blood flow patterns and arterial wall stiffness and is expected to provide additional insight in studying carotid artery biomechanics, as well as aid in carotid artery disease diagnosis and monitoring.
引用
收藏
页码:2516 / 2531
页数:16
相关论文
共 84 条
[1]   Adaptive Pulse Wave Imaging: Automated Spatial Vessel Wall Inhomogeneity Detection in Phantoms and in-Vivo [J].
Apostolakis, Iason Z. ;
Karageorgos, Grigorios M. ;
Nauleau, Pierre ;
Nandlall, Sacha D. ;
Konofagou, Elisa E. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2020, 39 (01) :259-269
[2]   Pulse wave imaging using coherent compounding in a phantom and in vivo [J].
Apostolakis, Iason Zacharias ;
McGarry, Matthew D. J. ;
Bunting, Ethan A. ;
Konofagou, Elisa E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (05) :1700-1730
[3]   Piecewise Pulse Wave Imaging (pPWI) for Detection and Monitoring of Focal Vascular Disease in Murine Aortas and Carotids In Vivo [J].
Apostolakis, Iason Zacharias ;
Nandlall, Sacha D. ;
Konofagou, Elisa E. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2016, 35 (01) :13-28
[4]   Feasibility and Validation of 4-D Pulse Wave Imaging in Phantoms and In Vivo [J].
Apostolakis, Iason-Zacharias ;
Nauleau, Pierre ;
Papadacci, Clement ;
McGarry, Matthew D. ;
Konofagou, Elisa E. .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2017, 64 (09) :1305-1317
[5]  
Ateshian G. A., 2018, J BIOMECH ENG-T ASME, V140
[6]   Combined 2-D Vector Velocity Imaging and Tracking Doppler for Improved Vascular Blood Velocity Quantification [J].
Avdal, Jorgen ;
Lovstakken, Lasse ;
Torp, Hans ;
Ekroll, Ingvild Kinn .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2017, 64 (12) :1795-1804
[7]   EFFECTS OF AGING ON CHANGING ARTERIAL COMPLIANCE AND LEFT-VENTRICULAR LOAD IN A NORTHERN CHINESE URBAN-COMMUNITY [J].
AVOLIO, AP ;
CHEN, SG ;
WANG, RP ;
ZHANG, CL ;
LI, MF ;
OROURKE, MF .
CIRCULATION, 1983, 68 (01) :50-58
[8]   Abdominal aortic aneurysm follow-up by shear wave elasticity imaging after endovascular repair in a canine model [J].
Bertrand-Grenier, Antony ;
Lerouge, Sophie ;
Tang, An ;
Salloum, Eli ;
Therasse, Eric ;
Kauffmann, Claude ;
Heon, Helene ;
Salazkin, Igor ;
Cloutier, Guy ;
Soulez, Gilles .
EUROPEAN RADIOLOGY, 2017, 27 (05) :2161-2169
[9]   Carotid Artery Plaque Vulnerability Assessment Using Noninvasive Ultrasound Elastography: Validation With MRI [J].
Cardinal, Marie-Helene Roy ;
Heusinkveld, Maarten H. G. ;
Qin, Zhao ;
Lopata, Richard G. P. ;
Naim, Cyrille ;
Soulez, Gilles ;
Cloutier, Guy .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2017, 209 (01) :142-151
[10]   NON-INVASIVE IN VIVO CHARACTERIZATION OF HUMAN CAROTID PLAQUES WITH ACOUSTIC RADIATION FORCE IMPULSE ULTRASOUND: COMPARISON WITH HISTOLOGY AFTER ENDARTERECTOMY [J].
Czernuszewicz, Tomasz J. ;
Homeister, Jonathon W. ;
Caughey, Melissa C. ;
Farber, Mark A. ;
Fulton, Joseph J. ;
Ford, Peter F. ;
Marston, William A. ;
Vallabhaneni, Raghuveer ;
Nichols, Timothy C. ;
Gallippi, Caterina M. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (03) :685-697