Ultratrace Determination of Tin, Germanium, and Selenium by Hydride Generation Coupled with a Novel Solution-Cathode Glow Discharge-Atomic Emission Spectrometry Method

被引:54
|
作者
Huang, Chuchu [1 ]
Li, Qing [1 ]
Mo, Jiamei [1 ]
Wang, Zheng [1 ]
机构
[1] Chinese Acad Sci, Shanghai Inst Ceram, Shanghai 200050, Peoples R China
基金
中国国家自然科学基金;
关键词
DIELECTRIC-BARRIER DISCHARGE; FLUORESCENCE SPECTROMETRY; ABSORPTION-SPECTROMETRY; ANALYTICAL PERFORMANCE; SIGNAL ENHANCEMENT; LIQUID CATHODE; SPECIATION; ANTIMONY; SAMPLES; CONTACT;
D O I
10.1021/acs.analchem.6b02807
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We herein describe a novel method of hydride generation (HG) coupled to a newly designed atmospheric pressure solution-cathode glow discharge (SCGD) spectrometric technique for the ultratrace determination of tin, germanium, and selenium. In this novel SCGD process, gas introduction was permitted using a hollow titanium tube as both the anode and sampling port. In these experiments, the analytes were converted into volatile hydrides upon passing through the hydride generator, and were introduced into the near-anode region of the SCGD system, where they were detected directly by atomic emission spectrometry (AES). A significant improvement in both selectivity and sensitivity was achieved, which was reflected in an improvement in the detection limits (DLs) by 3 orders of magnitude, in addition to successful valence analysis of Se without the requirement for chromatographic separation. In the absence of a strict sample pretreatment process and with a reduction in electrolyte consumption, the detection limits of Sn, Ge, and Se were determined to be 0.8, 0.5, and 0.2 mu g.L-1. Moreover, our HG-SCGD-AES system demonstrated excellent repeatability (<3% peak height relative standard deviation) and more than 2 orders of linear dynamic range. The optimal operating conditions are outlined herein, and the analytical performance of the system is evaluated as described. Furthermore, our method was applied for the analysis of Sn, Ge, and Se in both environmental and biological samples, and the obtained results were in good agreement with reference values.
引用
收藏
页码:11559 / 11567
页数:9
相关论文
共 50 条
  • [41] Multi-element Simultaneous sensitization of solution cathode glow discharge atomic emission spectrometry by using portable semiconductor anode refrigeration
    Chen, Xiangyu
    Cai, Zhaoqing
    Wang, Zheng
    TALANTA, 2022, 248
  • [42] A novel atmospheric pressure glow discharge system for sensitive determination of As, Sb, and Se by optical emission spectrometry following their hydride generation
    Greda, Krzysztof
    Kratzer, Jan
    Vlckova, Nikol
    Pohl, Pawel
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2024, 213
  • [43] Direct and Sensitive Determination of Antimony in Water by Hydrogen-Doped Solution Anode Glow Discharge-Optical Emission Spectrometry Without Hydride Generation
    Yang, Chun
    Cheng, Guo
    Cheng, Shuang-Quan
    Liu, Xing
    Liu, Ying
    Zheng, Hong-Tao
    Hu, Sheng-Hong
    Zhu, Zhen-Li
    ANALYTICAL CHEMISTRY, 2021, 93 (49) : 16393 - 16400
  • [44] Ultra-sensitive determination of mercury by atmospheric pressure glow discharge atomic emission spectrometry coupled with cold vapor generation
    Rong, Rong
    Cai, Zhaoqing
    Li, Xiaoyan
    Wang, Zheng
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (11) : 2377 - 2382
  • [45] Application of Microwave Plasma Atomic Emission Spectrometry and Hydride Generation for Determination of Arsenic and Selenium in Mineral Water
    Mikheev, I. V.
    Karpukhina, E. A.
    Usol'tseva, L. O.
    Samarina, T. O.
    Volkov, D. S.
    Proskurnin, M. A.
    INORGANIC MATERIALS, 2017, 53 (14) : 1422 - 1426
  • [46] Optimization of selenium determination in chicken's meat and eggs by the hydride-generation atomic absorption spectrometry method
    Turker, Ali Rehber
    Erol, Esra
    INTERNATIONAL JOURNAL OF FOOD SCIENCES AND NUTRITION, 2009, 60 (01) : 40 - 50
  • [47] High Sensitivity Determination of Antimony with Application for the Characterization of Its Migration in Bottled Water by a Dielectric Barrier Discharge (DBD) Coupled with Hydride Generation - Atomic Fluorescence Spectrometry (HG-AFS)
    Liu, Meitong
    Ding, Lan
    Liu, Jixin
    Na, Xing
    Mao, Xuefei
    ANALYTICAL LETTERS, 2021, 54 (06) : 990 - 1004
  • [48] Ultrasensitive determination of mercury by ICP-OES coupled with a vapor generation approach based on solution cathode glow discharge
    Cai, Zhaoqing
    Zou, Huijun
    Chen, Yirui
    Wang, Zheng
    CHINESE CHEMICAL LETTERS, 2022, 33 (05) : 2692 - 2696
  • [49] Determination of arsenic and selenium by hydride generation and headspace solid phase microextraction coupled with optical emission spectrometry
    Tyburska, Anna
    Jankowski, Krzysztof
    Rodzik, Agnieszka
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2011, 66 (07) : 517 - 521
  • [50] Solution Cathode Glow Discharge - Atomic Emission Spectrometry (SCGD-AES) With an Interference Filter for Spectral Discrimination for the Determination of Thallium in Surface Water
    Wang, Jinmei
    Fei, Guoxing
    Zheng, Peichao
    Yan, Nantian
    Chen, Guanghui
    Li, Wei
    Dong, Daming
    Guo, Lianbo
    ANALYTICAL LETTERS, 2025, 58 (03) : 415 - 424