Dynamic input to determine hip joint moments, power and work on the prosthetic limb of transfemoral amputees: ground reaction vs knee reaction

被引:40
作者
Frossard, Laurent [1 ]
Cheze, Laurence [2 ]
Dumas, Raphael [2 ]
机构
[1] Univ Quebec, Grp Res Adapted Phys Activ, Montreal, PQ H3C 3P8, Canada
[2] Univ Lyon, Lab Biomecan & Mecan Chocs, UMR T9406, F-69675 Lyon, France
基金
澳大利亚研究理事会;
关键词
Biomechanics of prosthetic/orthotic devices; prosthetic design; rehabilitation of amputees; testing of prosthetic and orthotic components; SPECIAL GAIT REEDUCATION; TRANS-FEMORAL AMPUTEES; ENERGY-EXPENDITURE; OSSEOINTEGRATED FIXATION; WALKING; REHABILITATION; BIOMECHANICS; AMPUTATION; ALIGNMENT; MUSCLES;
D O I
10.1177/0309364611409002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Calculation of lower limb kinetics is limited by floor-mounted force-plates. Objectives: Comparison of hip joint moments, power and mechanical work on the prosthetic limb of a transfemoral amputee calculated by inverse dynamics using either the ground reactions (force-plates) or knee reactions (transducer). Study design: Comparative analysis. Methods: Kinematics, ground reaction and knee reaction data were collected using a motion analysis system, two force-plates, and a multi-axial transducer mounted below the socket, respectively. Results: The inverse dynamics using ground reaction underestimated the peaks of hip energy generation and absorption occurring at 63% and 76% of the gait cycle (GC) by 28% and 54%, respectively. This method also overestimated by 24% a phase of negative work at the hip (37%-56% GC), and underestimated the phases of positive (57%-72% GC) and negative (73%-98%GC) work at the hip by 11% and 58%, respectively. Conclusions: A transducer mounted within the prosthesis has the capacity to provide more realistic kinetics of the prosthetic limb because it enables assessment of multiple consecutive steps and a wide range of activities without the issue of foot placement on force-plates.
引用
收藏
页码:140 / 149
页数:10
相关论文
共 47 条
[1]   The Endo-Exo Femur Prosthesis - A New Concept of Bone-Guided, Prosthetic Rehabilitation Following Above-Knee Amputation [J].
Aschoff, H. H. ;
Clausen, A. ;
Hoffmeister, T. .
ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE, 2009, 147 (05) :610-615
[2]  
Blumentritt S., 1998, J PROSTHET ORTHOT, V10, P61
[3]   CALCULATION OF VERTICAL GROUND REACTION FORCE ESTIMATES DURING RUNNING FROM POSITIONAL DATA [J].
BOBBERT, MF ;
SCHAMHARDT, HC ;
NIGG, BM .
JOURNAL OF BIOMECHANICS, 1991, 24 (12) :1095-1105
[4]  
Brånemark R, 2001, J REHABIL RES DEV, V38, P175
[5]   Hip and knee joints are more stabilized than driven during the stance phase of gait: An analysis of the 3D angle between joint moment and joint angular velocity [J].
Dumas, R. ;
Cheze, L. .
GAIT & POSTURE, 2008, 28 (02) :243-250
[6]  
Dumas R., 2004, Computer Methods in Biomechanics and Biomedical Engineering, V7, P159, DOI 10.1080/10255840410001727805
[7]   Loading applied on prosthetic knee of transfemoral amputee: Comparison of inverse dynamics and direct measurements [J].
Dumas, R. ;
Cheze, L. ;
Frossard, L. .
GAIT & POSTURE, 2009, 30 (04) :560-562
[8]   A survey of formal methods for determining functional joint axes [J].
Ehrig, Rainald M. ;
Taylor, William R. ;
Duda, Georg N. ;
Heller, Markus O. .
JOURNAL OF BIOMECHANICS, 2007, 40 (10) :2150-2157
[9]   Evaluation of a mixed approach combining stationary and wearable systems to monitor gait over long distance [J].
Favre, J. ;
Crevoisier, X. ;
Jolles, B. M. ;
Aminian, K. .
JOURNAL OF BIOMECHANICS, 2010, 43 (11) :2196-2202
[10]  
FOERSTER SA, 1995, J REHABIL RES DEV, V32, P128