Comparison of formula accuracy for intraocular lens power calculation based on measurements by a swept-source optical coherence tomography optical biometer

被引:73
作者
Savini, Giacomo [1 ]
Hoffer, Kenneth J. [4 ,5 ]
Balducci, Nicole [1 ]
Barboni, Piero [2 ,3 ]
Schiano-Lomoriello, Domenico [1 ]
机构
[1] IRCCS GB Bietti Fdn, Via Livenza 3, Rome, Italy
[2] Studio Oculist Azeglio, Bologna, Italy
[3] Osped San Raffaele, Milan, Italy
[4] Univ Calif Los Angeles, Stein Eye Inst, Los Angeles, CA USA
[5] St Marys Eye Ctr, Santa Monica, CA USA
关键词
SYSTEM;
D O I
10.1016/j.jcrs.2019.08.044
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To analyze the results of intraocular lens (IOL) power calculation using measurements by a swept-source optical coherence tomography (SS-OCT) optical biometer. Setting: IRCCS G.B. Bietti Foundation, Rome, Italy. Design: Evaluation of a diagnostic test instrument. Methods: Preoperative measurements by the OA-2000 (Tomey Inc.) were taken in a consecutive series of patients undergoing cataract surgery with one IOL model (AcrySof SN60WF; Alcon Laboratories, Inc.). Measurements were entered into the following formulas: Barrett Universal II, Emmetropia Verifying Optical (EVO), Haigis, Hoffer Q, Holladay 1, Holladay 2, Holladay 2 with axial length adjustment, Kane, Olsen, Panacea, SRIVT, T2, and VRF. When refraction was measured at 1 month postoperatively, the mean arithmetic prediction error, the median absolute error (MedAE), and the percentage of eyes with a error of +/- 0.50 D or less were calculated after constant optimization. Results: We enrolled 150 eyes. All formulas yielded excellent outcomes. The MedAE ranged between 0.200 D and 0.259 D, with a statistically significant difference among formulas (P = .0004). The lowest MedAE values were obtained with the Barrett, EVO, Kane, Olsen(standalone), Radial Basis Function (RBF), and T2 formulas. The percentage of eyes with a prediction error of +/- 0.50 D or less ranged between 80.00% and 90.67%, with a statistically significant difference (P < .0001). The Barrett, EVO, Holladay 2 with axial length adjustment, Kane, RBF, and T2 achieved the highest percentages (>= 88%). Conclusions: Measurements provided by the SS-OCT optical biometer enable accurate IOL power calculation because all formulas yielded a prediction error within 0.50 D in at least 80% of eyes. Copyright (C) 2019 Published by Wolters Kluwer on behalf of ASCRS and ESCRS
引用
收藏
页码:27 / 33
页数:7
相关论文
共 25 条
[1]  
Barrett G, BARRETT UNIVERSAL 2
[2]   THE LENS OPACITIES CLASSIFICATION SYSTEM-III [J].
CHYLACK, LT ;
WOLFE, JK ;
SINGER, DM ;
LESKE, MC ;
BULLIMORE, MA ;
BAILEY, IL ;
FRIEND, J ;
MCCARTHY, D ;
WU, SY .
ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (06) :831-836
[3]   Comparison of the Kane formula with existing formulas for intraocular lens power selection [J].
Connell, Benjamin J. ;
Kane, Jack X. .
BMJ OPEN OPHTHALMOLOGY, 2019, 4 (01)
[4]   Comparison of 9 intraocular lens power calculation formulas [J].
Cooke, David L. ;
Cooke, Timothy L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (08) :1157-1164
[5]   Benchmark standards for refractive outcomes after NHS cataract surgery [J].
Gale, R. P. ;
Saldana, M. ;
Johnston, R. L. ;
Zuberbuhler, B. ;
McKibbin, M. .
EYE, 2009, 23 (01) :149-152
[6]   Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis [J].
Haigis, W ;
Lege, B ;
Miller, N ;
Schneider, B .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (09) :765-773
[7]  
Haigis W., 2004, Intraocular Lens Power Calculations, P41
[8]  
Hoffer KJ, 2007, J CATARACT REFR SURG, V33, P2, DOI 10.1016/j.jcrs.2006.08.054
[9]   Protocols for Studies of Intraocular Lens Formula Accuracy [J].
Hoffer, Kenneth J. ;
Aramberri, Jaime ;
Haigis, Wolfgang ;
Olsen, Thomas ;
Savini, Giacomo ;
Shammas, H. John ;
Bentow, Stanley .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2015, 160 (03) :403-405
[10]   Clinical results using the Holladay 2 intraocular lens power formula [J].
Hoffer, KJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (08) :1233-1237