Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry

被引:3
作者
Huang, Jinhai [1 ,2 ]
Savini, Giacomo [3 ]
Wang, Chengfang [1 ]
Lu, Weicong [1 ,2 ]
Gao, Rongrong [1 ]
Li, Yuanguang [1 ]
Wang, Qinmei [1 ,2 ]
Zhao, Yune [1 ,2 ]
机构
[1] Wenzhou Med Univ, Sch Ophthalmol & Optometry, Hosp Eye, Wenzhou 325027, Zhejiang, Peoples R China
[2] Minist Hlth, Key Lab Vis Sci, Wenzhou, Zhejiang, Peoples R China
[3] IRCCS, GB Bietti Eye Fdn, Rome, Italy
关键词
CHAMBER DEPTH MEASUREMENT; REPEATABILITY; REPRODUCIBILITY; PENTACAM; ORBSCAN; TOMOGRAPHY; ANALYZER; SYSTEM;
D O I
10.1155/2015/328798
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To assess the reliability and comparability of measuring central corneal thickness (CCT) and thinnest corneal thickness (TCT) using a new Scheimpflug-Placido analyzer (TMS-5, Japan) and ultrasound (US) pachymetry. Methods. Seventy-six healthy subjects were prospectively measured 3 times by 1 operator using the TMS-5, 3 additional consecutive scans were performed by a second operator, and ultrasound (US) pachymetry measurements were taken. The test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intraoperator repeatability and interoperator reproducibility. Agreement among the devices was assessed using Bland-Altman plots and 95% limits of agreement (LoA). Results. The intraoperators TRT and CoV were <19 mu m and 2.0%, respectively. The interoperators TRT and CoV were <12 mu m and 1.0%, respectively, and ICC was >0.90. The mean CCT and TCT measurements using the TMS-5 were 15.97 mu m (95% LoA from -26.42 to -5.52 mu m) and 20.32 mu m (95% LoA from -30.67 to -9.97 mu m) smaller, respectively, than those using US pachymetry. Conclusions. The TMS-5 shows good repeatability and reproducibility for measuring CCT and TCT in normal subjects but only moderate agreement with US pachymetry results. Caution is warranted before using these techniques interchangeably.
引用
收藏
页数:6
相关论文
共 29 条
[21]   Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS pentacam [J].
O'Donnell, C ;
Maldonado-Codina, C .
CORNEA, 2005, 24 (08) :920-924
[22]   Central corneal pachymetry in patients undergoing laser in situ keratomileusis [J].
Price, FW ;
Koller, DL ;
Price, MO .
OPHTHALMOLOGY, 1999, 106 (11) :2216-2220
[23]   Risk assessment for ectasia after corneal refractive surgery [J].
Randleman, J. Bradley ;
Woodward, Maria ;
Lynn, Michael J. ;
Stulting, R. Doyle .
OPHTHALMOLOGY, 2008, 115 (01) :37-50
[24]   Comparison of anterior segment measurements by 3 Scheimpflug tomographers and 1 Placido corneal topographer [J].
Savini, Giacomo ;
Carbonelli, Michele ;
Sbreglia, Alessandra ;
Barboni, Piero ;
Deluigi, Giulia ;
Hoffer, Kenneth J. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (09) :1679-1685
[25]   Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes [J].
Savini, Giacomo ;
Carbonelli, Michele ;
Barboni, Piero ;
Hoffer, Kenneth J. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (02) :302-309
[26]   Comparison of Central Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-Slit Topography [J].
Sedaghat, Mohammad Reza ;
Daneshvar, Ramin ;
Kargozar, Abbas ;
Derakhshan, Akbar ;
Daraei, Mona .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2010, 150 (06) :780-789
[27]   Comparison of optical low coherence reflectometry and ultrasound pachymetry in the measurement of central corneal thickness before and after photorefractive keratectomy [J].
Spadea, Leopoldo ;
Giammaria, Daniele ;
Di Genova, Lucia ;
Fiasca, Arianna .
JOURNAL OF REFRACTIVE SURGERY, 2007, 23 (07) :661-666
[28]   Anterior Chamber Depth Measurement in Phakic and Pseudophakic Eyes [J].
Su, Po-Fang ;
Lo, Andy Y. ;
Hu, Chao-Yu ;
Chang, Shu-Wen .
OPTOMETRY AND VISION SCIENCE, 2008, 85 (12) :1193-1200
[29]   Repeatability of corneal power and wavefront aberration measurements with a dual-Scheimpflug Placido corneal topographer [J].
Wang, Li ;
Shirayama, Mariko ;
Koch, Douglas D. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (03) :425-430