Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry

被引:3
作者
Huang, Jinhai [1 ,2 ]
Savini, Giacomo [3 ]
Wang, Chengfang [1 ]
Lu, Weicong [1 ,2 ]
Gao, Rongrong [1 ]
Li, Yuanguang [1 ]
Wang, Qinmei [1 ,2 ]
Zhao, Yune [1 ,2 ]
机构
[1] Wenzhou Med Univ, Sch Ophthalmol & Optometry, Hosp Eye, Wenzhou 325027, Zhejiang, Peoples R China
[2] Minist Hlth, Key Lab Vis Sci, Wenzhou, Zhejiang, Peoples R China
[3] IRCCS, GB Bietti Eye Fdn, Rome, Italy
关键词
CHAMBER DEPTH MEASUREMENT; REPEATABILITY; REPRODUCIBILITY; PENTACAM; ORBSCAN; TOMOGRAPHY; ANALYZER; SYSTEM;
D O I
10.1155/2015/328798
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To assess the reliability and comparability of measuring central corneal thickness (CCT) and thinnest corneal thickness (TCT) using a new Scheimpflug-Placido analyzer (TMS-5, Japan) and ultrasound (US) pachymetry. Methods. Seventy-six healthy subjects were prospectively measured 3 times by 1 operator using the TMS-5, 3 additional consecutive scans were performed by a second operator, and ultrasound (US) pachymetry measurements were taken. The test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intraoperator repeatability and interoperator reproducibility. Agreement among the devices was assessed using Bland-Altman plots and 95% limits of agreement (LoA). Results. The intraoperators TRT and CoV were <19 mu m and 2.0%, respectively. The interoperators TRT and CoV were <12 mu m and 1.0%, respectively, and ICC was >0.90. The mean CCT and TCT measurements using the TMS-5 were 15.97 mu m (95% LoA from -26.42 to -5.52 mu m) and 20.32 mu m (95% LoA from -30.67 to -9.97 mu m) smaller, respectively, than those using US pachymetry. Conclusions. The TMS-5 shows good repeatability and reproducibility for measuring CCT and TCT in normal subjects but only moderate agreement with US pachymetry results. Caution is warranted before using these techniques interchangeably.
引用
收藏
页数:6
相关论文
共 29 条
[1]   Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry [J].
Al-Mezaine H.S. ;
Al-Amro S.A. ;
Kangave D. ;
Sadaawy A. ;
Wehaib T.A. ;
Al-Obeidan S. .
International Ophthalmology, 2008, 28 (5) :333-338
[2]   Repeatability and comparability of corneal thickness measurements obtained from Dual Scheimpflug Analyzer and from ultrasonic pachymetry [J].
Al-Mohtaseb, Zaina N. ;
Wang, Li ;
Weikert, Mitchell P. .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2013, 251 (07) :1855-1860
[3]  
Bland JM, 1996, BRIT MED J, V313, P744
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry [J].
Chen, Shihao ;
Huang, Jinhai ;
Wen, Daizong ;
Chen, Wuhe ;
Huang, David ;
Wang, Qinmei .
ACTA OPHTHALMOLOGICA, 2012, 90 (05) :449-455
[6]  
Gordon MO, 2002, ARCH OPHTHALMOL-CHIC, V120, P714
[7]   CENTRAL AND PERIPHERAL CORNEAL THICKNESS IN KERATOCONUS AND NORMAL-PATIENT GROUPS [J].
GROMACKI, SJ ;
BARR, JT .
OPTOMETRY AND VISION SCIENCE, 1994, 71 (07) :437-441
[8]   Corneal thickness, curvature, and elevation readings in normal corneas: Combined Placido-Scheimpflug system versus combined Placido-scanning-slit system [J].
Guilbert, Emmanuel ;
Saad, Alain ;
Grise-Dulac, Alice ;
Gatinel, Damien .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (07) :1198-1206
[9]   Repeatability, Reproducibility, and Agreement Between Three Different Scheimpflug Systems in Measuring Corneal and Anterior Segment Biometry [J].
Hernandez-Camarena, Julio C. ;
Chirinos-Saldana, Patricia ;
Navas, Alejandro ;
Ramirez-Miranda, Arturo ;
de la Mota, Angie ;
Jimenez-Corona, Aida ;
Graue-Hernandez, Enrique O. .
JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (09) :616-U44
[10]   Comparison of Corneal Thickness Measurements Using Galilei, HR Pentacam, and Ultrasound [J].
Hosseini, Hamid Reza Jahadi ;
Katbab, Asadollah ;
Khalili, Mohammad Reza ;
Abtahi, Mohammad Bagher .
CORNEA, 2010, 29 (10) :1091-1095