Experimental and Theoretical Evaluation of a High Resolution CMOS Based Detector Under X-Ray Imaging Conditions

被引:50
作者
Michail, Christos M. [1 ]
Spyropoulou, Vasiliki A. [1 ]
Fountos, George P. [2 ]
Kalyvas, Nektarios I. [2 ]
Valais, Ioannis G. [2 ]
Kandarakis, Ioannis S. [2 ]
Panayiotakis, George S. [1 ]
机构
[1] Univ Patras, Sch Med, Dept Med Phys, Patras 26500, Greece
[2] Inst Educ Technol, Dept Med Instruments Technol, Athens 12210, Greece
关键词
CMOS; digital mammography; digital radiography; image quality; DIGITAL MAMMOGRAPHY; QUANTUM EFFICIENCY; PHYSICAL-CHARACTERIZATION; QUALITY MEASUREMENTS; PRESAMPLED MTF; PERFORMANCE; PHOSPHORS; SCREENS; IMAGER; NOISE;
D O I
10.1109/TNS.2010.2094206
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Fundamental imaging performance in terms of Modulation Transfer Function (MTF), Noise Power Spectrum (NPS) and Detective Quantum Efficiency (DQE) was investigated for a high resolution CMOS based imaging sensor. The device consists of a 33.91 mg/cm(2) Gd2O2S:Tb scintillator screen, placed in direct contact with a CMOS photodiode array. The CMOS photodiode array, featuring 1200 x 1600 pixels with a pixel pitch of 22.5 mu m, was used as an optical photon detector. In addition to the conventional frequency dependent parameters characterizing image quality, image information content was assessed through the application of information capacity (IC). The MTF was measured using the slanted-edge method to avoid aliasing while the Normalized NPS (NNPS) was determined by two-dimensional (2D) Fourier transforming of uniformly exposed images. Both measurements were performed under the representative radiation quality (RQA) settings, RQA-5 (70 kVp digital-radiography) and RQA-M2 (28 kVp digital-mammography) recommended by the International Electrotechnical Commission Reports 62220-1 and 62220-1-2 respectively. The DQE was assessed from the measured MTF, NPS and the direct entrance surface air-Kerma (ESAK) obtained from X-ray spectra measurement with a portable cadmium telluride (CdTe) detector. The ESAK values ranged between 11-87 mu Gy for RQA-5 and 6-40 mu Gy for RQA-M2. Additionally the output electrons per X-ray photon of the detector and its signal transfer characteristics were assessed via an analytical model, within the framework of the linear cascaded systems (LCS) theory. It was found that the detector response function was linear for the exposure ranges under investigation. Additionally our results showed that for the same RQA quality the output electrons per X-ray photon, as well as the measured and analytically predicted MTF, were not significantly affected by ESAK. MTF and DQE where found better compared to previously published data for other CCD and CMOS sensors, while the NNPS appeared to be comparable in the frequency range under investigation (0-10 cycles/mm).
引用
收藏
页码:314 / 322
页数:9
相关论文
共 58 条
[21]   Measurement of the detective quantum efficiency (DQE) of digital X-ray detectors according to the novel standard IEC 62220-1 [J].
Illers, H ;
Buhr, E ;
Hoeschen, C .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :39-44
[22]   System performance of a prototype flat-panel imager operated under marnmographic conditions [J].
Jee, KW ;
Antonuk, LE ;
El-Mohri, Y ;
Zhao, QH .
MEDICAL PHYSICS, 2003, 30 (07) :1874-1890
[23]   Modeling quantum noise of phosphors used in medical X-ray imaging detectors [J].
Kalivas, N ;
Kandarakis, I ;
Cavouras, D ;
Costaridou, L ;
Nomicos, CD ;
Panayiotakis, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 430 (2-3) :559-569
[24]   Evaluating x-ray detectors for radiographic applications: A comparison of ZnSCdS:Ag with Gd2O2S:Tb and Y2O2S:Tb screens [J].
Kandarakis, I ;
Cavouras, D ;
Panayiotakis, GS ;
Nomicos, CD .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (07) :1351-1373
[25]   Experimental investigation of the optical signal, gain, signal-to-noise ratio and information content characteristics of X-ray phosphor screens [J].
Kandarakis, I ;
Cavouras, D ;
Panayiotakis, GS ;
Nomicos, CD .
APPLIED PHYSICS B-LASERS AND OPTICS, 2001, 72 (07) :877-883
[26]   Development of a lens-coupled CMOS detector for an X-ray inspection system [J].
Kim, HK ;
Ahn, JK ;
Cho, GS .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 545 (1-2) :210-216
[27]  
Kim HK, 2001, IEEE T NUCL SCI, V48, P662, DOI 10.1109/23.940143
[28]   Investigation of scintillation screens for X-ray imaging [J].
Kim, Jung-Min ;
Kim, Ho Kyung ;
Cheong, Min Ho ;
Cho, Min Kook ;
Shon, Cheol-Soon ;
Lim, Chang Hwy .
ADVANCED NONDESTRUCTUVE EVALUATION I, PTS 1 AND 2, PROCEEDINGS, 2006, 321-323 :1056-1059
[29]  
Kim KH, 2004, IEEE NUCL SCI CONF R, P1425
[30]  
KIM KH, 2003, P IEEE NUCL SCI S RE, V5, P3190