Experimental and Theoretical Evaluation of a High Resolution CMOS Based Detector Under X-Ray Imaging Conditions

被引:50
作者
Michail, Christos M. [1 ]
Spyropoulou, Vasiliki A. [1 ]
Fountos, George P. [2 ]
Kalyvas, Nektarios I. [2 ]
Valais, Ioannis G. [2 ]
Kandarakis, Ioannis S. [2 ]
Panayiotakis, George S. [1 ]
机构
[1] Univ Patras, Sch Med, Dept Med Phys, Patras 26500, Greece
[2] Inst Educ Technol, Dept Med Instruments Technol, Athens 12210, Greece
关键词
CMOS; digital mammography; digital radiography; image quality; DIGITAL MAMMOGRAPHY; QUANTUM EFFICIENCY; PHYSICAL-CHARACTERIZATION; QUALITY MEASUREMENTS; PRESAMPLED MTF; PERFORMANCE; PHOSPHORS; SCREENS; IMAGER; NOISE;
D O I
10.1109/TNS.2010.2094206
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Fundamental imaging performance in terms of Modulation Transfer Function (MTF), Noise Power Spectrum (NPS) and Detective Quantum Efficiency (DQE) was investigated for a high resolution CMOS based imaging sensor. The device consists of a 33.91 mg/cm(2) Gd2O2S:Tb scintillator screen, placed in direct contact with a CMOS photodiode array. The CMOS photodiode array, featuring 1200 x 1600 pixels with a pixel pitch of 22.5 mu m, was used as an optical photon detector. In addition to the conventional frequency dependent parameters characterizing image quality, image information content was assessed through the application of information capacity (IC). The MTF was measured using the slanted-edge method to avoid aliasing while the Normalized NPS (NNPS) was determined by two-dimensional (2D) Fourier transforming of uniformly exposed images. Both measurements were performed under the representative radiation quality (RQA) settings, RQA-5 (70 kVp digital-radiography) and RQA-M2 (28 kVp digital-mammography) recommended by the International Electrotechnical Commission Reports 62220-1 and 62220-1-2 respectively. The DQE was assessed from the measured MTF, NPS and the direct entrance surface air-Kerma (ESAK) obtained from X-ray spectra measurement with a portable cadmium telluride (CdTe) detector. The ESAK values ranged between 11-87 mu Gy for RQA-5 and 6-40 mu Gy for RQA-M2. Additionally the output electrons per X-ray photon of the detector and its signal transfer characteristics were assessed via an analytical model, within the framework of the linear cascaded systems (LCS) theory. It was found that the detector response function was linear for the exposure ranges under investigation. Additionally our results showed that for the same RQA quality the output electrons per X-ray photon, as well as the measured and analytically predicted MTF, were not significantly affected by ESAK. MTF and DQE where found better compared to previously published data for other CCD and CMOS sensors, while the NNPS appeared to be comparable in the frequency range under investigation (0-10 cycles/mm).
引用
收藏
页码:314 / 322
页数:9
相关论文
共 58 条
  • [1] X-ray spectroscopy and dosimetry with a portable CdTe device
    Abbene, Leonardo
    La Manna, Angelo
    Fauci, Francesco
    Gerardi, Gaetano
    Stumbo, Simone
    Raso, Giuseppe
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) : 373 - 377
  • [2] Amptek, XRAY GAMM RAY DET XR
  • [3] [Anonymous], 61267 IEC
  • [4] Contrast cancellation technique applied to digital x-ray imaging using silicon strip detectors
    Avila, C
    Lopez, J
    Sanabria, JC
    Baldazzi, G
    Bollini, D
    Gombia, M
    Cabal, AE
    Ceballos, C
    Garcia, AD
    Gambaccini, M
    Taibi, A
    Sarnelli, A
    Tuffanelli, A
    Giubellino, P
    Marzari-Chiesa, A
    Prino, F
    Tomassi, E
    Grybos, P
    Idzik, M
    Swientek, K
    Wiacek, P
    Montaño, LM
    Ramello, L
    Sitta, M
    [J]. MEDICAL PHYSICS, 2005, 32 (12) : 3755 - 3766
  • [5] Review of CMOS image sensors
    Bigas, M
    Cabruja, E
    Forest, J
    Salvi, J
    [J]. MICROELECTRONICS JOURNAL, 2006, 37 (05) : 433 - 451
  • [6] Modeling the imaging performance of prototype organic x-ray imagers
    Blakesley, J. C.
    Speller, R.
    [J]. MEDICAL PHYSICS, 2008, 35 (01) : 225 - 239
  • [7] Determination of the presampled MTF in computed tomography
    Boone, JM
    [J]. MEDICAL PHYSICS, 2001, 28 (03) : 356 - 360
  • [8] Photon-counting spectral computed tomography using silicon strip detectors: a feasibility study
    Bornefalk, Hans
    Danielsson, Mats
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (07) : 1999 - 2022
  • [9] Measurements of X-ray imaging performance of granular phosphors with direct-coupled CMOS sensors
    Cho, Min Kook
    Kim, Ho Kyung
    Graeve, Thorsten
    Yum, Seung Man
    Lim, Chang Hwy
    Cho, Hyosung
    Kim, Jung-Min
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) : 1338 - 1343
  • [10] Principles of digital radiography with large-area, electronically readable detectors: A review of the basics
    Chotas, HG
    Dobbins, JT
    Ravin, CE
    [J]. RADIOLOGY, 1999, 210 (03) : 595 - 599