Implantation of a Novel Telemetric Intraocular Pressure Sensor in Patients With Glaucoma (ARGOS Study): 1-Year Results

被引:70
作者
Koutsonas, Antonis [1 ]
Walter, Peter [1 ]
Roessler, Gernot [1 ]
Plange, Niklas [1 ]
机构
[1] Rhein Westfal TH Aachen, Dept Ophthalmol, D-52074 Aachen, Germany
关键词
glaucoma; intraocular pressure sensor; self tonometry; PHACOEMULSIFICATION; PROGRESSION;
D O I
10.1167/iovs.14-14925
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. We investigated the safety of a telemetric IOP sensor and the accuracy of its IOP measurements in six patients with open-angle glaucoma and cataract. METHODS. The study design was a prospective, single-center clinical trial. Here we present 1year follow-up data. A ring-shaped telemetric IOP sensor was implanted in the ciliary sulcus after implantation of the intracapsular lens, during planned cataract surgery. The sensor is encapsulated in silicone rubber and consists of a miniature device with eight pressuresensitive capacitors and a circular microcoil antenna. IOP measurements are performed with a reader unit held in front of the eye. IOP is calculated as the differences between the absolute pressure inside the eye (pressure sensor) and that outside the eye (reader unit). RESULTS. The sensor was successfully implanted in all patients. Four patients developed sterile anterior chamber inflammation that resolved completely within 9 days after surgery with antiinflammatory treatment. All patients showed mild to moderate pupillary distortion and pigment dispersion after surgery. Telemetric IOP measurement was performed in all patients at all visits, and the patients successfully performed self-tonometry at home after receiving instructions. Telemetric IOP values showed similar profiles compared to those of Goldmann applanation tonometry (GAT). Three patients showed a relevant IOP step during follow-up, and in one patient, negative values were obtained throughout the study. CONCLUSIONS. Despite early postoperative anterior chamber inflammation, the IOP sensor was well tolerated by all patients. We describe the first prospective clinical study of a noncontact IOP sensor that potentially enables continuous IOP monitoring in patients with glaucoma. The sensor shape and size needs to be adapted to avoid pupillary distortion and to confirm that IOP measurements are accurately recorded in comparison to those of GAT.
引用
收藏
页码:1063 / 1069
页数:7
相关论文
共 24 条
[1]   Fluctuation of intraocular pressure and glaucoma progression in the early manifest glaucoma trial [J].
Bengtsson, Boel ;
Leske, M. Cristina ;
Hyman, Leslie ;
Heijl, Anders .
OPHTHALMOLOGY, 2007, 114 (02) :205-209
[2]   Biology of the corneal endothelium in health and disease [J].
Bourne, WM .
EYE, 2003, 17 (08) :912-918
[3]   Sequential office pressure measurements in the management of glaucoma [J].
Collaer, N ;
Zeyen, TY ;
Caprioli, J .
JOURNAL OF GLAUCOMA, 2005, 14 (03) :196-200
[5]   24-hour Intraocular Pressure Fluctuation Monitoring Using an Ocular Telemetry Sensor: Tolerability and Functionality in Healthy Subjects [J].
De Smedt, Stefan ;
Mermoud, Andre ;
Schnyder, Corinne .
JOURNAL OF GLAUCOMA, 2012, 21 (08) :539-544
[6]   Changes in intraocular pressure and anterior segment morphometry after uneventful phacoemulsification cataract surgery [J].
Dooley, I. ;
Charalampidou, S. ;
Malik, A. ;
Loughman, J. ;
Molloy, L. ;
Beatty, S. .
EYE, 2010, 24 (04) :520-526
[7]   Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach [J].
Doughty, MJ ;
Zaman, ML .
SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) :367-408
[8]   Reduction of intraocular pressure and glaucoma progression - Results from the early manifest glaucoma trial [J].
Heijl, A ;
Leske, MC ;
Bengtsson, B ;
Hyman, L ;
Bengtsson, B ;
Hussein, M .
ARCHIVES OF OPHTHALMOLOGY, 2002, 120 (10) :1268-1279
[9]   Advances in telemetric continuous intraocular pressure assessment [J].
Kakaday, T. ;
Hewitt, A. W. ;
Voelcker, N. H. ;
Li, J. S. J. ;
Craig, J. E. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (08) :992-996
[10]   A Comparison of Endothelial Cell Loss After Phacoemulsification in Penetrating Keratoplasty Patients and Normal Patients [J].
Kim, Eun Chul ;
Kim, Man Soo .
CORNEA, 2010, 29 (05) :510-515