Patent Foramen Ovale Closure Using a Bioabsorbable Closure Device Safety and Efficacy at 6-Month Follow-Up

被引:26
|
作者
Van den Branden, Ben J. [1 ]
Post, Martijn C. [1 ]
Plokker, Herbert W. [1 ]
ten Berg, Jurrien M. [1 ]
Suttorp, Maarten J. [1 ]
机构
[1] St Antonius Hosp, Dept Intervent Cardiol, NL-3435 CM Nieuwegein, Netherlands
关键词
closure; device; patent foramen ovale; transseptal catheter; SINGLE-CENTER EXPERIENCE; ATRIAL SEPTAL-DEFECT; TRANSCATHETER CLOSURE; PERCUTANEOUS CLOSURE; PARADOXICAL EMBOLISM; REPAIR IMPLANT; OCCLUDER; ECHOCARDIOGRAPHY; FEASIBILITY; PREVENTION;
D O I
10.1016/j.jcin.2010.06.012
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives The aim of this study was to assess the mid-term safety and efficacy of percutaneous patent foramen ovale (PFO) closure using a bioabsorbable device (BioSTAR, NMT Medical, Boston, Massachusetts). Background Closure of PFO in patients with cryptogenic stroke has proven to be safe and effective using different types of permanent devices. Methods All consecutive patients who underwent percutaneous PFO closure with the bioabsorbable closure device between November 2007 and January 2009 were included. Residual shunt was assessed using contrast transthoracic echocardiography. Results Sixty-two patients (55% women, mean age 47.7 +/- 11.8 years) underwent PFO closure. The in-hospital complications were a surgical device retrieval in 2 patients (3.2%), device reposition in 1 (1.6%), and a minimal groin hematoma in 6 patients (9.7%). The short-term complications at 1-month follow-up (n = 60) were a transient ischemic attack in the presence of a residual shunt in 1 patient and new supraventricular tachycardia in 7 patients (11.3%). At 6-month follow-up (n = 60), 1 patient without residual shunt developed a transient ischemic attack and 1 developed atrial fibrillation. A mild or moderate residual shunt was noted in 51.7%, 33.9%, and 23.7% after 1-day, 1-month, and 6-month follow-up, respectively. A large shunt was present in 8.3%, 3.4%, and 0% after 1-day, 1-month, and 6-month follow-up. Conclusions Closure of PFO using the bioabsorbable device is associated with a low complication rate and a low recurrence rate of embolic events. However, a relatively high percentage of mild or moderate residual shunting is still present at 6-month follow-up. (J Am Coll Cardiol Intv 2010;3: 968-73) (C) 2010 by the American College of Cardiology Foundation
引用
收藏
页码:968 / 973
页数:6
相关论文
共 50 条
  • [1] Follow-up of residual shunt after patent foramen ovale closure
    Orzan, F.
    Liboni, W.
    Bonzano, A.
    Molinari, F.
    Ribezzo, M.
    Rebaudengo, N.
    Grippi, G.
    Negri, E.
    ACTA NEUROLOGICA SCANDINAVICA, 2010, 122 (04): : 257 - 261
  • [2] Efficacy of patent foramen ovale closure for treating migraine: a prospective follow-up study
    Qi, Yi
    Zhang, Yushun
    Luo, Xiaohui
    Cheng, Gesheng
    Du, Yajuan
    Liu, Rui
    Xie, Hui
    Cheng, Yawen
    Guo, Yichen
    Luo, Guogang
    JOURNAL OF INVESTIGATIVE MEDICINE, 2021, 69 (01) : 7 - 12
  • [3] The BioSTAR® device versus the CardioSEAL® device in patent foramen ovale closure: comparison of mid-term efficacy and safety
    Van den Branden, Ben D.
    Luermans, Justin G.
    Post, Martijn C.
    Plokker, Herbert W.
    ten Berg, Jurrien M.
    Suttorp, Maarten J.
    EUROINTERVENTION, 2010, 6 (04) : 498 - 504
  • [4] Patent Foramen Ovale Stroke and Device Closure
    Suradi, Hussam S.
    Hijazi, Ziyad M.
    CARDIOLOGY CLINICS, 2016, 34 (02) : 231 - +
  • [5] Device closure of residual shunt after percutaneous closure of patent foramen ovale
    Majunke, Nicolas
    Wallenborn, Julia
    Baranowski, Andreas
    Wunderlich, Nina
    Sievert, Horst
    EUROINTERVENTION, 2010, 5 (07) : 833 - 837
  • [6] Echocardiographic Follow-Up of Patent Foramen Ovale and the Factors Affecting Spontaneous Closure
    Yildirim, Ali
    Aydin, Alperen
    Demir, Tevfik
    Aydin, Fatma
    Ucar, Birsen
    Kilic, Zubeyir
    ACTA CARDIOLOGICA SINICA, 2016, 32 (06) : 731 - 737
  • [7] Safety and Efficacy of Patent Foramen Ovale Device Closure: Evaluation of Outcomes in a Tertiary Referral Center
    Favaedi, Maryam
    Danesh-Moghadam, Maryam
    Khajali, Zahra
    Firouzi, Ata
    Parsaee, Mozhgan
    Saedi, Sedigheh
    RESEARCH IN CARDIOVASCULAR MEDICINE, 2023, 12 (02) : 52 - 55
  • [8] Percutaneous Closure of Patent Foramen Ovale with a Bioabsorbable Occluder Device: Single-Centre Experience
    Ussia, Gian Paolo
    Cammalleri, Valeria
    Mule, Massimiliano
    Scarabelli, Marilena
    Barbanti, Marco
    Scardaci, Francesco
    Mangiafico, Sarah
    Imme, Sebastiano
    Capodanno, Davide
    Tamburino, Corrado
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 74 (04) : 607 - 614
  • [9] Very long-term follow-up after percutaneous closure of patent foramen ovale
    Eeckhout, Eric
    Martin, Solenne
    Delabays, Alain
    Michel, Patrik
    Girod, Gregoire
    EUROINTERVENTION, 2015, 10 (12) : 1474 - 1479
  • [10] Patent foramen ovale closure with the BioSTAR® bioabsorbable implant
    Lopes, Joao Pedro
    Sousa, Lidia
    Fiarresga, Antonio
    Ferreira, Filipa
    Calrinho, Ana
    Martins, Jose Diogo
    Agapito, Ana
    Pinto, Fatima F.
    Ferreira, Rui
    REVISTA PORTUGUESA DE CARDIOLOGIA, 2010, 29 (11) : 1737 - 1742