High-resolution soft X-ray beamline ADRESS at the Swiss Light Source for resonant inelastic X-ray scattering and angle-resolved photoelectron spectroscopies

被引:295
作者
Strocov, V. N. [1 ]
Schmitt, T. [1 ]
Flechsig, U. [1 ]
Schmidt, T. [1 ]
Imhof, A. [1 ]
Chen, Q. [1 ]
Raabe, J. [1 ]
Betemps, R.
Zimoch, D.
Krempasky, J.
Wang, X. [1 ,2 ]
Grioni, M. [2 ]
Piazzalunga, A. [1 ,3 ]
Patthey, L. [1 ]
机构
[1] Paul Scherrer Inst, Swiss Light Source, CH-5232 Villigen, Switzerland
[2] Ecole Polytech Fedederale Lausanne, Inst Phys Matiere Condense, Lausanne, Switzerland
[3] Politecn Milan, Dipartimento Fis, I-20133 Milan, Italy
关键词
Swiss Light Source; fixed-gap undulator; soft X-ray beamline; X-ray optics; plane-grating monochromator; grating optimization; ellipsoidal refocusing optics; resonant inelastic X-ray scattering; angle-resolved photoelectron spectroscopy; ELLIPTICALLY-POLARIZING UNDULATOR; MONOCHROMATOR; PERFORMANCE;
D O I
10.1107/S0909049510019862
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The concepts and technical realisation of the high-resolution soft X-ray beamline ADRESS operating in the energy range from 300 to 1600 eV and intended for resonant inelastic X-ray scattering (R1XS) and angle-resolved photoelectron spectioscopy (ARPES) are described. The photon source is an undulator of novel fixed-gap design where longitudinal movement of permanent magnetic arrays controls not only the light polarization (including circular and 0-180 rotatable lineal polarizations) but also the energy without changing the gap The beamline optics is based on the well established scheme of planegrating monochromatoi operating in collimated light. The ultimate resolving power E/AE is above 33000 at 1 keV photon energy The choice of blazed versus lamellar gratings and optimization of their profile parameters is described. Owing to glancing angles on the minors as well as optimized groove densities and profiles of the gratings. the beamhne is capable of delivering high photon flux up to 1 x 10" photons (0.01% BW)-I at 1 keV. Ellipsoidal refocusing optics used for the R1XS endstation demagmfies the vertical spot size clown to 4 pm, which allows shtless operation and thus maximal transmission of the highresolution R1XS spectrometel delivering E/6,E > 11000 at 1 keV photon energy. Apart from the beamhne optics, an overview of the control system is given. the diagnostics and software tools are described, and strategies used for the optical alignment are discussed An introduction to the concepts and instrumental realisation of the ARMES and RIXS endstations is given
引用
收藏
页码:631 / 643
页数:13
相关论文
共 29 条
  • [1] BEAMLINE OPTIMIZATION AND PHASE-SPACE TRANSFORMATION
    BAHRDT, J
    FLECHSIG, U
    SENF, F
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (03) : 2719 - 2723
  • [2] ADJUSTABLE PHASE INSERTION DEVICES AS X-RAY SOURCES
    CARR, R
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 306 (1-2) : 391 - 396
  • [3] PERFORMANCE OF THE ELLIPTICALLY POLARIZING UNDULATOR ON SPEAR
    CARR, R
    KORTRIGHT, JB
    RICE, M
    LIDIA, S
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02) : 1862 - 1864
  • [4] PERFORMANCE OF THE DRAGON SOFT-X-RAY BEAMLINE
    CHEN, CT
    SETTE, F
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) : 1616 - 1621
  • [5] Extended SX-700 type monochromator combining normal and grazing incidence optics for a new undulator beamline at SLS
    Flechsig, U
    Patthey, L
    Quitmann, C
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 467 : 479 - 481
  • [6] Plane-grating monochromator at BESSY II using collimated light
    Follath, R
    Senf, F
    Gudat, W
    [J]. JOURNAL OF SYNCHROTRON RADIATION, 1998, 5 : 769 - 771
  • [7] New plane-grating monochromators for third generation synchrotron radiation light sources
    Follath, R
    Senf, F
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 390 (03) : 388 - 394
  • [8] SAXES, a high resolution spectrometer for resonant x-ray emission in the 400-1600 eV energy range
    Ghiringhelli, G.
    Piazzalunga, A.
    Dallera, C.
    Trezzi, G.
    Braicovich, L.
    Schmitt, T.
    Strocov, V. N.
    Betemps, R.
    Patthey, L.
    Wang, X.
    Grioni, M.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (11)
  • [9] Observation of Two Nondispersive Magnetic Excitations in NiO by Resonant Inelastic Soft-X-Ray Scattering
    Ghiringhelli, G.
    Piazzalunga, A.
    Dallera, C.
    Schmitt, T.
    Strocov, V. N.
    Schlappa, J.
    Patthey, L.
    Wang, X.
    Berger, H.
    Grioni, M.
    [J]. PHYSICAL REVIEW LETTERS, 2009, 102 (02)
  • [10] JANSA G, 2007, P 6 INT WORKSH PERS, P26101