Comparison of underivatized silica and zwitterionic sulfobetaine hydrophilic interaction liquid chromatography stationary phases for global metabolomics of human plasma

被引:13
|
作者
Sonnenberg, Rosalynde Ann [1 ]
Naz, Shama [1 ]
Cougnaud, Lise [1 ]
Vuckovic, Dajana [1 ]
机构
[1] Concordia Univ, Dept Chem & Biochem, 7141 Sherbrooke St West, Montreal, PQ H4B 1R6, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
HILIC; Global metabolomics; Sulfobetaine stationary phase; Silica stationary phase; Human plasma; Mass spectrometry; MASS-SPECTROMETRY; SELECTIVITY; SEPARATION; RETENTION; MS; QUANTITATION; STRATEGIES; BEHAVIOR;
D O I
10.1016/j.chroma.2019.460419
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
To increase metabolome coverage in global LC-MS metabolomics, often both reversed-phase liquid chromatography (RPLC) and hydrophilic-interaction liquid chromatography (HILIC) are implemented in parallel. However, there is a lack of consensus in the literature on the best HILIC stationary phase to employ for global metabolomics of human biological fluids. The objective of this study was to compare in detail the performance of two commonly employed HILIC phases: zwitterionic sulfobetaine ZIC-HILIC stationary phase and an underivatized silica HILIC stationary phase. During method development, the effect of salt concentration in the mobile phase was also investigated, and 5 mM ammonium acetate was selected. The stationary phases were evaluated using a mixture of 37 polar standards covering a range of logP values (-10 to 3.73), molecular weights (59-776 Da), charges (15 anions, 11 cations, and 11 neutral) as well as 17 lipid standards to understand phospholipid behaviour on the two stationary phases. The criteria used for the comparison included the quality of the chromatographic peak shape, adequate analyte retention, peak separation capability, and metabolite coverage. The zwitterionic ZIC-HILIC column provided better chromatographic performance over the silica stationary phase with 14 standards achieving good quality peaks compared to the 7 with the silica column. Only 2 standards were undetected with the ZIC-HILIC column compared to the 14 undetected with the silica column. In human plasma, 1966 and 1650 metabolites were observed on the ZIC-HILIC column in positive and negative electrospray ionization (ESI) respectively. On the silica HILIC column, 1773 and 2028 metabolites were observed in positive and negative ESI respectively, showing comparable performance of the two phases. Next, the effect of adding 10 mM ammonium phosphate to the samples to improve the analyte peak shape and metabolite coverage was investigated for both ZIC-HILIC and silica HILIC. In contrast with recently reported results for pZIC-HILIC, there was no clear evidence that ammonium phosphate addition was beneficial for human plasma samples. In conclusion, ZIC-HILIC provided better chromatographic performance for polar plasma metabolomics than underivatized silica in terms of chromatographic peak shape and chromatographic resolution, while maintaining comparable metabolite coverage. The addition of ammonium phosphate to human plasma was not beneficial for either of the two stationary phases. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Cyclofructan 6 based stationary phases for hydrophilic interaction liquid chromatography
    Qiu, Haixiao
    Loukotkova, Lucie
    Sun, Ping
    Tesarova, Eva
    Bosakova, Zuzana
    Armstrong, Daniel W.
    JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (02) : 270 - 279
  • [22] Dipeptide-bonded stationary phases for hydrophilic interaction liquid chromatography
    Skoczylas, Magdalena
    Bocian, Szymon
    Buszewski, Boguslaw
    RSC ADVANCES, 2016, 6 (98): : 96389 - 96397
  • [23] Study of the retention behavior of iodinated X-ray contrast agents in hydrophilic interaction liquid chromatography, comparing bare silica and zwitterionic stationary phases
    Salas, Daniela
    Borrull, Francesc
    Maria Marce, Rosa
    Fontanals, Nuria
    JOURNAL OF SEPARATION SCIENCE, 2014, 37 (9-10) : 1111 - 1117
  • [24] WATER UPTAKE ON SILICA-BASED STATIONARY PHASES IN HYDROPHILIC INTERACTION CHROMATOGRAPHY
    Soukup, Jan
    Jandera, Pavel
    CECE 2014: 11TH INTERNATIONAL INTERDISCIPLINARY MEETING ON BIOANALYSIS, 2014, : 106 - 109
  • [25] A comparison of hydrophilic interaction liquid chromatography and capillary electrophoresis for the metabolomics analysis of human serum
    Narduzzi, Luca
    Delgado-Povedano, Maria del Mar
    Lara, Francisco J.
    Le Bizec, Bruno
    Garcia-Campana, Ana Maria
    Dervilly, Gaud
    Hernandez-Mesa, Maykel
    JOURNAL OF CHROMATOGRAPHY A, 2023, 1706
  • [26] Development and evaluation of new imidazolium-based zwitterionic stationary phases for hydrophilic interaction chromatography
    Qiao, Lizhen
    Dou, Abo
    Shi, Xianzhe
    Li, Hua
    Shan, Yuanhong
    Lu, Xin
    Xu, Guowang
    JOURNAL OF CHROMATOGRAPHY A, 2013, 1286 : 137 - 145
  • [27] Investigation of the temperature dependence of water adsorption on silica-based stationary phases in hydrophilic interaction liquid chromatography
    Barto, Endre
    Felinger, Attila
    Jandera, Pavel
    JOURNAL OF CHROMATOGRAPHY A, 2017, 1489 : 143 - 148
  • [28] Modelling of Hydrophilic Interaction Liquid Chromatography Stationary Phases Using Chemometric Approaches
    Navarro-Reig, Meritxell
    Ortiz-Villanueva, Elena
    Tauler, Roma
    Jaumot, Joaquim
    METABOLITES, 2017, 7 (04):
  • [29] Evaluation of hydrophilic interaction liquid chromatography stationary phases for analysis of opium alkaloids
    Bagheri, Mohsen
    Taheri, Mohammadreza
    Farhadpour, Mohsen
    Rezadoost, Hassan
    Ghassempour, Alireza
    Aboul-Enein, Hassan Y.
    JOURNAL OF CHROMATOGRAPHY A, 2017, 1511 : 77 - 84
  • [30] Evaluation of thermally pretreated silica stationary phases under hydrophilic interaction chromatography conditions
    Mignot, Melanie
    Periat, Aurelie
    Peulon-Agasse, Valerie
    Cardinael, Pascal
    Veuthey, Jean-Luc
    Guillarme, Davy
    JOURNAL OF SEPARATION SCIENCE, 2016, 39 (09) : 1611 - 1618