Numerical analysis of displacements of mandible bone parts using various elements for fixation of subcondylar fractures

被引:0
作者
Kromka-Szydek, Magdalena [1 ]
Jedrusik-Pawlowska, Magdalena [2 ]
Milewski, Grzegorz [1 ]
Lekston, Zdzislaw [3 ]
Cieslik, Tadeusz [2 ]
Drugacz, Jan [2 ]
机构
[1] Cracow Univ Technol, Dept Expt Mech & Biomech, Inst Appl Mech, PL-31864 Krakow, Poland
[2] Med Univ Silesia, Dept Craniomaxillofacial Surg, Katowice, Poland
[3] Univ Silesia, Inst Mat Sci, Katowice, Poland
关键词
mandible condylar process fracture; plate osteosynthesis; NiTi shape-memory staples; MES analysis; CONDYLAR PROCESS; BIOMECHANICAL EVALUATION; CLOSED TREATMENT;
D O I
暂无
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The paper presents an analysis of the distribution of dislocations in the case of fixation of a mandible low subcondylar fracture. Three types of elements have been used for fixation: NiTi shape-memory staples, miniaturized titanium plates and the Synthes compression plate, also called the Synthes zygomatic plate. The analysis was conducted using the finite element method. The degree of the mobility of the fractures was analysed as well as the tendency towards their separation on the basis of value analysis and the distribution of dislocation areas. The results obtained allowed a preliminary assessment of predicted healing effects and the possibility of being able to predict developing complications after osteosynthesis with the use of the elements analysed.
引用
收藏
页码:11 / 18
页数:8
相关论文
共 16 条
  • [1] CIESLIK T, 2002, ENG BIOMATERIALS, V5, P23
  • [2] CIESLIK T, 1998, CZAS STOM, V51, P349
  • [3] CIESLIK T, 2002, ENG BIOMATERIALS, V5, P51
  • [4] Open versus closed treatment of fractures of the mandibular condylar process - a prospective randomized multi-centre study
    Eckelt, Uwe
    Schneider, Matthias
    Erasmus, Francois
    Gerlach, Klaus Louis
    Kuhlisch, Eberhard
    Loukota, Richard
    Rasse, Michael
    Schubert, Johannes
    Terheyden, Hendrik
    [J]. JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2006, 34 (05) : 306 - 314
  • [5] Surgical complications with open treatment of mandibular condylar process fractures
    Ellis, E
    McFadden, D
    Simon, P
    Throckmorton, G
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2000, 58 (09) : 950 - 958
  • [6] Occlusal results after open or closed treatment of fractures of the mandibular condylar process
    Ellis, E
    Simon, P
    Throckmorton, GS
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2000, 58 (03) : 260 - 268
  • [7] Harle F., 1999, ATLAS CRANIOMAXILLOF
  • [8] A biomechanical evaluation of mandibular condyle fracture plating techniques
    Haug, RH
    Peterson, GP
    Goltz, M
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2002, 60 (01) : 73 - 80
  • [9] JEDRUSIKPAWLOWS.M, 2007, DENT MED PROB, V44, P449
  • [10] Knets I., 1996, P 10 C EUR SOC BIOM, P97