Development of integrated prompt gamma imaging and positron emission tomography system for in vivo 3-D dose verification: a Monte Carlo study

被引:14
作者
Choi, Hyun Joon [1 ]
Jang, Ji Won [1 ]
Shin, Wook-Geun [1 ]
Park, Hyojun [1 ]
Incerti, Sebastien [2 ,3 ]
Min, Chul Hee [1 ]
机构
[1] Yonsei Univ, Dept Radiat Convergence Engn, Wonju 26493, South Korea
[2] Bordeaux Univ, UMR 5797, F-33170 Gradignan, France
[3] CNRS, IN2P3, CENBG, UMR 5797, F-33170 Gradignan, France
基金
新加坡国家研究基金会;
关键词
proton therapy; prompt gamma imaging; positron emission tomography; in vivo dose verification; Monte Carlo; CONVOLUTIONAL NEURAL-NETWORK; RANGE VERIFICATION; PROTON-BEAMS; THERAPY; PATIENT; CT; SEGMENTATION; VALIDATION; PREDICTION;
D O I
10.1088/1361-6560/ab857c
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An accurate knowledge of in vivo proton dose distribution is key to fully utilizing the potential advantages of proton therapy. Two representative indirect methods for in vivo range verification, namely, prompt gamma (PG) imaging and positron emission tomography (PET), are available. This study proposes a PG-PET system that combines the advantages of these two methods and presents detector geometry and background reduction techniques optimized for the PG-PET system. The characteristics of the secondary radiations emitted by a water phantom by interaction with a 150 MeV proton beam were analysed using Geant4.10.00, and the 2-D PG distributions were obtained and assessed for different detector geometries. In addition, the energy window (EW), depth-of-interaction (DOI), and time-of-flight (TOF) techniques are proposed as the background reduction techniques. To evaluate the performance of the PG-PET system, the 3-D dose distribution in the water phantom caused by two proton beams of energies 80 MeV and 100 MeV was verified using 16 optimal detectors. The thickness of the parallel-hole tungsten collimator of pitch 8 mm and width 7 mm was determined as 200 mm, and that of the GAGG scintillator was determined as 30 mm, by an optimization study. Further, 3-7 MeV and 2-7 MeV were obtained as the optimal EWs when the DOI and both the DOI and TOF techniques were applied for data processing, respectively; the detector performances were improved by about 38% and 167%, respectively, compared with that when applying only the 3-5 MeV EW. In this study, we confirmed that the PG distribution can be obtained by simply combining the 2-D parallel hole collimator and the PET detector module. In the future, we will develop an accurate 3-D dose evaluation technique using deep learning algorithms based on the image sets of dose, PG, and PET distributions for various proton energies.
引用
收藏
页数:16
相关论文
共 55 条
  • [1] The influence of the optimization starting conditions on the robustness of intensity-modulated proton therapy plans
    Albertini, F.
    Hug, E. B.
    Lomax, A. J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (10) : 2863 - 2878
  • [2] [Anonymous], 2017, ARXIV
  • [3] Discriminative Prior - Prior Image Constrained Compressed Sensing Reconstruction for Low-Dose CT Imaging
    Chen, Yang
    Liu, Jin
    Xie, Lizhe
    Hu, Yining
    Shu, Huazhong
    Luo, Limin
    Zhang, Libo
    Gui, Zhiguo
    Coatrieux, Gouenou
    [J]. SCIENTIFIC REPORTS, 2017, 7
  • [4] Development of a Geant4-based independent patient dose validation system with an elaborate multileaf collimator simulation model
    Choi, Hyun Joon
    Park, Hyojun
    Shin, Wook-Geun
    Kim, Jung-in
    Min, Chul Hee
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (02): : 94 - 106
  • [5] A feasibility study for predicting optimal radiation therapy dose distributions of prostate cancer patients from patient anatomy using deep learning
    Dan Nguyen
    Long, Troy
    Jia, Xun
    Lu, Weiguo
    Gu, Xuejun
    Iqbal, Zohaib
    Jiang, Steve
    [J]. SCIENTIFIC REPORTS, 2019, 9 (1)
  • [6] Status report on silicon photomultiplier development and its applications
    Dolgoshein, B.
    Balagura, V.
    Buzhan, P.
    Danilov, M.
    Filatov, L.
    Garutti, E.
    Groll, M.
    Ilyin, A.
    Kantserov, V.
    Kaplin, V.
    Karakash, A.
    Kayumov, F.
    Klemin, S.
    Korbel, V.
    Meyer, H.
    Mizuk, R.
    Morgunov, V.
    Novikov, E.
    Pakhlov, P.
    Popova, E.
    Rusinov, V.
    Sefkow, F.
    Tarkovsky, E.
    Tikhomirov, I.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 563 (02) : 368 - 376
  • [7] Charged hadron tumour therapy monitoring by means of PET
    Enghardt, W
    Crespo, P
    Fiedler, F
    Hinz, R
    Parodi, K
    Pawelke, J
    Pönisch, F
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 525 (1-2) : 284 - 288
  • [8] GAMMA-RADIATION IMAGING-SYSTEM BASED ON COMPTON-EFFECT
    EVERETT, DB
    FLEMING, JS
    TODD, RW
    NIGHTINGALE, JM
    [J]. PROCEEDINGS OF THE INSTITUTION OF ELECTRICAL ENGINEERS-LONDON, 1977, 124 (11): : 995 - 1000
  • [9] Esophagus segmentation in CT via 3D fully convolutional neural network and random walk
    Fechter, Tobias
    Adebahr, Sonja
    Baltas, Dimos
    Ben Ayed, Ismail
    Desrosiers, Christian
    Dolz, Jose
    [J]. MEDICAL PHYSICS, 2017, 44 (12) : 6341 - 6352
  • [10] Online proton therapy monitoring: clinical test of a Silicon-photodetector-based in-beam PET
    Ferrero, Veronica
    Fiorina, Elisa
    Morrocchi, Matteo
    Pennazio, Francesco
    Baroni, Guido
    Battistoni, Giuseppe
    Belcari, Nicola
    Camarlinghi, Niccolo
    Ciocca, Mario
    Del Guerra, Alberto
    Donetti, Marco
    Giordanengo, Simona
    Giraudo, Giuseppe
    Patera, Vincenzo
    Peroni, Cristiana
    Rivetti, Angelo
    Rolo, Manuel Dionisio da Rocha
    Rossi, Sandro
    Rosso, Valeria
    Sportelli, Giancarlo
    Tampellini, Sara
    Valvo, Francesca
    Wheadon, Richard
    Cerello, Piergiorgio
    Bisogni, Maria Giuseppina
    [J]. SCIENTIFIC REPORTS, 2018, 8