Monte Carlo derivation of AAPM TG-43 dosimetric parameters for GZP6 Co-60 HDR sources

被引:5
|
作者
Tabrizi, Sanaz Hariri [1 ]
Asl, Alireza Kamali [1 ]
Azma, Zohreh [1 ]
机构
[1] Shahid Beheshti Univ, Radiat Med Engn Dept, Tehran, Iran
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2012年 / 28卷 / 02期
关键词
HDR brachytherapy; Monte Carlo; TG-43; protocol; Treatment planning software; BRACHYTHERAPY SOURCES; TECHNICAL NOTE; IR-192;
D O I
10.1016/j.ejmp.2011.04.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Cobalt 60 source is generally available on high dose rate (HDR) afterloading equipment especially for treatment of gynecological lesions. The GZP6 remote afterloader (Nuclear Power Institute of China) utilizes Co-60 sources for treatment of intracavitary and intraluminal malignancies. In this study, the AAPM TG-43 dosimetric parameters of three sources in GZP6 system have been studied using MCNP4C Monte Carlo (MC) code; and the results are compared with other available Co-60 HDR sources. The presented parameters consist of air kerma strength, dose rate constant, radial dose function and anisotropy function. They show less than 1% uncertainty. The TG-43 based dosimetry data can be used not only to validate the dedicated treatment planning software (TPS), but also to introduce new complementary software to enhance the system performance in gynecological treatments. (C) 2011 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:153 / 160
页数:8
相关论文
共 25 条
  • [1] Impact of magnetic fields on calculated AAPM TG-43 parameters for 192Ir and 60Co HDR brachytherapy sources: A Monte Carlo study
    Ranjbar, Sahar
    Gholami, Somayeh
    Longo, Francesco
    Sharp, Ryan
    Meigooni, Ali S.
    APPLIED RADIATION AND ISOTOPES, 2020, 159
  • [2] A Monte Carlo study on dose distribution validation of GZP6 Co-60 stepping source
    Toossi, Mohammad Taghi Bahreyni
    Abdollahi, Maliheh
    Ghorbani, Mahdi
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2013, 18 (02) : 112 - 116
  • [3] Monte Carlo calculation of the TG-43 dosimetric parameters of a new BEBIG Ir-192HDR source
    Granero, D
    Pérez-Calatayud, J
    Ballester, F
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 (01) : 79 - 85
  • [4] Experimental and Monte Carlo determination of the TG-43 dosimetric parameters for the model 9011 THINSeed™ brachytherapy source
    Kennedy, R. M.
    Davis, S. D.
    Micka, J. A.
    DeWerd, L. A.
    MEDICAL PHYSICS, 2010, 37 (04) : 1681 - 1688
  • [5] An EGSnrc Monte Carlo-calculated database of TG-43 parameters
    Taylor, R. E. P.
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2008, 35 (09) : 4228 - 4241
  • [6] Monte Carlo calculation of the TG-43 dosimetry parameters for the INTRABEAM source with spherical applicators
    Alvarez, David Santiago Ayala
    Watson, Peter G. F.
    Popovic, Marija
    Heng, Veng Jean
    Evans, Michael D. C.
    Seuntjens, Jan
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (21)
  • [7] Monte Carlo dosimetry of the 60Co sources of a new GZP3 HDR afterloading system
    Junxiang Wu
    Shengwei Kang
    Pei Wang
    Jie Li
    Xianliang Wang
    Radiation Oncology, 18
  • [8] Monte Carlo dosimetry of the 60Co sources of a new GZP3 HDR afterloading system
    Wu, Junxiang
    Kang, Shengwei
    Wang, Pei
    Li, Jie
    Wang, Xianliang
    RADIATION ONCOLOGY, 2023, 18 (01)
  • [9] Radial dose functions of GZP6 intracavitary brachytherapy 60Co sources: treatment planning system versus Monte Carlo calculations
    Mesbahi, A.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2008, 5 (04): : 181 - 186
  • [10] Monte Carlo and experimental derivation of TG43 dosimetric parameters for CSM-type Cs-137 sources
    Pérez-Calatayud, J
    Granero, D
    Casal, E
    Ballester, F
    Puchades, V
    MEDICAL PHYSICS, 2005, 32 (01) : 28 - 36