AMPEROMETRIC CARBON PASTE ENZYME ELECTRODES FOR URIC ACID DETERMINATION WITH DIFFERENT MEDIATORS

被引:3
|
作者
Erden, Pinar Esra [1 ]
Pekyardimci, Sule [1 ]
Kilic, Esma [1 ]
机构
[1] Ankara Univ, Dept Chem, Fac Sci, TR-06100 Ankara, Turkey
关键词
Amperometry; 1,4-Benzoquinone; Carbon paste; Enzyme electrode; Poly(vinylferrocene); Uric acid; Biosensors; Carbon; Enzymes; VOLTAMMETRIC DETECTION; INJECTION-ANALYSIS; ASCORBIC-ACID; BIOSENSOR; SERUM; PERFORMANCE; SYSTEM;
D O I
10.1135/cccc2011049
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
In this study, two new amperometric carbon paste enzyme electrodes for determination of uric acid were developed. The carbon paste was prepared by mixing uricase enzyme, 1,4-benzoquinone or poly(vinylferrocene) (PVF) as a mediator, graphite powder, paraffin oil and then the paste was placed into cavity of a teflon electrode body. Determination of uric acid was performed by oxidation of enzymatically generated H2O2. The effects of enzyme loading, mediator amount, buffer type, pH, buffer concentration, working potential and temperature were investigated for both electrodes. The working range of the 1,4-benzoquinone modified enzyme electrode was 1.9 x 10(-8)-2.7 x 10(-3) M, detection limit 1.9 x 10(-8) M and response time 150 s. Optimum buffer type, pH, buffer concentration, working potential, temperature and amounts of enzyme and mediator for 1,4-benzoquinone modified enzyme electrode were found to be Tris, 8.0, 0.20 M, +0.25 V, 30 degrees C, 2.0 Unit and 13%, respectively. The working range of the PVF modified enzyme electrode was 7.4 x 10(-8)-7.0 x 10(-3) M, detection limit 7.4 x 10(-8) M and response time 120 s. Optimum buffer type, pH, buffer concentration, working potential, temperature and amounts of enzyme and mediator for PVF modified enzyme electrode were found to be phosphate, 8.0, 0.05 M, +0.70 and +0.30 V, 40 degrees C, 2.0 Unit and 10.9%, respectively. The repeatability, storage stability of the enzyme electrodes and interference effects were also investigated. Enzyme electrodes were used for determination of uric acid in serum samples and the results were in a good agreement with those obtained by commercial enzymatic kits.
引用
收藏
页码:1055 / 1073
页数:19
相关论文
共 50 条
  • [1] Amperometric Enzyme Electrodes for Xanthine Determination with Different Mediators
    Erden, Pinar Esra
    Pekyardimci, Sule
    Kilic, Esma
    ACTA CHIMICA SLOVENICA, 2012, 59 (04) : 824 - 832
  • [2] Amperometric determination of rutin on carbon paste electrodes
    Macikova, Pavla
    Hrbac, Jan
    Halouzka, Vladimir
    Skopalova, Jana
    JOURNAL OF BIOCHEMICAL TECHNOLOGY, 2010, 2 (05) : S20 - S21
  • [3] AMPEROMETRIC DETERMINATION OF RUTIN ON CARBON PASTE ELECTRODES
    Macikova, Pavla
    Hrbac, Jan
    Halouzka, Vladimir
    Skopalova, Jana
    X PRACOVNI SETKANI FYZIKALNICH CHEMIKU A ELECTROCHEMIKU/ 10TH WORKSHOP OF PHYSICAL CHEMISTS AND ELECTROCHEMISTS, 2010, : 147 - 149
  • [4] IMMOBILIZED ENZYME CARBON PASTE ELECTRODES AS AMPEROMETRIC SENSORS
    LORENZO, E
    GONZALEZ, E
    PARIENTE, F
    HERNANDEZ, L
    ELECTROANALYSIS, 1991, 3 (4-5) : 319 - 323
  • [5] Evaluation of redox mediators for amperometric biosensors: Ru-complex modified carbon-paste/enzyme electrodes
    Ivanova, EV
    Sergeeva, VS
    Oni, J
    Kurzawa, C
    Ryabov, AD
    Schuhmann, W
    BIOELECTROCHEMISTRY, 2003, 60 (1-2) : 65 - 71
  • [6] THE DETERMINATION OF GLUCOSE, HYPOXANTHINE AND URIC-ACID WITH USE OF BI-ENZYME AMPEROMETRIC ELECTRODES
    KULYS, JJ
    LAURINAVICIUS, VSA
    PESLIAKIENE, MV
    GUREVICIENE, VV
    ANALYTICA CHIMICA ACTA, 1983, 148 (APR) : 13 - 18
  • [7] VOLTAMMETRIC AND AMPEROMETRIC DETERMINATION OF DOXORUBICIN USING CARBON PASTE ELECTRODES
    Jemelkova, Zuzana
    Zima, Jiri
    Barek, Jiri
    COLLECTION OF CZECHOSLOVAK CHEMICAL COMMUNICATIONS, 2009, 74 (10) : 1503 - 1515
  • [8] Acid stability of carbon paste enzyme electrodes
    Wang, Joseph
    Musameh, Mustafa
    Mo, Jian-Wei
    ANALYTICAL CHEMISTRY, 2006, 78 (19) : 7044 - 7047
  • [9] A new amperometric carbon paste enzyme electrode for ethanol determination
    Koyuncu, Derya
    Erden, Pinar E.
    Pekyardimci, Sule
    Kilic, Esma
    ANALYTICAL LETTERS, 2007, 40 (10) : 1904 - 1922