Intra-rater repeatability of gait parameters in healthy adults during self-paced treadmill-based virtual reality walking

被引:19
作者
Al-Amri, Mohammad [1 ,2 ]
Al Balushi, Hilal [1 ]
Mashabi, Abdulrhman [1 ]
机构
[1] Cardiff Univ, Sch Healthcare Sci, Coll Biomed & Life Sci, Cardiff, S Glam, Wales
[2] Cardiff Univ, Arthrit Res UK Biomech & Bioengn Ctr, Cardiff, S Glam, Wales
关键词
Reliability; instrumented treadmill; gait analysis; GRAIL; self-paced walking; OVERGROUND WALKING; SPORTS-MEDICINE; RELIABILITY; YOUNG; SYSTEM; VARIABILITY; KINEMATICS; LOCOMOTION; AGREEMENT; BLAND;
D O I
10.1080/10255842.2017.1404994
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Self-paced treadmill walking is becoming increasingly popular for the gait assessment and re-education, in both research and clinical settings. Its day-to-day repeatability is yet to be established. This study scrutinised the test-retest repeatability of key gait parameters, obtained from the Gait Real-time Analysis Interactive Lab (GRAIL) system. Twenty-three male able-bodied adults (age: 34.56 +/- 5.12years) completed two separate gait assessments on the GRAIL system, separated by 5 +/- 3days. Key gait kinematic, kinetic, and spatial-temporal parameters were analysed. The Intraclass-Correlation Coefficients (ICC), Standard Error Measurement (SEM), Minimum Detectable Change (MDC), and the 95% limits of agreements were calculated to evaluate the repeatability of these gait parameters. Day-to-day agreements were excellent (ICCs>0.87) for spatial-temporal parameters with low MDC and SEM values, <0.153 and<0.055, respectively. The repeatability was higher for joint kinetic than kinematic parameters, as reflected in small values of SEM (<0.13 Nm/kg and<3.4 degrees) and MDC (<0.335 Nm/kg and<9.44 degrees). The obtained values of all parameters fell within the 95% limits of agreement. Our findings demonstrate the repeatability of the GRAIL system available in our laboratory. The SEM and MDC values can be used to assist researchers and clinicians to distinguish real' changes in gait performance over time.
引用
收藏
页码:1669 / 1677
页数:9
相关论文
共 35 条
[1]   Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine [J].
Atkinson, G ;
Nevill, AM .
SPORTS MEDICINE, 1998, 26 (04) :217-238
[2]   Gait analysis methods in rehabilitation [J].
Baker, Richard .
JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2006, 3 (1)
[3]   A treadmill ergometer for three-dimensional ground reaction forces measurement during walking [J].
Belli, A ;
Bui, P ;
Berger, A ;
Geyssant, A ;
Lacour, JR .
JOURNAL OF BIOMECHANICS, 2001, 34 (01) :105-112
[4]  
Bruton A., 2000, PHYSIOTHERAPY, V86, P94, DOI [DOI 10.1016/S0031-9406(05)61211-4, 10.1016/S0031-9406(05)61211-4]
[5]   Correlation, Agreement, and Bland-Altman Analysis: Statistical Analysis of Method Comparison Studies [J].
Bunce, Catey .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 148 (01) :4-6
[6]   The reliability of kinetic and kinematic variables used to analyse normal running gait [J].
Diss, CE .
GAIT & POSTURE, 2001, 14 (02) :98-103
[7]   Reliability of gait parameters during treadmill walking in community- dwelling healthy seniors [J].
Faude, Oliver ;
Donath, Lars ;
Roth, Ralf ;
Fricker, Livia ;
Zahner, Lukas .
GAIT & POSTURE, 2012, 36 (03) :444-448
[8]  
Geijtenbeek T, 2011, P 10 INT C VIRT REAL
[9]   Can treadmill walking be used to assess propulsion generation? [J].
Goldberg, Evan J. ;
Kautz, Steven A. ;
Neptune, Richard R. .
JOURNAL OF BIOMECHANICS, 2008, 41 (08) :1805-1808
[10]   Interpreting change scores of tests and measures used in physical therapy [J].
Haley, SM ;
Fragala-Pinkham, MA .
PHYSICAL THERAPY, 2006, 86 (05) :735-743