Robot-Assisted Versus Fluoroscopy-Assisted Cortical Bone Trajectory Screw Instrumentation in Lumbar Spinal Surgery: A Matched-Cohort Comparison

被引:91
作者
Le, Xiaofeng [1 ]
Tian, Wei [1 ]
Shi, Zhan [1 ]
Han, Xiaoguang [1 ]
Liu, Yajun [1 ]
Liu, Bo [1 ]
He, Da [1 ]
Yuan, Qiang [1 ]
Sun, Yuqing [1 ]
Xu, Yunfeng [1 ]
机构
[1] Beijing Jishuitan Hosp, Dept Spine Surg, Beijing, Peoples R China
基金
国家高技术研究发展计划(863计划);
关键词
Computer-assisted surgery; Cortical bone trajectory; Robotic surgery; Spine instrumentation; PEDICLE SCREW; INTERBODY FUSION; FIXATION; PLACEMENT; ACCURACY; SEGMENT;
D O I
10.1016/j.wneu.2018.08.157
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: The aim of the present study was to compare the accuracy and safety of TiRobot system-assisted with those of fluoroscopy-assisted cortical bone trajectory screw placement in lumbar spinal surgery. METHODS: We included 58 patients who required instrumentation in a retrospective matched-cohort study. The primary outcome measures were the accuracy of screw placement using the modified Gertzbein-Robbins scale and the incidence of proximal facet joint violation. Secondary outcome measures included operative time, intraoperative blood loss, postoperative hospital stay, radiation exposure, and complications. RESULTS: A total of 231 screws were placed (TiRobot group [RG], 86 screws; fluoroscopy group [FG], 145 screws). In the RG, 87.2% of the screws had perfect positions (grade A). The remaining screws were grade B (8.1%) and C (4.7%). In the FG, 66.9% of the screws had perfect positions (grade A). The remaining screws were grade B (20.0%), C (9.0%), and D (4.1%). The proportion of clinically acceptable screws (grade A or B) was greater in the RG than in the FG. In the RG, the most common direction of screw deviation was cephalad endplate and was the vertebral cortex in FG. No difference was found in facet joint violation between the 2 groups. The operative time and blood loss were slightly greater in the RG than in the FG. No statistically significant difference was found in the postoperative hospital stay between the 2 groups. The mean cumulative radiation time was greater for the RG than for the FG, but the radiation exposure to the surgeon was significantly lower in the RG than in the FG. CONCLUSIONS: Robot-assisted screw placement is more accurate and safe compared with fluoroscopy-assisted placement for lumbar spinal cortical bone trajectory instrumentation.
引用
收藏
页码:E745 / E751
页数:7
相关论文
共 25 条
[1]   Effect of Physiological Loads on Cortical and Traditional Pedicle Screw Fixation [J].
Baluch, Daniel A. ;
Patel, Alpesh A. ;
Lullo, Brett ;
Havey, Robert M. ;
Voronov, Leonard I. ;
Ngoc-Lam Nguyen ;
Carandang, Gerard ;
Ghanayem, Alexander J. ;
Patwardhan, Avinash G. .
SPINE, 2014, 39 (22) :E1297-E1302
[2]  
Delgado-Fernandez J, 2017, ASIAN SPINE J, V11, P817, DOI 10.4184/asj.2017.11.5.817
[3]   ACCURACY OF PEDICULAR SCREW PLACEMENT INVIVO [J].
GERTZBEIN, SD ;
ROBBINS, SE .
SPINE, 1990, 15 (01) :11-14
[4]   Isthmus-guided Cortical Bone Trajectory for Pedicle Screw Insertion [J].
Iwatsuki, Koichi ;
Yoshimine, Toshiki ;
Ohnishi, Yu-ichiro ;
Ninomiya, Kosi ;
Ohkawa, Toshika .
ORTHOPAEDIC SURGERY, 2014, 6 (03) :244-248
[5]   Current applications of robotics in spine surgery: a systematic review of the literature [J].
Joseph, Jacob R. ;
Smith, Brandon W. ;
Liu, Xilin ;
Park, Paul .
NEUROSURGICAL FOCUS, 2017, 42 (05)
[6]   Short-Term Results of Transforaminal Lumbar Interbody Fusion Using Pedicle Screw with Cortical Bone Trajectory Compared with Conventional Trajectory [J].
Kasukawa, Yuji ;
Miyakoshi, Naohisa ;
Hongo, Michio ;
Ishikawa, Yoshinori ;
Kudo, Daisuke ;
Shimada, Yoichi .
ASIAN SPINE JOURNAL, 2015, 9 (03) :440-448
[7]   The Biomechanical Effect of Pedicle Screws' Insertion Angle and Position on the Superior Adjacent Segment in 1 Segment Lumbar Fusion [J].
Kim, Ho-Joong ;
Chun, Heoung-Jae ;
Kang, Kyoung-Tak ;
Moon, Seong-Hwan ;
Kim, Hak-Sun ;
Park, Jin-Oh ;
Moon, Eun-Su ;
Kim, Bo-Ram ;
Sohn, Joon-Seok ;
Ko, Yu-Na ;
Lee, Hwan-Mo .
SPINE, 2012, 37 (19) :1637-1644
[8]   Incidence and Risk Factors of Adjacent Cranial Facet Joint Violation Following Pedicle Screw Insertion Using Cortical Bone Trajectory Technique [J].
Matsukawa, Keitaro ;
Kato, Takashi ;
Yato, Yoshiyuki ;
Sasao, Hiroshi ;
Imabayashi, Hideaki ;
Hosogane, Naobumi ;
Asazuma, Takashi ;
Chiba, Kazuhiro .
SPINE, 2016, 41 (14) :E851-E856
[9]   In Vivo Analysis of Insertional Torque During Pedicle Screwing Using Cortical Bone Trajectory Technique [J].
Matsukawa, Keitaro ;
Yato, Yoshiyuki ;
Kato, Takashi ;
Imabayashi, Hideaki ;
Asazuma, Takashi ;
Nemoto, Koichi .
SPINE, 2014, 39 (04) :E240-E245
[10]   Morphometric Measurement of Cortical Bone Trajectory for Lumbar Pedicle Screw Insertion Using Computed Tomography [J].
Matsukawa, Keitaro ;
Yato, Yoshiyuki ;
Nemoto, Osamu ;
Imabayashi, Hideaki ;
Asazuma, Takashi ;
Nemoto, Koichi .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2013, 26 (06) :E248-E253