Present limitations of CVD diamond detectors for IMRT applications

被引:25
作者
De Angelis, C. [1 ]
Casati, M. [2 ]
Bruzzi, M. [3 ]
Onori, S. [1 ]
Bucciolini, M. [2 ]
机构
[1] Ist Super Sanita, INFN, Dipartimento Tecnol & Salute, I-00161 Rome, Italy
[2] Univ Florence, INFN, Dipartimento Fisiopatol, I-50121 Florence, Italy
[3] Univ Florence, INFN, Dipartimento Engn, I-50121 Florence, Italy
关键词
CVD diamond detector; solid-state dosimetry; IMRT dosimetry;
D O I
10.1016/j.nima.2007.08.186
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The aim of the work was to test the suitability of chemical vapor deposited (CVD) diamond detectors for dosimetry in IMRT fields. We used in-house CVD detectors prepared with state-of-the-art polycrystalline diamond films (Element Six Ltd., UK). The parameters considered were time stability, dynamic response, dose-rate dependence and energy dependence. Output factors and TPR were measured in conventional photon fields and dose measurements were performed in IMRT fields using the step-and-shoot technique. Results prove that CVD diamond detectors are suitable for dosimetry in conventional treatments, but they still do not fit the IMRT dosimetry requirements, mainly because of their slow dynamic response. In particular, the slow dynamics affects linearity at low Monitor Units and renders it impossible to follow the sharp transients of IMRT fields. Time stability and dose-rate dependence as well must be improved to reduce their influence on dose assessment. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:195 / 203
页数:9
相关论文
共 25 条
  • [1] Status of the R&D activity on diamond particle detectors
    Adam, W
    Bellini, B
    Berdermann, E
    Bergonzo, P
    de Boer, W
    Bogani, F
    Borchi, E
    Brambilla, A
    Bruzzi, M
    Colledani, C
    Conway, J
    D'Angelo, P
    Dabrowski, W
    Delpierre, P
    Doroshenko, J
    Dulinski, W
    van Eijk, B
    Fallou, A
    Fischer, P
    Fizzotti, F
    Furetta, C
    Gan, KK
    Ghodbane, N
    Grigoriev, E
    Hallewell, G
    Han, S
    Hartjes, F
    Hrubec, J
    Husson, D
    Kagan, H
    Kaplon, J
    Karl, C
    Kass, R
    Keil, M
    Knöpfle, KT
    Koeth, T
    Krammer, M
    Logiudice, A
    Lu, R
    mac Lynne, L
    Manfredotti, C
    Marshall, RD
    Meier, D
    Menichelli, D
    Meuser, S
    Mishina, M
    Moroni, L
    Noomen, J
    Oh, A
    Perera, L
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 511 (1-2) : 124 - 131
  • [2] ADAM W, 2002, CVD IEEE T NUCL SCI, V49, P530
  • [3] Improvement of the dosimetric properties of chemical-vapor-deposited diamond films by neutron irradiation
    Bruzzi, M
    Menichelli, D
    Pini, S
    Bucciolini, M
    Mólnar, J
    Fenyvesi, A
    [J]. APPLIED PHYSICS LETTERS, 2002, 81 (02) : 298 - 300
  • [4] Diamond dosimetry: Outcomes of the CANDIDO and CONRAD INFN projects
    Bucciolini, M
    Borchi, E
    Bruzzi, M
    Casati, M
    Cirrone, P
    Cuttone, G
    De Angelis, C
    Lovik, I
    Onori, S
    Raffaele, L
    Sciortino, S
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 552 (1-2) : 189 - 196
  • [5] Verification of IMRT fields by film dosimetry
    Bucciolini, M
    Buonamici, FB
    Casati, M
    [J]. MEDICAL PHYSICS, 2004, 31 (01) : 161 - 168
  • [6] Dosimetric characterization of CVD diamonds irradiated with 62 MeV proton beams
    Cirrone, GAP
    Cuttone, G
    Lo Nigro, S
    Mongelli, V
    Raffaele, L
    Sabini, MG
    Valastro, L
    Bucciolini, M
    Onori, S
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 552 (1-2) : 197 - 202
  • [7] Improvements in CVD diamond properties for radiotherapy dosimetry
    De Angelis, C.
    Bucciolini, M.
    Casati, M.
    Lovik, I.
    Bruzzi, M.
    Lagomarsino, S.
    Sciortino, S.
    Onori, S.
    [J]. RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) : 38 - 42
  • [8] An investigation of the operating characteristics of two PTW diamond detectors in photon and electron beams
    De Angelis, C
    Onori, S
    Pacilio, M
    Cirrone, GAP
    Cuttone, G
    Raffaele, L
    Bucciolini, M
    Mazzocchi, S
    [J]. MEDICAL PHYSICS, 2002, 29 (02) : 248 - 254
  • [9] Radiation monitoring with CVD diamonds in BABAR
    Edwards, AJ
    Bruinsma, M
    Burchat, P
    Kagan, H
    Kass, R
    Kirkby, D
    Petersen, BA
    Pulliam, T
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 552 (1-2) : 176 - 182
  • [10] PTW-diamond detector: Dose rate and particle type dependence
    Fidanzio, A
    Azario, L
    Miceli, R
    Russo, A
    Piermattei, A
    [J]. MEDICAL PHYSICS, 2000, 27 (11) : 2589 - 2593