Pulse transit time: validation of blood pressure measurement under positive airway pressure ventilation

被引:19
|
作者
Schmalgemeier, Heidi [1 ]
Bitter, Thomas [1 ]
Bartsch, Stephan [1 ]
Bullert, Kevin [1 ]
Fischbach, Thomas [1 ]
Eckert, Siegfried [1 ]
Horstkotte, Dieter [1 ]
Oldenburg, Olaf [1 ]
机构
[1] Ruhr Univ Bochum, Univ Hosp, Heart & Diabet Ctr N Rhine Westphalia, Dept Cardiol, D-32545 Bad Oeynhausen, Germany
关键词
Pulse transit time; Arterial blood pressure measurement; Continuous positive airway pressure (CPAP); Sleep disordered breathing; OBSTRUCTIVE SLEEP-APNEA; HEART-FAILURE; RISK; ECG;
D O I
10.1007/s11325-011-0609-7
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The objective of the present study was to validate noninvasive, continuous blood pressure (BP) measurement using pulse transit time (PTT) to represent absolute values and detect BP changes under continuous positive airway pressure (CPAP) treatment. We applied CPAP to 78 patients of a cardiological sleep lab using 0 (baseline), 4, 8, and 12 cmH(2)O for 10 min at every level and measured BP simultaneous via PTT-based method and standard oscillometric method (OM). Quality of signal perception was acceptable to convert PTT into BP values in 64 patients (82%). When comparing both methods, we found a strong linear correlation of systolic and diastolic BP (baseline, r = 0.94 for systolic BP; r = 0.95 for diastolic BP, p < 0.001) while no significant differences between absolute values obtained with OM and PTT measurement. Mean bias at baseline was 4.1 +/- 3.2 mmHg for systolic BP and 2.3 +/- 2.2 mmHg for diastolic BP. With increasing CPAP levels, PTT and OM measurements differed continuously up to a systolic difference of 6.6 +/- 4.9 mmHg and a diastolic difference of 4.4 +/- 3.5 mmHg. There was no definite trend of PTT method to either over- or underestimate BP. We found that PTT- and OM-based BP results are closely correlated while applying CPAP over a period of 40 min. With higher CPAP level, bias between both methods increased.
引用
收藏
页码:1105 / 1112
页数:8
相关论文
共 50 条
  • [31] Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method
    Heiko Gesche
    Detlef Grosskurth
    Gert Küchler
    Andreas Patzak
    European Journal of Applied Physiology, 2012, 112 : 309 - 315
  • [32] Ballistocardiogram as Proximal Timing Reference for Pulse Transit Time Measurement: Potential for Cuffless Blood Pressure Monitoring
    Kim, Chang-Sei
    Carek, Andrew M.
    Mukkamala, Ramakrishna
    Inan, Omer T.
    Hahn, Jin-Oh
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (11) : 2657 - 2664
  • [33] An investigation of pulse transit time as a blood pressure measurement method in patients undergoing carotid artery stenting
    Can, Yusuf
    Kilic, Harun
    Akdemir, Ramazan
    Acar, Bilgehan
    Edem, Efe
    Kocyigit, Ibrahim
    Vatan, Mehmet B.
    Aksoy, Murat
    Can, Nimet
    Gunduz, Huseyin
    BLOOD PRESSURE MONITORING, 2016, 21 (03) : 168 - 170
  • [34] Wearable Continuous Blood Pressure Monitoring Devices Based on Pulse Wave Transit Time and Pulse Arrival Time: A Review
    Zhou, Zi-Bo
    Cui, Tian-Rui
    Li, Ding
    Jian, Jin-Ming
    Li, Zhen
    Ji, Shou-Rui
    Li, Xin
    Xu, Jian-Dong
    Liu, Hou-Fang
    Yang, Yi
    Ren, Tian-Ling
    MATERIALS, 2023, 16 (06)
  • [35] Pulse transit time-estimated blood pressure: a comparison of beat-to-beat and intermittent measurement
    Satoshi Hoshide
    Akiomi Yoshihisa
    Fumihiro Tsuchida
    Hiroyuki Mizuno
    Hiroki Teragawa
    Takatoshi Kasai
    Hitoshi Koito
    Shin-ichi Ando
    Yoshihiko Watanabe
    Yasuchika Takeishi
    Kazuomi Kario
    Hypertension Research, 2022, 45 : 1001 - 1007
  • [36] Acute Effects of Continuous Positive Airway Pressure on Pulse Pressure in CHF
    Blanco, Jacobo Bacariza
    Esquinas, Antonio M.
    ARQUIVOS BRASILEIROS DE CARDIOLOGIA, 2015, 104 (02) : 175 - 176
  • [37] Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference
    Gao, Mingwu
    Olivier, N. Bari
    Mukkamala, Ramakrishna
    PHYSIOLOGICAL REPORTS, 2016, 4 (10):
  • [38] Relationship between pulse transit time and blood pressure is impaired in patients with chronic heart failure
    Daniel R. Wagner
    Norbert Roesch
    Patrick Harpes
    Heinrich Körtke
    Pierre Plumer
    Amir Saberin
    Viviane Chakoutio
    Denis Oundjede
    Charles Delagardelle
    Jean Beissel
    Georges Gilson
    Ingrid Kindermann
    Michael Böhm
    Clinical Research in Cardiology, 2010, 99 : 657 - 664
  • [39] PULSE TRANSIT-TIME - RELATIONSHIP TO BLOOD-PRESSURE AND MYOCARDIAL PERFORMANCE
    OBRIST, PA
    LIGHT, KC
    MCCUBBIN, JA
    HUTCHESON, JS
    HOFFER, JL
    PSYCHOPHYSIOLOGY, 1979, 16 (03) : 292 - 301
  • [40] Towards Ubiquitous Blood Pressure Monitoring in an Armband using Pulse Transit Time
    Toreyin, Hakan
    Javaid, Abdul Q.
    Ashouri, Hazar
    Ode, Oludotun
    Inan, Omer T.
    2015 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS), 2015, : 560 - 563