Applicability and accuracy improvement of transient elastography using the M and XL probes by experienced operators

被引:21
作者
Carrion, J. A. [1 ,2 ]
Puigvehi, M. [1 ,2 ]
Coll, S. [1 ,2 ]
Garcia-Retortillo, M. [1 ,2 ]
Canete, N. [1 ,2 ]
Fernandez, R. [1 ,2 ]
Marquez, C. [1 ,2 ]
Gimenez, M. D. [1 ,2 ]
Garcia, M. [3 ]
Bory, F. [1 ]
Sola, R. [1 ,2 ]
机构
[1] Univ Autonoma Barcelona, Liver Sect, Dept Gastroenterol, Hosp Mar, E-08193 Barcelona, Spain
[2] Hosp Mar, Med Res Inst, IMIM, Barcelona 08003, Spain
[3] Hosp Mar, Dept Pathol, Barcelona 08003, Spain
关键词
fibrosis; liver; noninvasive; stiffness; LIVER STIFFNESS MEASUREMENT; CHRONIC HEPATITIS-C; FIBROSIS; DIAGNOSIS; VIRUS;
D O I
10.1111/jvh.12296
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Transient elastography (TE) is the reference method to obtain liver stiffness measurements (LSM), but no results are obtained in 3.1% and unreliable in 15.8%. We assessed the applicability and diagnostic accuracy of TE re-evaluation using M and XL probes. From March 2011 to April 2012 868 LSM were performed with the M probe by trained operators (50-500 studies) (LSM1). Measurements were categorized as inadequate (no values or ratio <60% and/or IQR/LSM>30%) or adequate. Inadequate LSM1 were re-evaluated by experienced operators (>500 explorations) (LSM2) and inadequate LSM2 using XL probe (LSMXL). Inadequate LSM1 were obtained in 187 (21.5%) patients, IQR/LSM >30% in 97 (51%), ratio <60% in 24 (13%) and TE failed to obtain a measurement in 67 (36%). LSM2 achieved adequate registers in 123 (70%) of 175 registers previously considered as inadequate. Independent variables (OR, 95%CI) related to inadequate LSM1 were body mass index (1.11, 1.04-1.18), abdominal circumference (1.03, 1.01-1.06) and age (1.03, 1.01-1.04) and to inadequate LSM2 were skin-capsule distance (1.21, 1.09-1.34) and abdominal circumference (1.05, 1.01-1.10). The diagnostic accuracy (AUROC) to identify significant fibrosis improved from 0.89 (LSM1) to 0.91 (LSM2) (P=0.046) in 334 patients with liver biopsy or clinically significant portal hypertension. A third evaluation (LSMXL) obtained adequate registers in 41 (93%) of 44 patients with inadequate LSM2. Operator experience increases the applicability and diagnostic accuracy of TE. The XL probe may be recommended for patients with inadequate values obtained by experienced operators using the M probe. (NCT01900808).
引用
收藏
页码:297 / 306
页数:10
相关论文
共 19 条
[1]   Liver Stiffness Is Influenced by a Standardized Meal in Patients With Chronic Hepatitis C Virus at Different Stages of Fibrotic Evolution [J].
Arena, Umberto ;
Platon, Monica Lupsor ;
Stasi, Cristina ;
Moscarella, Stefania ;
Assarat, Ali ;
Bedogni, Giorgio ;
Piazzolla, Valeria ;
Badea, Radu ;
Laffi, Giacomo ;
Marra, Fabio ;
Mangia, Alessandra ;
Pinzani, Massimo .
HEPATOLOGY, 2013, 58 (01) :65-72
[2]   Determination of reliability criteria for liver stiffness evaluation by transient elastography [J].
Boursier, Jerome ;
Zarski, Jean-Pierre ;
de Ledinghen, Victor ;
Rousselet, Marie-Christine ;
Sturm, Nathalie ;
Lebail, Brigitte ;
Fouchard-Hubert, Isabelle ;
Gallois, Yves ;
Oberti, Frederic ;
Bertrais, Sandrine ;
Cales, Paul .
HEPATOLOGY, 2013, 57 (03) :1182-1191
[3]   Hepatic elastography. Position paper of the Catalan Society of Gastroenterology [J].
Carrion, Jose A. ;
Navasa, Miquel ;
Buti, Maria ;
Torras, Xavier ;
Xiol, Xavier ;
Vergara, Mercedes ;
Planas, Ramon ;
Sola, Ricard ;
Forns, Xavier .
GASTROENTEROLOGIA Y HEPATOLOGIA, 2011, 34 (07) :504-510
[4]   Liver Stiffness Identifies Two Different Patterns of Fibrosis Progression in Patients with Hepatitis C Virus Recurrence After Liver Transplantation [J].
Carrion, Jose A. ;
Torres, Ferran ;
Crespo, Gonzalo ;
Miquel, Rosa ;
Garcia-Valdecasas, Juan-Carlos ;
Navasa, Miquel ;
Forns, Xavier .
HEPATOLOGY, 2010, 51 (01) :23-34
[5]   Noninvasive Methods to Assess Liver Disease in Patients With Hepatitis B or C [J].
Castera, Laurent .
GASTROENTEROLOGY, 2012, 142 (06) :1293-+
[6]   Pitfalls of Liver Stiffness Measurement: A 5-Year Prospective Study of 13,369 Examinations [J].
Castera, Laurent ;
Foucher, Juliette ;
Bernard, Pierre-Henri ;
Carvalho, Francoise ;
Allaix, Daniele ;
Merrouche, Wassil ;
Couzigou, Patrice ;
de Ledinghen, Victor .
HEPATOLOGY, 2010, 51 (03) :828-835
[7]   Feasibility of liver transient elastography with FibroScan® using a new probe for obese patients [J].
de Ledinghen, Victor ;
Vergniol, Julien ;
Foucher, Juliette ;
El-Hajbi, Farid ;
Merrouche, Wassil ;
Rigalleau, Vincent .
LIVER INTERNATIONAL, 2010, 30 (07) :1043-1048
[8]   Diagnostic accuracy of FibroScan and comparison to liver fibrosis biomarkers in chronic viral hepatitis: A multicenter prospective study (the FIBROSTIC study) [J].
Degos, Francoise ;
Perez, Paul ;
Roche, Bruno ;
Mahmoudi, Amel ;
Asselineau, Julien ;
Voitot, Helene ;
Bedossa, Pierre .
JOURNAL OF HEPATOLOGY, 2010, 53 (06) :1013-1021
[9]   Noninvasive markers of fibrosis: key concepts for improving accuracy in daily clinical practice [J].
Duarte-Rojo, Andres ;
Trinidad Altamirano, Jose ;
Feld, Jordan J. .
ANNALS OF HEPATOLOGY, 2012, 11 (04) :426-439
[10]   Reproducibility of transient elastography in the evaluation of liver fibrosis in patients with chronic liver disease [J].
Fraquelli, Mirella ;
Rigamonti, Cristina ;
Casazza, Giovanni ;
Conte, Dario ;
Donato, Maria Francesca ;
Ronchi, Guido ;
Colombo, Massimo .
GUT, 2007, 56 (07) :968-973