Microstructures and bond strengths of plasma-sprayed hydroxyapatite coatings on porous titanium substrates

被引:47
作者
Oh, IK
Nomura, N
Chiba, A
Murayama, Y
Masahashi, N
Lee, BT
Hanada, S
机构
[1] Korea Inst Ind Technol, Kwangju 506824, South Korea
[2] Iwate Univ, Dept Welf Engn, Morioka, Iwate 0208551, Japan
[3] Tohoku Univ, Inst Mat Res, Aoba Ku, Sendai, Miyagi 9808577, Japan
[4] Kongju Natl Univ, Sch Adv Mat Engn, Kong Ju 314701, Chungnam, South Korea
关键词
D O I
10.1007/s10856-005-2534-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Hydroxyapatite (HA) coating was carried out by plasma spraying on bulk Ti substrates and porous Ti substrates having a Young's modulus similar to that of human bone. The microstructures and bond strengths of HA coatings were investigated in this study. The HA coatings with thickness of 200-250 mu m were free from cracks at interfaces between the coating and Ti substrates. XRD analysis revealed that the HA powder used for plasma spraying had a highly crystallized apatite structure, while the HA coating contained several phases other than HA. The bond strength between the HA coating and the Ti substrates evaluated by standard bonding test ASTM C633-01) were strongly affected by the failure behavior of the HA coating. A mechanism to explain the failure is discussed in terms of surface roughness of the plasma-sprayed HA coatings on the bulk and porous Ti substrates. (C) 2005 Springer Science + Business Media, Inc.
引用
收藏
页码:635 / 640
页数:6
相关论文
共 30 条
[1]   RESIDUAL-STRESS MEASUREMENT IN THERMAL SPRAYED HYDROXYAPATITE COATINGS [J].
BROWN, SR ;
TURNER, IG ;
REITER, H .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1994, 5 (9-10) :756-759
[2]   Addressing processing problems associated with plasma spraying of hydroxyapatite coatings [J].
Cheang, P ;
Khor, KA .
BIOMATERIALS, 1996, 17 (05) :537-544
[3]   Plasma-sprayed hydroxyapatite coating on titanium alloy with ZrO2 second phase and ZrO2 intermediate layer [J].
Chou, BY ;
Chang, E .
SURFACE & COATINGS TECHNOLOGY, 2002, 153 (01) :84-92
[4]   Hydroxyapatite-Ti functionally graded biomaterial fabricated by powder metallurgy [J].
Chu, CL ;
Zhu, JC ;
Yin, ZD ;
Wang, SD .
MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 1999, 271 (1-2) :95-100
[5]  
Collier J P, 1993, J Arthroplasty, V8, P389, DOI 10.1016/S0883-5403(06)80037-9
[6]  
COOK SD, 1991, CLIN ORTHOP RELAT R, V265, P280
[7]   IN-VIVO MECHANICAL AND HISTOLOGICAL CHARACTERISTICS OF HA-COATED IMPLANTS VARY WITH COATING VENDOR [J].
DALTON, JE ;
COOK, SD .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (02) :239-245
[8]   ON THE MECHANICAL-BEHAVIOR OF BRITTLE COATINGS AND LAYERS [J].
EVANS, AG ;
CRUMLEY, GB ;
DEMARAY, RE .
OXIDATION OF METALS, 1983, 20 (5-6) :193-216
[9]   Phase transformations in plasma sprayed hydroxyapatite coatings [J].
Feng, CF ;
Khor, KA ;
Liu, EJ ;
Cheang, P .
SCRIPTA MATERIALIA, 1999, 42 (01) :103-109
[10]   BONDING OF BONE TO APATITE-COATED IMPLANTS [J].
GEESINK, RGT ;
DEGROOT, K ;
KLEIN, CPAT .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (01) :17-22