Fat quantification using three-point dixon technique: In vitro validation

被引:44
作者
Kovanlikaya, A
Guclu, C
Desai, C
Becerra, R
Gilsanz, V
机构
[1] Childrens Hosp Los Angeles, Dept Radiol, Los Angeles, CA 90027 USA
[2] Gen Elect Med Syst, Waukesha, WI USA
关键词
fat content; MRI; three-point Dixon;
D O I
10.1016/j.acra.2005.01.019
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. To test the repeatability, reproducibility and accuracy of the three-point Dixon (3PD) sequence for estimating true fat volume ratios using a fat/water phantom. Materials and Methods. A phantom, constructed from test tubes of varying fat content, was imaged using the 3PD sequence on a 1.5T MRI scanner by two operators four times each. Fat volume ratios were calculated from these images and compared with true fat volumes. Results. Measures of fat volume ratios calculated from the 3PD MR images correlated strongly with values for true fat volumes (r = 0.96). Conclusion. The 3PD technique was found to be highly reproducible and accurate, and may be useful for in vivo quantification of fat in lean tissues, such as the liver, pancreas or skeletal muscle. © AUR, 2005.
引用
收藏
页码:636 / 639
页数:4
相关论文
共 49 条
[21]   Quantification of pancreas fat on dual-energy computed tomography: comparison with six-point Dixon magnetic resonance imaging [J].
Fumi Kameda ;
Masahiro Tanabe ;
Hideko Onoda ;
Mayumi Higashi ;
Shoko Ariyoshi ;
Kenichiro Ihara ;
Etsushi Iida ;
Matakazu Furukawa ;
Munemasa Okada ;
Katsuyoshi Ito .
Abdominal Radiology, 2020, 45 :2779-2785
[22]   Quantification of pancreas fat on dual-energy computed tomography: comparison with six-point Dixon magnetic resonance imaging [J].
Kameda, Fumi ;
Tanabe, Masahiro ;
Onoda, Hideko ;
Higashi, Mayumi ;
Ariyoshi, Shoko ;
Ihara, Kenichiro ;
Lida, Etsushi ;
Furukawa, Matakazu ;
Okada, Munemasa ;
Ito, Katsuyoshi .
ABDOMINAL RADIOLOGY, 2020, 45 (09) :2779-2785
[23]   Pelvic imaging using a T1W fat-suppressed three-dimensional dual echo dixon technique at 3T [J].
Cornfeld, Daniel M. ;
Israel, Gary ;
McCarthy, Shirley M. ;
Weinreb, Jeffery C. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (01) :121-127
[24]   Carotid and aortic plaque imaging using 3D gradient-echo imaging and the three-point Dixon method with improved motion-sensitized driven-equilibrium (iMSDE) [J].
Koori, Norikazu ;
Kamekawa, Hiroki ;
Naito, Takehiro ;
Takatsu, Yasuo ;
Fuse, Hiraku ;
Miyakawa, Shin ;
Yasue, Kenji ;
Takahashi, Masato ;
Kurata, Kazuma .
MAGNETIC RESONANCE IMAGING, 2024, 111 :202-209
[25]   Correlation between Vertebral Marrow Fat Fraction in MRI Using DIXON Technique and BMD in DXA in Patients of Suspected Osteoporosis [J].
Naik, Suprava ;
Jain, Mantu ;
Bhoi, Sanjeev Kumar ;
Tripathy, Sujit .
INDIAN JOURNAL OF RADIOLOGY AND IMAGING, 2024, 34 (02) :239-245
[26]   Fat suppression at three-dimensional T1-weighted MR imaging of the hands: Dixon method versus CHESS technique [J].
Kirchgesner, T. ;
Perlepe, V. ;
Michoux, N. ;
Larbi, A. ;
Vande Berg, B. .
DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2018, 99 (01) :23-28
[27]   Usefulness of two-point Dixon fat-water separation technique in gadoxetic acid-enhanced liver magnetic resonance imaging [J].
Ding, Ying ;
Rao, Sheng-Xiang ;
Chen, Cai-Zhong ;
Li, Ren-Chen ;
Zeng, Meng-Su .
WORLD JOURNAL OF GASTROENTEROLOGY, 2015, 21 (16) :5017-5022
[28]   Quantification of Fatty Degeneration Within the Supraspinatus Muscle by Using a 2-Point Dixon Method on 3-T MRI [J].
Nozaki, Taiki ;
Tasaki, Atsushi ;
Horiuchi, Saya ;
Osakabe, Chiharu ;
Ohde, Sachiko ;
Saida, Yukihisa ;
Yoshioka, Hiroshi .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 205 (01) :116-122
[29]   Quantification of spinal bone marrow fat fraction using three-material decomposition technique on dual-energy CT: A phantom study [J].
Li, Baojun ;
Hua, Ning ;
Li, Janelle ;
Andreu-Arasa, V. Carlota ;
LeBedis, Christina ;
Anderson, Stephan W. .
MEDICAL PHYSICS, 2023, 50 (07) :4105-4111
[30]   Correcting gradient chain induced fat quantification errors in radial multi-echo Dixon imaging using a gradient modulation transfer function [J].
Zoellner, Christoph ;
Kronthaler, Sophia ;
Weiss, Kilian ;
Boehm, Christof ;
Stelter, Jonathan ;
Rahmer, Juergen ;
Boernert, Peter ;
Peeters, Johannes M. ;
Junker, Daniela ;
Karampinos, Dimitrios C. .
FRONTIERS IN PHYSICS, 2023, 11