The burden of the variability introduced by the HEp-2 assay kit and the CAD system in ANA indirect immunofluorescence test

被引:27
作者
Infantino, M. [1 ]
Meacci, F. [1 ]
Grossi, V. [1 ]
Manfredi, M. [1 ]
Benucci, M. [2 ]
Merone, M. [3 ]
Soda, P. [3 ]
机构
[1] S Giovanni di Dio Hosp, Immunol & Allergy Lab, Via Torregalli 3, I-50143 Florence, Italy
[2] S Giovanni di Dio Hosp, Rheumatol Unit, Via Torregalli 3, I-50143 Florence, Italy
[3] Univ Campus Biomed, Dept Engn, Comp Syst & Bioinformat Lab, Rome, Italy
关键词
Indirect immunofluorescence; HEp-2; Antinuclear antibodies; Automated classification system; Standardization; ANTINUCLEAR ANTIBODIES; AUTOANTIBODIES; GUIDELINES; DIAGNOSIS; IMMUNOASSAYS; CHALLENGES;
D O I
10.1007/s12026-016-8845-3
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
According to the recent recommendations of the American College of Rheumatology, ANATask Force, IIF technique should be considered the gold standard in antinuclear antibodies (ANAs) testing. To overcome the lack of standardization, biomedical industries have developed several computer-aided diagnosis (CAD) systems. Two hundred and sixty-one consecutive samples with suspected autoimmune diseases were tested for ANA by means of IIF on routinely HEp-2 assay kit (Euroimmun AG). Assignment of result wasmade if consensus for positive/negative was reached by at least 2 out of 3 expert physicians. ANAIIF was also carried out using 3 CAD systems: Zenit G-Sight (n = 84), Helios (n = 85) and NOVA View (n = 92); human evaluation was repeated on the same substrate of each CAD system (Immco, Aesku and Inova HEp-2 cells, respectively). To anonymize the results, we randomly named these three systems as A, B and C. We ran a statistical analysis computing several measures of agreement between the ratings, and we also improved the evaluation by using the Wilcoxon's test for nonparametric data. Agreement between the human readings on routinely HEp-2 assay kit and human readings on CAD HEp-2 assay was substantial for A (k = 0.82) and B (k = 0.72), and almost perfect for C (k = 0.89). Such readingswere statistically different only in case A. Comparing experts' readings with the readings of CAD systems, when the samples were prepared using CAD HEp-2 assay kits, we found almost perfect agreement for B and C (k = 0.86; k = 0.82) and substantial agreement for A (k = 0.73). Again, human and CAD readings were statistically different only in A. When we compared the readings of medical experts on routinely HEp-2 assay kit with the output of the CAD systems thatworked using their own slides, we found substantial agreement for all the systems (A: k = 0.62; B: k = 0.65; C: k = 0.71). Such readingswere not statistically different. The change of the assay kit and/or the introduction of a CAD system affect the laboratory reporting, with an evident impact on the autoimmune laboratory workflow. The CAD systems may represent one of themost important novel elements of harmonization in the autoimmunity field, reducing intra-and inter-laboratory variability in a new vision of the diagnostic autoimmune platform.
引用
收藏
页码:345 / 354
页数:10
相关论文
共 59 条
[21]   Benchmarking HEp-2 Cells Classification Methods [J].
Foggia, Pasquale ;
Percannella, Gennaro ;
Soda, Paolo ;
Vento, Mario .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (10) :1878-1889
[22]   Challenges to the use of autoantibodies as predictors of disease onset, diagnosis and outcomes [J].
Fritzler, Marvin J. .
AUTOIMMUNITY REVIEWS, 2008, 7 (08) :616-620
[23]   The Emergence of Multiplexed Technologies as Diagnostic Platforms in Systemic Autoimmune Diseases [J].
Fritzler, Marvin J. ;
Fritzler, Mark L. .
CURRENT MEDICINAL CHEMISTRY, 2006, 13 (21) :2503-2512
[24]   Commercial blot assays in the diagnosis of systemic rheumatic diseases [J].
Ghirardello, Anna ;
Bendo, Raffaele ;
Rampudda, Maria Elisa ;
Bassi, Nicola ;
Zampieri, Sandra ;
Doria, Andrea .
AUTOIMMUNITY REVIEWS, 2009, 8 (08) :645-649
[25]   Establishment of HEp-2 cell preparation for automated analysis of ANA fluorescence pattern [J].
Hahm, D ;
Anderer, U .
CYTOMETRY PART A, 2006, 69A (03) :178-181
[26]   Automatic analysis of immunofluorescence patterns of HEp-2 cells [J].
Hiemann, Rico ;
Hilger, Nadja ;
Michel, Joerg ;
Nitschke, Joerg ;
Boehm, Alexander ;
Anderer, Ursula ;
Weigert, Martin ;
Sack, Ulrich .
AUTOIMMUNITY, PART A: BASIC PRINCIPLES AND NEW DIAGNOSTIC TOOLS, 2007, 1109 :358-371
[27]   Challenges of automated screening and differentiation of non-organ specific autoantibodies on HEp-2 cells [J].
Hiemann, Rico ;
Buettner, Thomas ;
Krieger, Thorsten ;
Roggenbuc, Dirk ;
Sack, Ulrich ;
Conrad, Karsten .
AUTOIMMUNITY REVIEWS, 2009, 9 (01) :17-22
[28]  
Kavanaugh A, 2000, ARCH PATHOL LAB MED, V124, P71
[29]   A novel automated indirect immunofluorescence autoantibody evaluation [J].
Kivity, Shaye ;
Gilburd, Boris ;
Agmon-Levin, Nancy ;
Garcia Carrasco, Marina ;
Tzafrir, Yaron ;
Sofer, Yael ;
Mandel, Matilda ;
Buttner, Thomas ;
Roggenbuck, Dirk ;
Matucci-Cerinic, Marco ;
Danko, Katalin ;
Lopez Hoyos, Marcos ;
Shoenfeld, Yehuda .
CLINICAL RHEUMATOLOGY, 2012, 31 (03) :503-509
[30]   MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174