Flow system device for the on-line determination of total mercury in seawater

被引:28
作者
Wurl, O
Elsholz, O
Ebinghaus, R
机构
[1] Univ Appl Sci, D-21033 Hamburg, Germany
[2] GKSS Forschungszentrum Geesthacht GmbH, Inst Phys & Chem Anal, D-21502 Geesthacht, Germany
关键词
mercury; seawater; cold vapour atomic absorption spectrometry;
D O I
10.1016/S0039-9140(99)00343-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The instrumental set-up for the on-line sampling, digestion and quantification of total mercury in seawater is described. Based on a flow system and cold vapour atomic absorption spectrometry (CVAAS) detection limit was improved by a gold amalgam preconcentration to 0.5 ng l(-1). The manifold design was optimized by the variation of the gas flow, length of reaction coils, shape and material of the gas-liquid separator and flow rate of the reductant. A calibration following the equations p = 0.039c + 0.0813 (p, peak area; c, concentration) and a correlation coefficient of r = 0.9996 was obtained. The relative standard deviation of three measurements of 0.5 ng l(-1) Hg2+ was 3.8%. The long-time (12 h) reproducibility was 6.2% RSD (n = 25) of 0.5 ng l(-1) Hg2+. With a recovery rate >90% mercury can be determined after on-line UV digestion. For a complete analysis, 6 min is required. The technique is fast, simple to handle and robust. The apparatus is designed for the use on research vessels under sea conditions. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:51 / 57
页数:7
相关论文
共 13 条
[11]   ONLINE MICROWAVE SAMPLE PRETREATMENT FOR THE DETERMINATION OF MERCURY IN WATER AND URINE BY FLOW-INJECTION COLD-VAPOR ATOMIC-ABSORPTION SPECTROMETRY [J].
WELZ, B ;
TSALEV, DL ;
SPERLING, M .
ANALYTICA CHIMICA ACTA, 1992, 261 (1-2) :91-103
[12]  
Welz B., 1999, Atomic Absorption Spectrometry
[13]  
WURL O, 1999, IN PRESS FRESENIUS J