Comparison of a new piezoelectric train-of-four neuromuscular monitor, the ParaGraph, and the Relaxometer mechanomyograph

被引:18
作者
Dahaba, AA
Von Klobucar, F
Rehak, PH
List, WF
机构
[1] Karl Franzens Univ Graz, Dept Anaesthesiol & Intens Care Med, LKH, A-8036 Graz, Austria
[2] Karl Franzens Univ Graz, Dept Surg, Biomed Engn & Comp Unit, A-8010 Graz, Austria
关键词
neuromuscular block; cisatracurium; monitoring; neuromuscular function; equipment; myographs;
D O I
10.1093/bja/82.5.780
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The ParaGraph is a new device for monitoring neuromuscular function using a piezoelectric motion sensor. In 20 patients, monitoring of neuromuscular block produced by cisatracurium 0.1 mg kg(-1) was compared using the ParaGraph and a Relaxometer 2 mechanomyograph. The ParaGraph was quick to set up, and easy to operate and interpret. There were no significant differences in the time to 100% depression of T1/T0, time to 25% recovery of T1/T0 or time to recovery of T1/T0 from 25% to 75%, measured by the two monitors. When the difference between the two monitors was plotted against the average of the two measurements, the limits of agreement for T1/T0 (-42.95, +53.98%) and the train-of-four ratio, T4/T1 (-0.28, +0.21) were too wide to allow the values given by the two monitors for individual patients to be used interchangeably.
引用
收藏
页码:780 / 782
页数:3
相关论文
共 6 条
[1]   STIMULUS FREQUENCY AND DOSE-RESPONSE CURVE TO D-TUBOCURARINE IN MAN [J].
ALI, HH ;
SAVARESE, JJ .
ANESTHESIOLOGY, 1980, 52 (01) :36-39
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
KERN SE, 1994, ANESTH ANALG, V78, P978
[4]   THE TOF-GUARD NEUROMUSCULAR-TRANSMISSION MONITOR - A COMPARISON WITH THE MYOGRAPH-2000 [J].
LOAN, PB ;
PAXTON, LD ;
MIRAKHUR, RK ;
CONNOLLY, FM ;
MCCOY, EP .
ANAESTHESIA, 1995, 50 (08) :699-702
[5]   COMPARISON OF THUMB ACCELERATION AND THENAR EMG IN A PHARMACODYNAMIC STUDY OF ALCURONIUM [J].
MERETOJA, OA ;
WERNER, MU ;
WIRTAVUORI, K ;
LUOSTO, T .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1989, 33 (07) :545-548
[6]   Good clinical research practice (GCRP) in pharmacodynamic studies of neuromuscular blocking agents [J].
VibyMogensen, J ;
Engbaek, J ;
Eriksson, LI ;
Gramstad, L ;
Jensen, E ;
Jensen, FS ;
KoscielniakNielsen, Z ;
Skovgaard, LT ;
Ostergaard, D .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1996, 40 (01) :59-74