On-line analysis of a continuous-flow ozonolysis reaction using Raman spectroscopy

被引:16
|
作者
Pelletier, M. J. [1 ]
Fabiilli, M. L. [1 ]
Moon, Brian [1 ]
机构
[1] Pfizer Global Res & Dev, Res Analyt, Pharmaceut Sci, Devlopmental API, Ann Arbor, MI 48105 USA
关键词
Raman; quantitative analysis; ozonolysis; trans-stilbene; benzaldehyde; alpha-methoxybenzyl hydroperoxide; spectral stripping; partial least squares; PLS;
D O I
10.1366/000370207782217725
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe the use of an on-line Raman analyzer to quantitatively track the levels of trans-stilbene, benzaldehyde, and alpha-metboxybenzyl hydroperoxide in a continuous flow ozonolysis reactor. Quantitative analysis is carried out using spectral stripping in order to overcome baseline artifacts inherent in simple peak area determinations and to incorporate prior knowledge into the analytical model. The performance of spectral stripping is compared to partial least squares (PLS) analysis. The effect of two process variables, feed flow rate and ozone concentration, are also explored. Finally, the benefit of utilizing an on-line technique is illustrated by demonstrating the bias introduced by sampling for off-line analysis.
引用
收藏
页码:1107 / 1115
页数:9
相关论文
共 50 条
  • [21] CONTINUOUS-FLOW ANALYSIS - THE AUTOANALYZER
    CRANDELL, C
    JOURNAL OF AUTOMATIC CHEMISTRY, 1985, 7 (03): : 145 - 148
  • [22] IMPROVEMENTS IN CONTINUOUS-FLOW ANALYSIS
    CONETTA, A
    DORSHEIMER, WT
    DUCROS, MJF
    AMERICAN LABORATORY, 1981, 13 (09) : 116 - &
  • [23] On-line monitoring of trihalomethanes in drinking water using continuous-flow purge and cryofocusing gas chromatography-mass spectrometry
    Chen, TC
    Her, GR
    JOURNAL OF CHROMATOGRAPHY A, 2001, 927 (1-2) : 229 - 235
  • [24] Factors affecting on-line estimation of diastereomer composition using Raman spectroscopy
    Wong, Sze-Wing
    Georgakis, Christos
    Botsaris, Gregory
    Saranteas, Kostas
    Bakale, Roger
    CONTROL ENGINEERING PRACTICE, 2007, 15 (10) : 1257 - 1267
  • [25] On-line characterization of YBCO coated conductors using Raman spectroscopy methods
    Maroni, V. A.
    Reeves, J. L.
    Schwab, G.
    APPLIED SPECTROSCOPY, 2007, 61 (04) : 359 - 366
  • [26] SIMPLE CONTINUOUS-FLOW HELIUM CRYOSTAT FOR SPECTROSCOPY
    AURICH, F
    MARQUARD, J
    SCHUMACHER, KD
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1975, 8 (06): : 447 - 448
  • [27] On-line monitoring of chlorosilane streams by Raman spectroscopy
    Lipp, ED
    Grosse, RL
    APPLIED SPECTROSCOPY, 1998, 52 (01) : 42 - 46
  • [28] Raman spectroscopy for on-line process chemistry.
    Adar, F
    Morel, S
    Geiger, R
    Whitley, A
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2003, 226 : U125 - U125
  • [29] Aspirin synthesis on-line measurement by Raman spectroscopy
    Li, WH
    Lü, GW
    Huang, L
    Gao, H
    Du, WM
    ACTA PHYSICO-CHIMICA SINICA, 2003, 19 (02) : 105 - 108
  • [30] AN IMPROVED CONTINUOUS-FLOW METHOD FOR SERUM CREATININE USING THE JAFFE REACTION
    WHITE, WA
    ATTWOOD, EC
    ANNALS OF CLINICAL BIOCHEMISTRY, 1980, 17 (MAY) : 153 - 154