The effects of extending the spectral information acquired by a photon-counting detector for spectral CT

被引:12
作者
Schmidt, Taly Gilat [1 ]
Zimmerman, Kevin C. [1 ]
Sidky, Emil Y. [2 ]
机构
[1] Marquette Univ, Dept Biomed Engn, Milwaukee, WI 53233 USA
[2] Univ Chicago, Dept Radiol, Chicago, IL 60637 USA
关键词
photon counting; spectral CT; computed tomography; BASIS IMAGE NOISE; COMPUTED-TOMOGRAPHY; RAY; FEASIBILITY; STATISTICS;
D O I
10.1088/0031-9155/60/4/1583
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Photon-counting x-ray detectors with pulse-height analysis provide spectral information that may improve material decomposition and contrast-tonoise ratio (CNR) in CT images. The number of energy measurements that can be acquired simultaneously on a detector pixel is equal to the number of comparator channels. Some spectral CT designs have a limited number of comparator channels, due to the complexity of readout electronics. The spectral information could be extended by changing the comparator threshold levels over time, sub pixels, or view angle. However, acquiring more energy measurements than comparator channels increases the noise and/or dose, due to differences in noise correlations across energy measurements and decreased dose utilisation. This study experimentally quantified the effects of acquiring more energy measurements than comparator channels using a bench-top spectral CT system. An analytical and simulation study modeling an ideal detector investigated whether there was a net benefit for material decomposition or optimal energy weighting when acquiring more energy measurements than comparator channels. Experimental results demonstrated that in a two-threshold acquisition, acquiring the high-energy measurement independently from the low-energy measurement increased noise standard deviation in material-decomposition basis images by factors of 1.5-1.7 due to changes in covariance between energy measurements. CNR in energy-weighted images decreased by factors of 0.92-0.71. Noise standard deviation increased by an additional factor of root 2 due to reduced dose utilisation. The results demonstrated no benefit for two-material decomposition noise or energy-weighted CNR when acquiring more energy measurements than comparator channels. Understanding the noise penalty of acquiring more energy measurements than comparator channels is important for designing spectral detectors and for designing experiments and interpreting data from prototype systems with a limited number of comparator channels.
引用
收藏
页码:1583 / 1600
页数:18
相关论文
共 18 条
[11]   Optimal "image-based" weighting for energy-resolved CT [J].
Schmidt, Taly Gilat .
MEDICAL PHYSICS, 2009, 36 (07) :3018-3027
[12]   Photon counting computed tomography: Concept and initial results [J].
Shikhaliev, PM ;
Xu, T ;
Molloi, S .
MEDICAL PHYSICS, 2005, 32 (02) :427-436
[13]   Computed tomography with energy-resolved detection: a feasibility study [J].
Shikhaliev, Polad M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (05) :1475-1495
[14]   Vision 20/20: Single photon counting x-ray detectors in medical imaging [J].
Taguchi, Katsuyuki ;
Iwanczyk, Jan S. .
MEDICAL PHYSICS, 2013, 40 (10) :100901
[15]   An analytical model of the effects of pulse pileup on the energy spectrum recorded by energy resolved photon counting x-ray detectors [J].
Taguchi, Katsuyuki ;
Frey, Eric C. ;
Wang, Xiaolan ;
Iwanczyk, Jan S. ;
Barber, William C. .
MEDICAL PHYSICS, 2010, 37 (08) :3957-3969
[16]   SNR AND DQE ANALYSIS OF BROAD-SPECTRUM X-RAY-IMAGING [J].
TAPIOVAARA, MJ ;
WAGNER, RF .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (06) :519-529
[17]   Sufficient Statistics as a Generalization of Binning in Spectral X-ray Imaging [J].
Wang, Adam S. ;
Pelc, Norbert J. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2011, 30 (01) :84-93
[18]   Penalized-likelihood estimators and noise analysis for randoms-precorrected PET transmission scans [J].
Yavuz, M ;
Fessler, JA .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (08) :665-674