The effects of extending the spectral information acquired by a photon-counting detector for spectral CT

被引:12
作者
Schmidt, Taly Gilat [1 ]
Zimmerman, Kevin C. [1 ]
Sidky, Emil Y. [2 ]
机构
[1] Marquette Univ, Dept Biomed Engn, Milwaukee, WI 53233 USA
[2] Univ Chicago, Dept Radiol, Chicago, IL 60637 USA
关键词
photon counting; spectral CT; computed tomography; BASIS IMAGE NOISE; COMPUTED-TOMOGRAPHY; RAY; FEASIBILITY; STATISTICS;
D O I
10.1088/0031-9155/60/4/1583
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Photon-counting x-ray detectors with pulse-height analysis provide spectral information that may improve material decomposition and contrast-tonoise ratio (CNR) in CT images. The number of energy measurements that can be acquired simultaneously on a detector pixel is equal to the number of comparator channels. Some spectral CT designs have a limited number of comparator channels, due to the complexity of readout electronics. The spectral information could be extended by changing the comparator threshold levels over time, sub pixels, or view angle. However, acquiring more energy measurements than comparator channels increases the noise and/or dose, due to differences in noise correlations across energy measurements and decreased dose utilisation. This study experimentally quantified the effects of acquiring more energy measurements than comparator channels using a bench-top spectral CT system. An analytical and simulation study modeling an ideal detector investigated whether there was a net benefit for material decomposition or optimal energy weighting when acquiring more energy measurements than comparator channels. Experimental results demonstrated that in a two-threshold acquisition, acquiring the high-energy measurement independently from the low-energy measurement increased noise standard deviation in material-decomposition basis images by factors of 1.5-1.7 due to changes in covariance between energy measurements. CNR in energy-weighted images decreased by factors of 0.92-0.71. Noise standard deviation increased by an additional factor of root 2 due to reduced dose utilisation. The results demonstrated no benefit for two-material decomposition noise or energy-weighted CNR when acquiring more energy measurements than comparator channels. Understanding the noise penalty of acquiring more energy measurements than comparator channels is important for designing spectral detectors and for designing experiments and interpreting data from prototype systems with a limited number of comparator channels.
引用
收藏
页码:1583 / 1600
页数:18
相关论文
共 18 条
  • [1] ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY
    ALVAREZ, RE
    MACOVSKI, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) : 733 - 744
  • [2] Estimator for photon counting energy selective x-ray imaging with multibin pulse height analysis
    Alvarez, Robert E.
    [J]. MEDICAL PHYSICS, 2011, 38 (05) : 2324 - 2334
  • [3] A cascaded model of spectral distortions due to spectral response effects and pulse pileup effects in a photon-counting x-ray detector for CT
    Cammin, Jochen
    Xu, Jennifer
    Barber, William C.
    Iwanczyk, Jan S.
    Hartsough, Neal E.
    Taguchi, Katsuyuki
    [J]. MEDICAL PHYSICS, 2014, 41 (04)
  • [4] NOISE DUE TO PHOTON-COUNTING STATISTICS IN COMPUTED X-RAY TOMOGRAPHY
    CHESLER, DA
    RIEDERER, SJ
    PELC, NJ
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (01) : 64 - 74
  • [5] Multi-energy performance of a research prototype CT scanner with small-pixel counting detector
    Kappler, S.
    Henning, A.
    Krauss, B.
    Schoeck, F.
    Stierstorfer, K.
    Weidinger, T.
    Flohr, T.
    [J]. MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [6] Optimization of Medipix-2 Threshold Masks for Spectroscopic X-Ray Imaging
    Procz, Simon
    Luebke, Joerdis
    Zwerger, Andreas
    Mix, Michael
    Fiederle, Michael
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (04) : 1795 - 1799
  • [7] K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors
    Roessl, E.
    Proksa, R.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (15) : 4679 - 4696
  • [8] Cramer-Rao lower bound of basis image noise in multiple-energy x-ray imaging
    Roessl, E.
    Herrmann, C.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (05) : 1307 - 1318
  • [9] On the influence of noise correlations in measurement data on basis image noise in dual-energylike x-ray imaging
    Roessl, Ewald
    Ziegler, Andy
    Proksa, Roland
    [J]. MEDICAL PHYSICS, 2007, 34 (03) : 959 - 966
  • [10] Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography
    Schlomka, J. P.
    Roessl, E.
    Dorscheid, R.
    Dill, S.
    Martens, G.
    Istel, T.
    Baeumer, C.
    Herrmann, C.
    Steadman, R.
    Zeitler, G.
    Livne, A.
    Proksa, R.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (15) : 4031 - 4047