Adjusting mean arterial pressure alarms improves the time spent within blood pressure targets in patients with septic shock: A quasi-experimental study

被引:1
作者
Florens, Nans [1 ]
Chabert, Beatrice [1 ]
Stevic, Neven [1 ]
Vareon, Julie [1 ]
Malatray, Arnaud [1 ]
Argaud, Laurent [1 ]
Cour, Martin [1 ]
机构
[1] Hosp Civils Lyon, Grp Hosp Ctr, Serv Med Intens Reanimat, Hop Edouard Herriot, F-69437 Lyon, France
关键词
Mean arterial pressure; Norepinephrine; Monitoring; Septic shock; DRUG INFUSION PUMPS; VASOPRESSOR THERAPY; NOREPINEPHRINE; MULTICENTER; CHANGEOVERS; SCORE;
D O I
10.1016/j.aucc.2020.10.002
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Norepinephrine is the first-line vasoactive drug in septic shock. As underdosages or overdosages can be harmful for patients, it seems useful to maintain the mean arterial pressure (MAP) within preset bounds. Objectives: We sought to assess whether adjusted MAP alarms could improve MAP control in patients with septic shock. Methods: We conducted a quasi-experimental before-and-after study. During two consecutive periods, data on MAP control were obtained from patients with septic shock (n 1/4 50/period) treated with norepinephrine over more than 24 h. The norepinephrine administration protocol, including prescription of the MAP target range (e.g., 65-75 mmHg), was identical during the two periods. During the first period (control group), the preset alarms of the monitor were used (i.e., low and high systolic blood pressure alarms set at 90 and 160 mmHg, respectively). During the second period, adjusted MAP alarms were implemented, with upper and lower bounds corresponding to the prescribed MAP target range (MAP-Alarm group). The primary end point was the percentage of time outside the desired MAP target range during the first 24 h of norepinephrine infusion. Results: Baseline characteristics were not significantly different. The primary end point was significantly lower in the MAP-Alarm group than in the control group (25 +/- 13% versus 51 +/- 18%, respectively; p < 0.01). MAP was higher than the target 14 +/- 11% of the time in the MAP-Alarm group versus 37 +/- 17% in the control group (p < 0.01) and lower than the target 11 +/- 9% of the time in the MAP-Alarm versus 21 +/- 22% in the control group (p < 0.05). There was no significant difference between the two groups with regard to the dose of norepinephrine, duration of norepinephrine administration, and survival. Conclusions: These results suggest that adjusting MAP alarms to the desired MAP target range could dramatically improve the percentage of time spent within MAP targets in patients with septic shock but does not reduce exposure to norepinephrine. (c) 2020 Australian College of Critical Care Nurses Ltd. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:358 / 362
页数:5
相关论文
共 20 条
  • [1] Changeovers of vasoactive drug infusion pumps: impact of a quality improvement program
    Argaud, Laurent
    Cour, Martin
    Martin, Olivier
    Saint-Denis, Marc
    Ferry, Tristan
    Goyatton, Agnes
    Robert, Dominique
    [J]. CRITICAL CARE, 2007, 11 (06):
  • [2] High versus Low Blood-Pressure Target in Patients with Septic Shock
    Asfar, Pierre
    Meziani, Ferhat
    Hamel, Jean-Francois
    Grelon, Fabien
    Megarbane, Bruno
    Anguel, Nadia
    Mira, Jean-Paul
    Dequin, Pierre-Francois
    Gergaud, Soizic
    Weiss, Nicolas
    Legay, Francois
    Le Tulzo, Yves
    Conrad, Marie
    Robert, Rene
    Gonzalez, Frederic
    Guitton, Christophe
    Tamion, Fabienne
    Tonnelier, Jean-Marie
    Guezennec, Pierre
    Van der Linden, Thierry
    Vieillard-Baron, Antoine
    Mariotte, Eric
    Pradel, Gael
    Lesieur, Olivier
    Ricard, Jean-Damien
    Herve, Fabien
    du Cheyron, Damien
    Guerin, Claude
    Mercat, Alain
    Teboul, Jean-Louis
    Radermacher, Peter
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2014, 370 (17) : 1583 - 1593
  • [3] Quasi-experimental study designs series-paper 1: introduction: two historical lineages
    Baernighausen, Till
    Rottingen, John-Arne
    Rockers, Peter
    Shemilt, Ian
    Tugwell, Peter
    [J]. JOURNAL OF CLINICAL EPIDEMIOLOGY, 2017, 89 : 4 - 11
  • [4] Benefits of smart pumps for automated changeovers of vasoactive drug infusion pumps: a quasi-experimental study
    Cour, M.
    Hernu, R.
    Benet, T.
    Robert, J. M.
    Regad, D.
    Chabert, B.
    Malatray, A.
    Conrozier, S.
    Serra, P.
    Lassaigne, M.
    Vanhems, P.
    Argaud, L.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2013, 111 (05) : 818 - 824
  • [5] Predictors of haemodynamic instability during the changeover of norepinephrine infusion pumps
    Cour, Martin
    Benet, Thomas
    Hernu, Romain
    Simon, Marie
    Baudry, Thomas
    Vanhems, Philippe
    Argaud, Laurent
    [J]. ANNALS OF INTENSIVE CARE, 2016, 6
  • [6] BLOOD PRESSURE TARGETS FOR VASOPRESSOR THERAPY: A SYSTEMATIC REVIEW
    D'Aragon, Frederick
    Belley-Cote, Emilie P.
    Meade, Maureen O.
    Lauzier, Francois
    Adhikari, Neill K. J.
    Briel, Matthias
    Lalu, Manoj
    Kanji, Salmaan
    Asfar, Pierre
    Turgeon, Alexis F.
    Fox-Robichaud, Alison
    Marshall, John C.
    Lamontagne, Francois
    [J]. SHOCK, 2015, 43 (06): : 530 - 539
  • [7] Comparison of Dopamine and Norepinephrine in the Treatment of Shock.
    De Backer, Daniel
    Biston, Patrick
    Devriendt, Jacques
    Madl, Christian
    Chochrad, Didier
    Aldecoa, Cesar
    Brasseur, Alexandre
    Defrance, Pierre
    Gottignies, Philippe
    Vincent, Jean-Louis
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2010, 362 (09) : 779 - 789
  • [8] Norepinephrine in septic shock: when and how much?
    Hamzaoui, Olfa
    Scheeren, Thomas W. L.
    Teboul, Jean-Louis
    [J]. CURRENT OPINION IN CRITICAL CARE, 2017, 23 (04) : 342 - 347
  • [9] Target blood pressure for septic and vasodilatory shock: what is optimal?
    Ishii, Junki
    Ohshimo, Shinichiro
    Shime, Nobuaki
    Lamontagne, Francois
    Guyatt, Gordon
    Asfar, Pierre
    [J]. INTENSIVE CARE MEDICINE, 2018, 44 (04) : 548 - 549
  • [10] Feasibility of closed-loop titration of norepinephrine infusion in patients undergoing moderate- and high-risk surgery
    Joosten, Alexandre
    Alexander, Brenton
    Duranteau, Jacques
    Taccone, Fabio Silvio
    Creteur, Jacques
    Vincent, Jean-Louis
    Cannesson, Maxime
    Rinehart, Joseph
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2019, 123 (04) : 430 - 438